Ray-tracing analytical absorption correction for X-ray crystallography based on tomographic reconstructions

crossmark logo

aOxford e-Research Centre, Department of Engineering Science, University of Oxford, 7 Keble Road, Oxford OX1 3QG, United Kingdom, bDiamond Light Source, Harwell Science & Innovation Campus, Didcot OX11 0DE, United Kingdom, cRosalind Franklin Institute, Harwell Science & Innovation Campus, Didcot OX11 0QX, United Kingdom, dRutherford Appleton Laboratory, Research Complex at Harwell, Didcot OX11 0FA, United Kingdom, and eDepartment of Life Sciences, Imperial College London, Exhibition Road, London SW7 2AZ, United Kingdom
*Correspondence e-mail: wes.armour@oerc.ox.ac.uk, armin.wagner@diamond.ac.uk

Edited by A. Barty, DESY, Hamburg, Germany (Received 24 November 2023; accepted 7 March 2024; online 15 April 2024)

Processing of single-crystal X-ray diffraction data from area detectors can be separated into two steps. First, raw intensities are obtained by integration of the diffraction images, and then data correction and reduction are performed to determine structure-factor amplitudes and their uncertainties. The second step considers the diffraction geometry, sample illumination, decay, absorption and other effects. While absorption is only a minor effect in standard macromolecular crystallography (MX), it can become the largest source of uncertainty for experiments performed at long wavelengths. Current software packages for MX typically employ empirical models to correct for the effects of absorption, with the corrections determined through the procedure of minimizing the differences in intensities between symmetry-equivalent reflections; these models are well suited to capturing smoothly varying experimental effects. However, for very long wavelengths, empirical methods become an unreliable approach to model strong absorption effects with high fidelity. This problem is particularly acute when data multiplicity is low. This paper presents an analytical absorption correction strategy (implemented in new software AnACor) based on a volumetric model of the sample derived from X-ray tomography. Individual path lengths through the different sample materials for all reflections are determined by a ray-tracing method. Several approaches for absorption corrections (spherical harmonics correction, analytical absorption correction and a combination of the two) are compared for two samples, the membrane protein OmpK36 GD, measured at a wavelength of λ = 3.54 Å, and chlorite dismutase, measured at λ = 4.13 Å. Data set statistics, the peak heights in the anomalous difference Fourier maps and the success of experimental phasing are used to compare the results from the different absorption correction approaches. The strategies using the new analytical absorption correction are shown to be superior to the standard spherical harmonics corrections. While the improvements are modest in the 3.54 Å data, the analytical absorption correction outperforms spherical harmonics in the longer-wavelength data (λ = 4.13 Å), which is also reflected in the reduced amount of data being required for successful experimental phasing.

1. Introduction

In X-ray crystallography, intensities of reflections are proportional to the square of their structure-factor amplitudes ([I_{\bf h}\propto] [|F_{\bf h}|^{2}]). Several factors need to be considered when calculating structure-factor amplitudes from measured intensities, such as Lorentz, polarization, sample illumination, decay and absorption corrections (Monaco & Artioli, 2002[Monaco, H. L. & Artioli, G. (2002). Fundamentals of Crystallography, 2nd ed., edited by H. Giacovazzo, ch. 5, pp. 376-388. Oxford University Press.]). Away from absorption edges, sample absorption is approximately proportional to the cube of the wavelength (Arndt, 1984[Arndt, U. W. (1984). J. Appl. Cryst. 17, 118-119.]). It depends on the chemical composition, density, and shape and size of the sample which includes the crystal, as well as the surrounding materials like sample mount, mother liquor, or oils and glues used to mount the crystals. High-quality structure determination relies on accurate structure-factor amplitudes. Hence, correcting the measured intensities by calculating absorption correction factors is critical. For a crystal which is not surrounded by mother liquor or mounted in a loop, the Bragg intensities after absorption correction are given by [I_{\rm corr} = I_{\rm meas}/A_{\bf h}], and the absorption correction factor [A_{\bf h}] for the reflection h in a crystallography experiment is given by

[A_{\bf h} = {{1} \over {V}}\int\limits_{V}\exp\{-\mu[L_{1}(x,y,z)+L_{2}(x,y,z)]\}\,{\rm d}V, \eqno(1)]

where L1(x, y, z) and L2(x, y, z) (hereafter referred to as L1 and L2) are the incident and diffracted X-ray path lengths for each crystal element dV, and μ is the absorption coefficient of the crystal (Albrecht, 1939[Albrecht, G. (1939). Rev. Sci. Instrum. 10, 221-222.]). Since the resulting volumetric integral calculation is intractable for irregularly shaped crystals, absorption correction for multi-faced crystals has been performed by numerical methods (Busing & Levy, 1957[Busing, W. R. & Levy, H. A. (1957). Acta Cryst. 10, 180-182.]; DeTitta, 1985[DeTitta, G. T. (1985). J. Appl. Cryst. 18, 75-79.]). As an alternative approach, the crystal can be partitioned into fundamental tetrahedra to calculate the integral over all the tetrahedra (Howells, 1950[Howells, R. G. (1950). Acta Cryst. 3, 366-369.]; de Meulenaer & Tompa, 1965[Meulenaer, J. de & Tompa, H. (1965). Acta Cryst. 19, 1014-1018.]; Clark & Reid, 1995[Clark, R. C. & Reid, J. S. (1995). Acta Cryst. A51, 887-897.]). Both analytical and numerical absorption corrections require an accurate description of the shape and dimensions of the crystal. One solution from the APEX3 software (Bruker, 2012[Bruker (2012). APEX. Bruker AXS Inc., Madison, Wisconsin, USA.]) is to determine and index all the crystal faces visually and perform an analytical absorption correction. However, this is difficult when the shape of the crystal is not a regular polyhedron. In addition, the presence of other materials surrounding the crystal, such as mother liquor and sample mount, adds further complication: these materials with different absorption coefficients only contribute to the absorption effect, not to the diffraction. Semi-empirical methods (North et al., 1968[North, A. C. T., Phillips, D. C. & Mathews, F. S. (1968). Acta Cryst. A24, 351-359.]; Kopfmann & Huber, 1968[Kopfmann, G. & Huber, R. (1968). Acta Cryst. A24, 348-351.]) based on intensity measurements and assumptions on the incident and diffracted beams do not rely on knowledge of the sample shape. However, they require multi-axis goniometers, and the additional data needed for the azimuthal scans can contribute significantly to radiation damage on modern synchrotron light sources. Empirical methods which are independent of the sample geometry were developed either based on Fourier series of the incident and diffracted beams (Katayama et al., 1972[Katayama, C., Sakabe, N. & Sakabe, K. (1972). Acta Cryst. A28, 293-295.]; Walker & Stuart, 1983[Walker, N. & Stuart, D. (1983). Acta Cryst. A39, 158-166.]) or by using spherical harmonics (Blessing, 1995[Blessing, R. H. (1995). Acta Cryst. A51, 33-38.]) to minimize the residual between the intensities for symmetry-related reflections. With the introduction of large area detectors, these numerical methods to obtain an empirical correction for absorption have become popular. Spherical harmonics are now the basis for absorption correction in most data reduction software packages for macromolecular crystallography (MX), such as AIMLESS (Evans & Murshudov, 2013[Evans, P. R. & Murshudov, G. N. (2013). Acta Cryst. D69, 1204-1214.]), hkl3000 (Minor et al., 2006[Minor, W., Cymborowski, M., Otwinowski, Z. & Chruszcz, M. (2006). Acta Cryst. D62, 859-866.]), SADABS (Sheldrick, 1996[Sheldrick, G. M. (1996). SADABS. University of Göttingen, Germany.]) and DIALS (Winter et al., 2018[Winter, G., Waterman, D. G., Parkhurst, J. M., Brewster, A. S., Gildea, R. J., Gerstel, M., Fuentes-Montero, L., Vollmar, M., Michels-Clark, T., Young, I. D., Sauter, N. K. & Evans, G. (2018). Acta Cryst. D74, 85-97.]; Beilsten-Edmands et al., 2020[Beilsten-Edmands, J., Winter, G., Gildea, R., Parkhurst, J., Waterman, D. & Evans, G. (2020). Acta Cryst. D76, 385-399.]), while XDS uses alternative numerical methods without spherical harmonics (Kabsch, 2010[Kabsch, W. (2010). Acta Cryst. D66, 125-132.]). However, the efficacy of empirical methods depends on having a large number of symmetry-equivalent reflections, which can be difficult to achieve when data multiplicity is low, e.g. in the case of radiation-sensitive crystals in low-symmetry space groups.

As the analytical absorption correction does not depend on refining parameters to minimize differences between structure-factor amplitudes of symmetry-related reflections, its success does not rely on data multiplicity. To analytically calculate absorption correction factors for a sample with irregular shape, its shape and orientation have to be characterized in detail. Previous work using optical microscopy to reconstruct a 3D model of the sample, containing crystal, sample mount and mother liquor, showed that absorption correction was viable and advantageous at lower levels of data multiplicity (Leal et al., 2008[Leal, R. M. F., Teixeira, S. C. M., Rey, V., Forsyth, V. T. & Mitchell, E. P. (2008). J. Appl. Cryst. 41, 729-737.]; Strutz, 2011[Strutz, T. (2011). IEEE/ACM Trans. Comput. Biol. Bioinf. 8, 797-807.]). An alternative approach to obtain a 3D model of the sample is X-ray tomography, which has been applied to either characterize or visualize crystals (Merrifield et al., 2011[Merrifield, D. R., Ramachandran, V., Roberts, K. J., Armour, W., Axford, D., Basham, M., Connolley, T., Evans, G., McAuley, K. E., Owen, R. L. & Sandy, J. (2011). Meas. Sci. Technol. 22, 115703.]; Warren et al., 2013[Warren, A. J., Armour, W., Axford, D., Basham, M., Connolley, T., Hall, D. R., Horrell, S., McAuley, K. E., Mykhaylyk, V., Wagner, A. & Evans, G. (2013). Acta Cryst. D69, 1252-1259.]). The use of tomographic reconstructions and segmentations as a basis for absorption correction has previously been suggested by Brockhauser et al. (2008[Brockhauser, S., Di Michiel, M., McGeehan, J. E., McCarthy, A. A. & Ravelli, R. B. G. (2008). J. Appl. Cryst. 41, 1057-1066.]). This enables the calculation of X-ray path lengths through the different materials in the sample (crystal, sample mount and mother liquor), as illustrated in Fig. 1[link].

[Figure 1]
Figure 1
A sketch illustrating the ray-tracing method used to calculate an absorption correction factor for a crystal voxel n. L(n)m1 and L(n)m2 represent the path lengths of the incident and diffracted X-ray beams through the material m (loop, liquor and crystal).

While X-ray absorption is not normally considered an issue at standard wavelengths in MX, it is a major limiting factor in long-wavelength crystallography. Beamline I23 at Diamond Light Source, UK (Wagner et al., 2016[Wagner, A., Duman, R., Henderson, K. & Mykhaylyk, V. (2016). Acta Cryst. D72, 430-439.]), is a unique synchrotron instrument operating in a wavelength range between 1.1 and 5.9 Å, giving access to the absorption edges of several light elements of biological significance, such as calcium, potassium, chlorine, sulfur and phosphorus. The largest anomalous signal for sulfur is expected close to its absorption edge (λ = 5.02 Å). However, the difficulties in correcting for increased sample absorption at very long wavelengths compromise the overall data quality, resulting in reduced measured anomalous signal. Applying standard absorption correction protocols, the optimal wavelength for single-wavelength anomalous diffraction experiments based on sulfur (S-SAD) is found to be λ = 2.75 Å (El Omari et al., 2023[El Omari, K., Duman, R., Mykhaylyk, V., Orr, C. M., Latimer-Smith, M., Winter, G., Grama, V., Qu, F., Bountra, K., Kwong, H. S., Romano, M., Reis, R. I., Vogeley, L., Vecchia, L., Owen, C. D., Wittmann, S., Renner, M., Senda, M., Matsugaki, N., Kawano, Y., Bowden, T. A., Moraes, I., Grimes, J. M., Mancini, E. J., Walsh, M. A., Guzzo, C. R., Owens, R. J., Jones, E. Y., Brown, D. G., Stuart, D. I., Beis, K. & Wagner, A. (2023). Commun. Chem. 6, 219.]), clearly indicating the need for more sophisticated methods to exploit the full potential of long-wavelength crystallography.

In this paper, we introduce AnACor, a computer program that employs a ray-tracing method to estimate the path lengths of the incident and diffracted X-rays through the sample from a tomographic reconstruction, to calculate absorption correction factors for long-wavelength X-ray diffraction data. The effectiveness of AnACor is demonstrated for long-wavelength data sets collected at 3.54 Å, on a crystal of the membrane protein OmpK36 GD, and at 4.13 Å, on a crystal of the heme-binding enzyme chlorite dismutase (Cld). OmpK36 GD, referred to as simply `OmpK36', is a 373 amino acid outer membrane porin from Klebsiella pneumonia involved in nutrient and antibiotic diffusion in gram negative bacteria (Wong et al., 2019[Wong, J. L., Romano, M., Kerry, L. E., Kwong, H.-S., Low, W.-W., Brett, S. J., Clements, A., Beis, K. & Frankel, G. (2019). Nat. Commun. 10, 1-10.]), while Cld is a heme-b-containing homodimeric oxidoreductase from Cyanothece sp. PCC7425, consisting of 181 amino acids per monomer. The choice of these two samples for this study was motivated by their crystallization in low-symmetry space groups, posing a challenge for the conventional absorption correction methods used in standard X-ray diffraction scaling programs.

2. Methods

2.1. Experiment workflow and data preparation

Crystals of OmpK36 were prepared and cryo-protected as previously described with no modification (Wong et al., 2019[Wong, J. L., Romano, M., Kerry, L. E., Kwong, H.-S., Low, W.-W., Brett, S. J., Clements, A., Beis, K. & Frankel, G. (2019). Nat. Commun. 10, 1-10.]). OmpK36 crystallized as rods in space group C2, with three monomers present in the asymmetric unit. Large sample-to-sample variations required extensive screening of crystals. The crystal selected for this study had dimensions of 260 × 30 × 30 µm. Cld crystals were produced using a protocol based on previously reported conditions (Schaffner et al., 2017[Schaffner, I., Mlynek, G., Flego, N., Pühringer, D., Libiseller-Egger, J., Coates, L., Hofbauer, S., Bellei, M., Furtmüller, P. G., Battistuzzi, G., Smulevich, G., Djinović-Carugo, K. & Obinger, C. (2017). ACS Catal. 7, 7962-7976.]) with further details provided in the supporting information, section S2. The crystal used in this study had dimensions of 190 × 150 × 90 µm and indexed in space group P1, with two monomers in the asymmetric unit.

All experiments were performed at the long-wavelength MX beamline I23 at Diamond Light Source, UK. The in-vacuum sample environment comprises the cylindrical P12M detector and a multi-axis goniometer to enable collection of complete diffraction data from crystals in low-symmetry space groups even at the longest wavelengths. A tomography camera is integrated into the beamline sample environment, allowing easy transition between the two experimental modes (Kazantsev et al., 2021[Kazantsev, D., Duman, R., Wagner, A., Mykhaylyk, V., Wanelik, K., Basham, M. & Wadeson, N. (2021). J. Synchrotron Rad. 28, 889-901.]). The sample preparation for in-vacuum data collection followed the standard protocol for beamline I23 (Duman et al., 2021[Duman, R., Orr, C. M., Mykhaylyk, V., El Omari, K., Pocock, R., Grama, V. & Wagner, A. (2021). J. Vis. Exp. 170, e62364.]). For the OmpK36 crystal, 3 × 360° of data were collected at a wavelength of λ = 3.54 Å with 0.1 s exposure per 0.1° rotation angle and a beam transmission of 50%, with a top-hat X-ray beam adjusted to 240 × 150 µm. To ensure completeness of the data, two of the three data sets were collected using kappa goniometry, with the kappa axis rotated to −70° and the phi axis positioned at 0° and −120°. Each of the three data sets was measured with a photon flux of 1.36 × 1011 photons s−1, which resulted in a total absorbed dose of 6.5 MGy per data set, as calculated by Raddose3D (Zeldin et al., 2013[Zeldin, O. B., Gerstel, M. & Garman, E. F. (2013). J. Appl. Cryst. 46, 1225-1230.]). Since the Cld crystal diffracted to higher resolution than the OmpK36 crystal, we chose a low-dose data collection strategy. In total 22 × 360° were collected at a wavelength of λ = 4.13 Å with a 350 × 350 µm top-hat beam, using an exposure of 0.1 s per 0.1°. With a beam transmission of 5%, the measured photon flux of 6.7 × 109 photons s−1 yielded an absorbed dose of 0.1 MGy per data set. Two of the 22 data sets were collected with the kappa and phi goniometer axes at 0°, while the rest were recorded at κ = −70° and 20 different phi values, between −120° and 120°. The diffraction data were indexed and integrated with DIALS (Winter et al., 2018[Winter, G., Waterman, D. G., Parkhurst, J. M., Brewster, A. S., Gildea, R. J., Gerstel, M., Fuentes-Montero, L., Vollmar, M., Michels-Clark, T., Young, I. D., Sauter, N. K. & Evans, G. (2018). Acta Cryst. D74, 85-97.]), providing a kappa/phi orientation matrix, raw intensities, incident vectors, scattering vectors and goniometer angles.

The diffraction experiment was immediately followed by tomography data collection at the same X-ray wavelength. One 180° tomography data set was collected for each crystal, with the kappa and phi axes set at 0° and a beam size of 700 × 700 µm and 100% transmission, using a propagation distance of 4.9 mm between scintillator and sample. For OmpK36 1800 projections, 30 flat-field images (without sample) and 30 dark images (without X-rays) were collected with an exposure of 0.15 s per 0.1° rotation. The measured flux for this data set was 1.5 × 1012 photons s−1, resulting in a total absorbed dose of 4.8 MGy. For the Cld crystal, 900 projections, 20 flat-field and 20 dark images were collected with an exposure of 0.28 s per 0.2° rotation and a measured flux of 4.3 × 1011 photons s−1, yielding a total absorbed dose of 0.8 MGy.

The tomography data were processed using the SAVU pipeline (Kazantsev et al., 2022[Kazantsev, D., Wadeson, N. & Basham, M. (2022). SoftwareX, 19, 101157.]), with a processing routine consisting of standard flat-field correction, followed by ring artefact removal (Vo et al., 2018[Vo, N. T., Atwood, R. C. & Drakopoulos, M. (2018). Opt. Express, 26, 28396-28412.]) and reconstruction. For OmpK36, the reconstruction step was performed by iterative methods via the ToMoBAR module in SAVU (Kazantsev et al., 2021[Kazantsev, D., Duman, R., Wagner, A., Mykhaylyk, V., Wanelik, K., Basham, M. & Wadeson, N. (2021). J. Synchrotron Rad. 28, 889-901.]), as its edge-enhancing properties gave improved results. For Cld, where the data showed better contrast, the filter-back projection (TomoPy) module (Gürsoy et al., 2014[Gürsoy, D., De Carlo, F., Xiao, X. & Jacobsen, C. (2014). J. Synchrotron Rad. 21, 1188-1193.]) was used instead. No contrast transfer function correction was applied in the processing. Flat-field images, raw projections and flat-field-corrected projections for both samples are shown in Fig. 2[link]. For ease of segmentation, reconstruction was performed on cropped data, to eliminate as much of the background as possible and reduce the size of the images. The OmpK36 data were cropped from an initial volume of 1600 × 1200 × 1200 voxels to 1220 × 1001 × 1001 voxels, while the Cld data were reduced to 1310 × 1181 × 1181 voxels. The pixel size in the tomography images, determined from previous beamline calibrations, was 0.3 × 0.3 µm. Manual segmentation was performed with the visualization software Avizo (Thermo Fisher), providing a 3D model with every voxel annotated as one of the different sample materials. On the basis of the sample 3D models, the absorption correction factors were calculated and exported to the scaling module in DIALS (Beilsten-Edmands et al., 2020[Beilsten-Edmands, J., Winter, G., Gildea, R., Parkhurst, J., Waterman, D. & Evans, G. (2020). Acta Cryst. D76, 385-399.]) to further correct the diffraction intensities. Published structures, Protein Data Bank (PDB) ID 6rck (Wong et al., 2019[Wong, J. L., Romano, M., Kerry, L. E., Kwong, H.-S., Low, W.-W., Brett, S. J., Clements, A., Beis, K. & Frankel, G. (2019). Nat. Commun. 10, 1-10.]) for OmpK36, and PDB ID 5mau (Schaffner et al., 2017[Schaffner, I., Mlynek, G., Flego, N., Pühringer, D., Libiseller-Egger, J., Coates, L., Hofbauer, S., Bellei, M., Furtmüller, P. G., Battistuzzi, G., Smulevich, G., Djinović-Carugo, K. & Obinger, C. (2017). ACS Catal. 7, 7962-7976.]) for Cld, were used as starting models for the Dimple pipeline (https://ccp4.github.io/dimple/). The `- - anode' option (Thorn & Sheldrick, 2011[Thorn, A. & Sheldrick, G. M. (2011). J. Appl. Cryst. 44, 1285-1287.]) was used to calculate anomalous difference Fourier maps and anomalous peak heights and the option `- - free-r-flags' in the Refmac refinement (Murshudov et al., 1997[Murshudov, G. N., Vagin, A. A. & Dodson, E. J. (1997). Acta Cryst. D53, 240-255.]) step ensured the same Rfree flags for all absorption correction strategies. The Crank2 phasing pipeline (Skubák & Pannu, 2013[Skubák, P. & Pannu, N. S. (2013). Nat. Commun. 4, 2777.]) was used for experimental phasing by single-wavelength anomalous diffraction (SAD) with identical input parameters for the different strategies: the AFRO and PRASA modules were chosen for the FA estimation and substructure determination steps, respectively, with the latter step using 4000 trials and resolution cutoffs of 2.7 Å for Cld and 3.4 Å for OmpK36.

[Figure 2]
Figure 2
Tomography projection images for background [(a) and (d)], sample [(b) and (e)] and flat-field-corrected images [(c) and (f)] of OmpK36 (top) and Cld samples (bottom).

2.2. Analytical absorption correction

For the calculation of the absorption correction factors, the integral [equation (1)[link]] is calculated over the crystal volume (Angel, 2004[Angel, R. J. (2004). J. Appl. Cryst. 37, 486-492.]) as the only source of X-ray diffraction. To move from the continuous integral in equation (1)[link] to a discrete equation, we replace crystal elements dV by crystal voxels ΔV from the tomographic reconstruction (Leal et al., 2008[Leal, R. M. F., Teixeira, S. C. M., Rey, V., Forsyth, V. T. & Mitchell, E. P. (2008). J. Appl. Cryst. 41, 729-737.]). This allows substitution of the integral over the volume V with a sum over the crystal voxels. Hence, the integral in equation (1)[link] can be rewritten discretely as

[A_{\bf h} = {{1} \over {N}}\sum\limits_{n = 1}^{N}A_{\bf h}^{(n)},\eqno(2)]

where N is the number of crystal voxels in the 3D model exposed to the X-ray beam. The sample in a crystallography experiment typically contains more than one material; therefore, the calculation of the absorption correction factor [A_{\bf h}^{(n)}] for a crystal voxel can be rewritten as

[A_{\bf h}^{(n)} = \exp\left[-\textstyle\sum\limits_{m = 1}^{M}\mu_{m}L_{m}^{(n)}\right],\eqno(3)]

where Lm(n) represents the sum of the incident path length L(n)m1 and the diffracted path length L(n)m2 through the material m as shown in Fig. 1[link].

The final squared structure-factor amplitudes [|F_{\bf h}|^{2}] are obtained after combining their absorption correction factors with the overall scale factor, Lorentz and polarization corrections, and other standard correction and scaling techniques.

2.3. Absorption coefficients

Absorption coefficients are determined experimentally using the intensity values in the flat-field-corrected tomograms [Figs. 2[link](c), 2[link](f)] as estimates of the ratio between the incident and transmitted intensities. The distances through each material required for the calculation are obtained from the 3D segmentation models. The 3D models of the OmpK36 and Cld samples in different orientations are presented in Fig. 3[link]. To make sure the transmitted intensities on the tomograms and the path lengths from the segmentation model are aligned, a Python script is used to superpose the 2D projection of the model onto the tomogram. The areas of the flat-field-corrected tomograms affected by phase contrast are excluded from the analysis by applying morphological shrinking. Transmission values are taken from areas in the flat-field-corrected projection images where only solvent is present using the pixels with the 50% longest linear path lengths through the mother liquor. Next, Beer–Lambert's law is applied on a pixel-by-pixel basis to calculate the absorption coefficients. The mother liquor absorption coefficient is then defined as the median of the resulting absorption coefficients. This value is used in the calculation of the absorption coefficients for the other materials (e.g. crystal or protein/detergent aggregate) according to their corresponding path lengths. A library of loop absorption coefficients based on tomography reconstructions of empty loops is available for the different loops used on the I23 beamline. The measured absorption coefficients are presented in Table 1[link]. The composition and density of the protein/detergent aggregate are unknown, but its largest absorption coefficient of all materials is consistent with the flat-field-corrected projection image presented in Fig. 2[link](c).

Table 1
Linear absorption coefficients (µm−1) of different materials in OmpK36 (λ = 3.54 Å) and Cld (λ = 4.13 Å) samples

Sample Crystal Mother liquor Loop Protein/detergent aggregate
OmpK36 0.01053 0.01208 0.00931 0.0322
Cld 0.0160 0.01856 0.01724 N/A
[Figure 3]
Figure 3
Volume renderings of segmentations of OmpK36 [(a)–(c)] and Cld [(d)–(f)]. Transparent blue: mother liquor; gold: loop; pink: crystal; green: protein/detergent aggregate.

2.4. Implementation details

A ray-tracing method is applied to compute the path lengths L(n)m for each crystal voxel n of the reflection h in equation (3)[link]. For a crystal voxel n, it assumes an incoming and a diffracted X-ray beam originating from the voxel. These X-rays, after applying the rotational matrix of the goniometer [{\bf R}_{\omega}] of the reflection h, will propagate through the 3D segmented model. The coordinates of each voxel, along with its corresponding material label, are recorded. Then, the path lengths L(n)m of material m can be determined by the distance between the coordinates of the boundaries of the materials. By combining the absorption coefficients of the corresponding materials, the absorption factor [A^{(n)}_{\bf h}] for the crystal voxel n can be determined [equation (3)[link]]. Finally, the total absorption factor [A_{\bf h}] for the reflection h is calculated by summing [A^{(n)}_{\bf h}] for all crystal voxels according to equation (2)[link].

It is computationally intensive to rotate the overall 3D segmented model for each absorption factor calculation according to the rotational matrix of the goniometer [{\bf R}_{\omega}]. Instead, AnACor rotates the vectors of the incoming and diffracted beams to calculate the path lengths by inverting the goniometer matrix. The tomography experiments are always performed at kappa/phi orientations κ = 0° and ϕ = 0°. To correct data from diffraction experiments with varying kappa/phi orientations, it is essential to transform the vectors of both the incoming and diffracted beams with the kappa/phi orientation matrices [({\bf R}_{\kappa} {\bf R}_{\phi})^{-1}] taken from the DIALS experiment model. Hence, the overall transformed vectors of these beams are in the form of [{\bf s}_{t} = ({\bf R}_{\kappa}{\bf R} _{\phi}{\bf R}_{\omega})^{-1} {\bf s}_{r}], where [{\bf s}_{r}] is either the vector of the incoming or that of the diffracted beam taken from the DIALS reflection data. The resulting directional vectors [{\bf s}_{t}] are used in the ray-tracing method. The incident beam is assumed to have a top-hat profile, so no additional beam profile correction is used. If the crystal is larger than the incident X-ray beam, a discriminator in the ray-tracing algorithm is used to determine whether a crystal voxel is inside the X-ray beam.

The absorption correction software AnACor 1.0 is written in Python to facilitate future integration into DIALS (Winter et al., 2018[Winter, G., Waterman, D. G., Parkhurst, J. M., Brewster, A. S., Gildea, R. J., Gerstel, M., Fuentes-Montero, L., Vollmar, M., Michels-Clark, T., Young, I. D., Sauter, N. K. & Evans, G. (2018). Acta Cryst. D74, 85-97.]). In order to enhance computational efficiency, NumPy 1.23.2 (Harris et al., 2020[Harris, C. R., Millman, K. J., van der Walt, S. J., Gommers, R., Virtanen, P., Cournapeau, D., Wieser, E., Taylor, J., Berg, S., Smith, N. J., Kern, R., Picus, M., Hoyer, S., van Kerkwijk, M. H., Brett, M., Haldane, A., del Río, J. F., Wiebe, M., Peterson, P., Gérard-Marchant, P., Sheppard, K., Reddy, T., Weckesser, W., Abbasi, H., Gohlke, C. & Oliphant, T. E. (2020). Nature, 585, 357-362.]) is used for data loading and preprocessing. Numba 0.56.2 (Lam et al., 2015[Lam, S. K., Pitrou, A. & Seibert, S. (2015). Proceedings of the Second Workshop on the LLVM Compiler Infrastructure in HPC, pp. 1-6. Association for Computing Machinery.]) is used for JIT (just-in-time) compilation. A typical protein crystallography data set contains hundreds of thousands of reflections. There are typically millions of crystal voxels in a 3D model, and each path length calculation can involve determining thousands of voxels along the incident and diffracted X-ray paths. Consequently, calculating all absorption correction factors for samples in protein crystallography is computationally expensive. To mitigate this, a systematic sampling method with a sampling interval of 2000 is applied. This sampling approach relies on the sorted arrangement of the crystal voxels, which helps in identifying the subsections of the crystal where the path lengths (L1 and L2) are similar. Selecting every 2000th voxel from this sorted list ensures that sampling is consistently applied across the crystal. Therefore, it can capture the essential characteristics of the sample with far fewer data points, maintaining accuracy in equation (2)[link] calculations while reducing computational load.

Parallel computing is used by the built-in multiprocessing package in Python, and the calculations of all the reflections are evenly distributed to each CPU core. After applying sampling and parallel computing, on a cluster node with 48 CPU cores, the computational time for the analytical absorption correction of one data set of OmpK36 and Cld is about 40 and 30 min, respectively, with total RAM usage of around 200 GB.

To evaluate the accuracy of the ray-tracing method with and without tomographic volume sampling, the absorption factor calculations were compared with previously published numerical solutions (Maslen, 2004[Maslen, E. N. (2004). International Tables for Crystallography, 3rd ed., edited by E. Prince, Vol. C, ch. 6.3.3, pp. 600-608. Dordrecht: Kluwer.]). Three simulated shapes were considered: cubic, cylindrical and spherical, consisting of crystal material only. For consistency, a voxel size of 0.3 × 0.3 µm and the same sampling interval of 2000 were applied. Both approaches gave errors smaller than 0.5% for cubic and cylindrical shapes. The errors for the spherical shape were smaller than 0.75% with the exception of those at 90°. The results for a smaller voxel size of 0.1 × 0.1 µm indicate that the error is dominated by the pixel size rather than the sampling. More details can be found in the supporting information, section S1.

The codes and further explanations of the algorithm are available at https://github.com/yishunlu-222/AnACor_public.

2.5. Absorption correction strategies

Data scaling is performed by the dials.scale program in DIALS (Beilsten-Edmands et al., 2020[Beilsten-Edmands, J., Winter, G., Gildea, R., Parkhurst, J., Waterman, D. & Evans, G. (2020). Acta Cryst. D76, 385-399.]) using the following custom scaling model:

[g_{{\bf h}l} = C_{{\bf h}l}T_{{\bf h}l}S_{{\bf h}l}A_{{\bf h}l},\eqno(4)]

where [g_{{\bf h}l}] is the overall inverse scale factor that needs to be determined for the lth observation of symmetry-unique reflection h. The scale factors are determined by optimizing the scaling model parameters using a least-squares target function as previously described (Beilsten-Edmands et al., 2020[Beilsten-Edmands, J., Winter, G., Gildea, R., Parkhurst, J., Waterman, D. & Evans, G. (2020). Acta Cryst. D76, 385-399.]). [C_{{\bf h}l}], [T_{{\bf h}l}] and [S_{{\bf h}l}] are, respectively, the scale term, the decay term and the spherical harmonics correction term of the default physical model. The absorption correction factors [A_{{\bf h}l}] are precalculated by AnACor for each reflection [{\bf h}l] and not optimized during the scaling process.

The scale term [C_{{\bf h}l}] models intensity variations as a function of rotation, while the decay term [T_{{\bf h}l}] is a function of resolution and rotation. The spherical harmonics term [S_{{\bf h}l}] corrects the intensities with a model dependent on the incoming and scattered beam paths. The `absorption_level = high' option in dials.scale (Winter et al., 2022[Winter, G., Beilsten-Edmands, J., Devenish, N., Gerstel, M., Gildea, R. J., McDonagh, D., Pascal, E., Waterman, D. G., Williams, B. H. & Evans, G. (2022). Protein Sci. 31, 232-250.]) was used for all approaches that included this term, which reduces the program's restraints on [S_{{\bf h}l}] and uses six orders of spherical harmonics basis functions, to allow high and complex levels of absorption to be modelled. The `anomalous = False' option in dials.scale was used, as the low multiplicity of individual data sets was found to lead to unstable error model refinement for some data sets when the option `anomalous = True' was used.

To evaluate the analytical absorption correction by ray-tracing in AnACor, four approaches are compared:

(i) No absorption correction (labelled as NO) ([g_{{\bf h}l} = C_{{\bf h}l}T_{{\bf h}l}]).

(ii) Spherical harmonics correction (default in dials.scale, SH) ([g_{{\bf h}l} = C_{{\bf h}l}T_{{\bf h}l}S_{{\bf h}l}]).

(iii) Analytical absorption correction described in this work (AC) ([g_{{\bf h}l} = C_{{\bf h}l}T_{{\bf h}l}A_{{\bf h}l}]).

(iv) Analytical absorption correction described in this work, combined with spherical harmonics correction (ACSH) ([g_{{\bf h}l} = C_{{\bf h}l}T_{{\bf h}l}S_{{\bf h}l}A_{{\bf h}l}]).

The parameters for each part of the scaling model (except [A_{{\bf h}l}]) are jointly refined against the integrated intensities in each case and therefore will be different in each approach, i.e. [g^{\rm ACSH}_{{\bf h}l}\neq g^{\rm SH}_{{\bf h}l} A_{{\bf h}l}]. The combination of the analytical absorption correction with spherical harmonics allows the effect of absorption to be corrected by an accurate analytical model, while still enabling the spherical harmonics model to correct for any residual effects.

3. Results

In crystallography, various metrics, such as R factors (Weiss & Hilgenfeld, 1997[Weiss, M. S. & Hilgenfeld, R. (1997). J. Appl. Cryst. 30, 203-205.]; Diederichs & Karplus, 1997[Diederichs, K. & Karplus, P. A. (1997). Nat. Struct. Mol. Biol. 4, 269-275.]; Weiss, 2001[Weiss, M. S. (2001). J. Appl. Cryst. 34, 130-135.]), correlation coefficients (Karplus & Diederichs, 2012[Karplus, P. A. & Diederichs, K. (2012). Science, 336, 1030-1033.]) and signal-to-noise ratios, are used to evaluate data quality. Additionally, for long-wavelength crystallography peak heights in the phased anomalous difference Fourier maps are important quality indicators (Yang et al., 2003[Yang, C., Pflugrath, J. W., Courville, D. A., Stence, C. N. & Ferrara, J. D. (2003). Acta Cryst. D59, 1943-1957.]). These metrics are used in combination with the success of experimental phasing by SAD to assess the three different absorption correction strategies and compare them with scaling without absorption correction.

Merging and refinement statistics (based on three data sets for OmpK36 and 22 for Cld) are presented in Table 2[link]. As expected, for both samples, all four strategies result in similar resolution ranges, completeness and number of unique reflections. All three approaches to deal with absorption unsurprisingly lead to significant improvements in data quality over the data without correction.

Table 2
Merging and refinement statistics from OmpK36 and Cld

Columns represent the four absorption correction methods: spherical harmonics correction (SH), analytical absorption correction (AC), analytical absorption correction combined with spherical harmonics correction (ACSH), no absorption correction (NO). Values in parentheses are for the outer resolution shell. Further refinement statistics can be found in the supporting information (Tables S1 and S2). For the calculation of the anomalous slope, the resolution range is restricted to resolutions below which the anomalous signal is significant in the ACSH processed data, which is 3.9 Å for OmpK36 and the full resolution range for Cld.

  NO SH AC ACSH
OmpK36 (λ = 3.54 Å)
Merging statistics        
Resolution range (Å) 107.4–2.34 (2.424–2.34) 107.4–2.34 (2.424–2.34) 107.4–2.34 (2.424–2.34) 107.4–2.34 (2.424–2.34)
Multiplicity 10.8 (5.5) 11.0 (5.5) 11.0 (5.5) 11.1 (5.5)
Completeness (%) 98.77 (91.67) 98.85 (92.15) 98.85 (92.12) 98.86 (92.15)
Mean I/σ(I) 11.99 (1.03) 16.42 (1.58) 21.37 (2.00) 24.92 (2.66)
Rmerge 0.139 (0.473) 0.119 (0.419) 0.119 (0.458) 0.105 (0.427)
Rmeas 0.146 (0.525) 0.125 (0.462) 0.125 (0.506) 0.110 (0.472)
Rpim 0.043 (0.214) 0.035 (0.185) 0.035 (0.204) 0.031 (0.191)
CC1/2 0.996 (0.814) 0.997 (0.896) 0.997 (0.874) 0.998 (0.878)
CC* 0.999 (0.947) 0.999 (0.972) 0.999 (0.966) 0.999 (0.967)
Anomalous slope (d ≤3.9 Å) 1.13 1.31 1.69 1.91
Total reflections 654312 (31265) 668732 (31264) 668892 (31264) 672491 (31264)
Unique reflections 60652 (5606) 60652 (5634) 60652 (5633) 60652 (5634)
Refinement statistics        
Work set reflections 60585 (5605) 60631 (5634) 60630 (5632) 60639 (5634)
Free set reflections 3258 (328) 3260 (328) 3260 (328) 3260 (328)
Rwork 0.219 (0.390) 0.207 (0.338) 0.203 (0.332) 0.199 (0.294)
Rfree 0.255 (0.386) 0.244 (0.335) 0.240 (0.335) 0.235 (0.303)
PDB code 8qur 8quq 8qvv 8qvs
         
Cld (λ = 4.13 Å)
Merging statistics        
Resolution range (Å) 46.67–2.7 (2.797–2.7) 46.67–2.7 (2.797–2.7) 46.67–2.7 (2.797–2.7) 46.67–2.7 (2.797–2.7)
Multiplicity 38.8 (23.5) 40.3 (23.5) 41.1 (23.5) 41.1 (23.5)
Completeness (%) 99.43 (97.97) 99.43 (97.97) 99.43 (97.97) 99.43 (97.97)
Mean I/σ(I) 16.51 (4.83) 20.22 (6.61) 37.43 (13.47) 44.73 (15.68)
Rmerge 0.205 (0.281) 0.163 (0.240) 0.112 (0.197) 0.095 (0.183)
Rmeas 0.208 (0.287) 0.165 (0.245) 0.113 (0.201) 0.096 (0.187)
Rpim 0.033 (0.056) 0.025 (0.048) 0.017 (0.039) 0.014 (0.037)
CC1/2 0.997 (0.986) 0.997 (0.99) 0.999 (0.992) 0.999 (0.993)
CC* 0.999 (0.996) 0.999 (0.998) 1 (0.998) 1 (0.998)
Anomalous slope 1.28 1.36 2.48 2.50
Total reflections 531035 (31693) 551553 (31730) 562964 (31747) 563200 (31739)
Unique reflections 13696 (1351) 13696 (1351) 13696 (1351) 13696 (1351)
Refinement statistics        
Work set reflections 13696 (1351) 13696 (1351) 13696 (1351) 13696 (1351)
Free set reflections 686 (76) 686 (76) 686 (76) 686 (76)
Rwork 0.191 (0.240) 0.176 (0.223) 0.172 (0.210) 0.172 (0.209)
Rfree 0.234 (0.297) 0.223 (0.285) 0.218 (0.271) 0.218 (0.273)
PDB code 8quv 8quu 8quz 8qvb

For OmpK36, the analytical absorption correction (AC) gives equivalent merging R factors to spherical harmonics correction (SH), with an overall Rmerge of 0.119 for both. Notably, the AC strategy leads to an increase in the mean I/σ(I), from 16.42 (SH) to 21.37 (AC), and a stronger anomalous signal, as measured by the anomalous slope (1.69 with AC, as opposed to 1.31 with SH). The anomalous slope (Evans, 2006[Evans, P. (2006). Acta Cryst. D62, 72-82.]) is the slope of the central region of a normal probability plot of anomalous differences: a slope greater than one indicates that the anomalous differences are larger than their uncertainties in aggregate. The combination of AC and SH corrections (ACSH) gives further improvements in the merging R factors, signal-to-noise ratio and anomalous signal, with the Rmerge decreasing to 0.105, the mean I/σ(I) increasing to 24.92 and the anomalous slope increasing to 1.91. In Fig. 4[link](a), the anomalous peak heights from sulfur atoms for the three correction strategies are compared with no absorption correction for OmpK36. In total 12 sulfur atoms are found, from two methionine residues and two sulfates in the trimeric structure. A significant increase in peak heights is observed with all three absorption correction methods. AC generally gives better results than SH, with the exception of the heights of MET310 in chain B and SO4-1 in chain C, which are larger in the SH data. Overall, the ACSH strategy brings further improvements in peak heights, except for the weakest anomalous peak, SO4-2, where AC and ACSH perform similarly. Detailed information on the anomalous peaks of OmpK36 can be found in Tables S3–S6 in the supporting information. The refinement statistics for all strategies follow a similar trend to the merging statistics, with R factors being the lowest for ACSH. SAD phasing was performed as a further test of the efficacy of analytical absorption corrections. Phasing was attempted with one, two out of three and all three data sets available. The results, summarized in Table 3[link], show that the ACSH strategy outperforms the others in requiring only two data sets for successful phasing despite the overall completeness of 89.2% and multiplicity of 8.3. Both AC and SH need all three data sets (98.9% overall completeness, multiplicity of 11.0), while the NO strategy is unsuccessful. The numbers of correct residues automatically built into the experimental maps are identical between the three successful strategies, indicating that the quality of the maps is of similar standard and the lower data completeness used for the ACSH approach has no impact.

Table 3
SAD phasing results for OmpK36 (top) and Cld (bottom): statistics from Crank2 for all four absorption correction strategies

While for some strategies only two data sets were needed for successful phasing, the statistics from using three data sets are presented for comparison.

Strategy No. of data sets required for phasing Completeness (overall/high-resolution bin) Multiplicity (overall/high-resolution bin) Refinement R factor/Rfree No. of correct residues automatically built/total No. of residues
OmpK36
NO
SH 3 98.8/88.1 11.0/3.9 0.235/0.280 1041/1041
AC 3 98.8/88.1 11.0/3.9 0.227/0.274 1041/1041
ACSH 2 89.2/71.9 8.3/3.3 0.228/0.280 1041/1041
ACSH 3 98.8/88.1 11.0/3.9 0.218/0.257 1041/1041
           
Cld
NO
SH 3 94.7/82.2 5.8/2.5 0.259/0.336 354/376
AC 2 83.3/64.9 4.4/2.2 0.266/0.348 354/376
AC 3 94.7/82.2 5.9/2.5 0.260/0.320 362/376
ACSH 2 83.3/64.9 4.4/2.2 0.260/0.338 354/376
ACSH 3 94.7/82.2 5.9/2.5 0.259/0.302 362/376
[Figure 4]
Figure 4
Peak heights (>5σ) in the anomalous difference Fourier maps of anomalous scatterers in OmpK36 (a) and Cld (b) plotted in descending order of peak heights in the ACSH data, generated by Anode (Thorn & Sheldrick, 2011[Thorn, A. & Sheldrick, G. M. (2011). J. Appl. Cryst. 44, 1285-1287.]). Raw data are presented in the supporting information, Tables S3 to S6 (OmpK36) and S7 to S10 (Cld).

For Cld, the merging R factors, I/σ(I) and anomalous slopes are noticeably better for AC compared with SH. All merging statistics show further improvement for the combined ACSH correction. In contrast to OmpK36, where data quality indicators changed little between the SH and AC strategies, for Cld, the analytical absorption correction strategy (AC) gives substantially better data statistics compared with SH. For instance, in terms of the merging R factors, we observe a decrease of the Rmerge from 0.163 with SH to 0.112 with AC and a further decrease to 0.095 with the ACSH treatment. There is also an increase in the overall mean I/σ(I) from 20.22 for SH to 44.73 for the ACSH strategy with the high-resolution shell I/σ(I) following this trend. The anomalous slope value increases from 1.36 with SH to 2.48 and 2.5 for AC and ACSH, respectively. This indicates an impressive improvement in the anomalous signal as a result of applying analytical absorption corrections.

The anomalous peak heights for the different absorption correction strategies for Cld are shown in Fig. 4[link](b). In addition to three methionines and one cysteine per polypeptide chain, each Cld monomer also binds an Fe-containing heme ligand and a Cl anion. A single SO42− anion could be identified for the dimer, bringing the total number of anomalous scatterers to 13. SH leads to higher anomalous peak heights compared with no absorption correction. In line with the improved merging statistics, the anomalous signal in AC and ACSH is stronger than that in SH. ACSH gives the highest anomalous peak heights overall. While for OmpK36 the improvements in peak heights given by the AC and ACSH strategies over SH are quite modest, for Cld the increase from SH to AC/ACSH is more substantial. For the largest peaks, MET99 and CYS132, we observe increases in peak heights from 14 to 17 and 18σ for AC and ACSH, respectively. Further details of anomalous peak heights for Cld may be found in the supporting information, Tables S7–S10. The experimental phasing results for this sample (presented in Table 3[link]) show that the AC and ACSH strategies perform very similarly, with a successful phasing outcome requiring only two out of 22 data sets, with an overall completeness of 83.3% and overall multiplicity of 4.4. For the SH strategy, three data sets are needed, with a higher overall completeness of 94.7% and multiplicity of 5.8. These results follow the same pattern seen with the data quality indicators discussed above, where the AC strategy outperforms the SH approach. Experimental phasing is unsuccessful for the Cld data with no absorption corrections, even after merging all 22 data sets.

To illustrate the extent of the AC and SH corrections, histograms of the per-reflection analytical absorption correction factors ([A_{{\bf h}l}]) and spherical harmonics correction terms ([S_{{\bf h}l}]) are presented in Fig. 5[link] for OmpK36 and Cld. For both data sets, when employing the SH correction strategy, the resulting spherical harmonics terms ([S_{{\bf h}l}]) are distributed over a large range (0.5–1.5). When employing the ACSH strategy, the inclusion of the absorption correction factors ([A_{{\bf h}l}]) (shown on the right of Fig. 5[link]) leads to unimodal [S_{{\bf h}l}] distributions over a narrower range (0.7–1.3) centred around 1. As the `no correction value' for the SH model is [S_{{\bf h}l}] = 1.0, fitting the additional spherical harmonics terms in the ACSH strategy results in further improvement in the internal consistency compared with AC alone, allowing correction for additional systematic effects present in the data.

[Figure 5]
Figure 5
Histograms of absorption factors [A_{{\bf h}l}] and spherical harmonics terms [S_{{\bf h}l}] for OmpK36 (a) and Cld (b). [A_{{\bf h}l}] (green) as used in AC and ACSH strategies are on an absolute scale, whereas [S_{{\bf h}l}] for SH (orange) and ACSH (purple) are on a relative scale.

4. Discussion and conclusion

In this study we demonstrate the successful application of analytical absorption corrections based on 3D reconstructions from X-ray tomography implemented in AnACor. We describe the algorithm for calculating the path lengths from 3D models by a ray-tracing method. Two very long wavelength experiments from crystals of the proteins OmpK36 and Cld indicate that this approach substantially improves data quality and the success of experimental phasing compared with the standard scaling protocol based on spherical harmonics. Scaling without any absorption correction is presented as a control and unsurprisingly yields the poorest data quality statistics and anomalous peak heights, and for both samples experimental phasing is unsuccessful. This clearly indicates that data quality is severely affected by absorption effects, demonstrating the need for absorption corrections.

Data from OmpK36, which crystallizes in the monoclinic space group C2, were collected at a wavelength of λ = 3.54 Å. A clear trend is visible: the analytical absorption correction (AC) is better than the spherical harmonics correction (SH) and the combination of the two (ACSH) improves the data even further. While the overall improvements on statistics are small, the fact that the OmpK36 structure could be solved after ACSH correction using only 2/3 of the data needed for the AC and SH strategies clearly highlights the importance of such an improvement. For the Cld data (P1, λ = 4.13 Å) the same trend is observed. However, while the difference between AC and ACSH is small, they outperform the spherical harmonics correction. This is in particular reflected in the outcome from experimental phasing, where two data sets are sufficient for both AC and ACSH, while three data sets are needed to solve the structure from data corrected by SH. In general, the combined approach of ACSH gives the best results for both samples/wavelengths, as it can model additional systematic effects present in the experimental data.

X-ray absorption increases with the cube of the wavelength, so a change from λ = 1.0 Å to λ = 4.13 Å leads to a 70-fold increase in absorption coefficients. The analytical absorption correction compensates for this increase, reflected in the narrow unimodal distribution of the resulting spherical harmonics terms [S_{{\bf h}l}] centred around 1.0 in the two ACSH cases. Both samples used in this study crystallize in either monoclinic (OmpK36) or triclinic (Cld) space groups. This in combination with the asymmetry of the cylindrical P12M detector, with an aspect ratio of 2:1, leads to a low overall data multiplicity of five for OmpK36 and only three in the case of Cld, as well as poor data completeness for a single 360° data set. In contrast to the spherical harmonics, the analytical absorption correction is not dependent on multiple observations, and hence is ideally suited for crystals in low-symmetry space groups or for radiation-sensitive crystals at long wavelengths.

AnACor is able to correct data in multiple crystal orientations and for cases where the beam is smaller than the sample. Future work will allow the use of experimentally determined beam profiles and increase the efficiency and speed of the software. Currently, the bottleneck is the manual segmentation step to create the 3D models. The increased phase contrast at long wavelengths and limitations with the current beamline hardware, in particular the sphere of confusion of the goniometer, lead to blurred boundaries in the tomographic reconstructions. The resulting inaccuracies in the segmented 3D model can affect both the path length and the absorption coefficient calculations. The next stage of this work is therefore to understand, quantify and reduce these errors impacting the 3D model. Analytical absorption corrections are beneficial not only for long-wavelength macromolecular crystallography but also for highly absorbing samples in chemical crystallography. In this work the segmented 3D model is obtained by X-ray tomography on beamline I23 at Diamond Light Source. However, AnACor can also be used for analytical absorption corrections for data from other sources, as long as a file with annotated voxels is provided and the relation between the coordinate systems of the 3D model and the diffraction experiment is known.

Footnotes

Joint first authors

Acknowledgements

The authors acknowledge the use of the University of Oxford Advanced Research Computing (ARC) facility in carrying out this work (https://dx.doi.org/10.5281/zenodo.22558).

Funding information

AMO and JJAGK were supported by Diamond Light Source and the UK Science and Technology Facilities Council (STFC). AMO acknowledges the Biotechnology and Biological Sciences Research Council, and is the recipient of a Wellcome Investigator Award 210734/Z/18/Z and a Royal Society Wolfson Fellowship RSWF\R2\182017.

References

First citationAlbrecht, G. (1939). Rev. Sci. Instrum. 10, 221–222.  CrossRef Google Scholar
First citationAngel, R. J. (2004). J. Appl. Cryst. 37, 486–492.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationArndt, U. W. (1984). J. Appl. Cryst. 17, 118–119.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationBeilsten-Edmands, J., Winter, G., Gildea, R., Parkhurst, J., Waterman, D. & Evans, G. (2020). Acta Cryst. D76, 385–399.  Web of Science CrossRef IUCr Journals Google Scholar
First citationBlessing, R. H. (1995). Acta Cryst. A51, 33–38.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationBrockhauser, S., Di Michiel, M., McGeehan, J. E., McCarthy, A. A. & Ravelli, R. B. G. (2008). J. Appl. Cryst. 41, 1057–1066.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBruker (2012). APEX. Bruker AXS Inc., Madison, Wisconsin, USA.  Google Scholar
First citationBusing, W. R. & Levy, H. A. (1957). Acta Cryst. 10, 180–182.  CrossRef CAS IUCr Journals Web of Science Google Scholar
First citationClark, R. C. & Reid, J. S. (1995). Acta Cryst. A51, 887–897.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationDeTitta, G. T. (1985). J. Appl. Cryst. 18, 75–79.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationDiederichs, K. & Karplus, P. A. (1997). Nat. Struct. Mol. Biol. 4, 269–275.  CrossRef CAS Web of Science Google Scholar
First citationDuman, R., Orr, C. M., Mykhaylyk, V., El Omari, K., Pocock, R., Grama, V. & Wagner, A. (2021). J. Vis. Exp. 170, e62364.  Google Scholar
First citationEl Omari, K., Duman, R., Mykhaylyk, V., Orr, C. M., Latimer-Smith, M., Winter, G., Grama, V., Qu, F., Bountra, K., Kwong, H. S., Romano, M., Reis, R. I., Vogeley, L., Vecchia, L., Owen, C. D., Wittmann, S., Renner, M., Senda, M., Matsugaki, N., Kawano, Y., Bowden, T. A., Moraes, I., Grimes, J. M., Mancini, E. J., Walsh, M. A., Guzzo, C. R., Owens, R. J., Jones, E. Y., Brown, D. G., Stuart, D. I., Beis, K. & Wagner, A. (2023). Commun. Chem. 6, 219.  CrossRef PubMed Google Scholar
First citationEvans, P. (2006). Acta Cryst. D62, 72–82.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationEvans, P. R. & Murshudov, G. N. (2013). Acta Cryst. D69, 1204–1214.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationGürsoy, D., De Carlo, F., Xiao, X. & Jacobsen, C. (2014). J. Synchrotron Rad. 21, 1188–1193.  Web of Science CrossRef IUCr Journals Google Scholar
First citationHarris, C. R., Millman, K. J., van der Walt, S. J., Gommers, R., Virtanen, P., Cournapeau, D., Wieser, E., Taylor, J., Berg, S., Smith, N. J., Kern, R., Picus, M., Hoyer, S., van Kerkwijk, M. H., Brett, M., Haldane, A., del Río, J. F., Wiebe, M., Peterson, P., Gérard-Marchant, P., Sheppard, K., Reddy, T., Weckesser, W., Abbasi, H., Gohlke, C. & Oliphant, T. E. (2020). Nature, 585, 357–362.  Web of Science CrossRef CAS PubMed Google Scholar
First citationHowells, R. G. (1950). Acta Cryst. 3, 366–369.  CrossRef IUCr Journals Google Scholar
First citationKabsch, W. (2010). Acta Cryst. D66, 125–132.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationKarplus, P. A. & Diederichs, K. (2012). Science, 336, 1030–1033.  Web of Science CrossRef CAS PubMed Google Scholar
First citationKatayama, C., Sakabe, N. & Sakabe, K. (1972). Acta Cryst. A28, 293–295.  CrossRef IUCr Journals Google Scholar
First citationKazantsev, D., Duman, R., Wagner, A., Mykhaylyk, V., Wanelik, K., Basham, M. & Wadeson, N. (2021). J. Synchrotron Rad. 28, 889–901.  CrossRef CAS IUCr Journals Google Scholar
First citationKazantsev, D., Wadeson, N. & Basham, M. (2022). SoftwareX, 19, 101157.  Google Scholar
First citationKopfmann, G. & Huber, R. (1968). Acta Cryst. A24, 348–351.  CrossRef IUCr Journals Web of Science Google Scholar
First citationLam, S. K., Pitrou, A. & Seibert, S. (2015). Proceedings of the Second Workshop on the LLVM Compiler Infrastructure in HPC, pp. 1–6. Association for Computing Machinery.  Google Scholar
First citationLeal, R. M. F., Teixeira, S. C. M., Rey, V., Forsyth, V. T. & Mitchell, E. P. (2008). J. Appl. Cryst. 41, 729–737.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMaslen, E. N. (2004). International Tables for Crystallography, 3rd ed., edited by E. Prince, Vol. C, ch. 6.3.3, pp. 600–608. Dordrecht: Kluwer.  Google Scholar
First citationMerrifield, D. R., Ramachandran, V., Roberts, K. J., Armour, W., Axford, D., Basham, M., Connolley, T., Evans, G., McAuley, K. E., Owen, R. L. & Sandy, J. (2011). Meas. Sci. Technol. 22, 115703.  CrossRef Google Scholar
First citationMeulenaer, J. de & Tompa, H. (1965). Acta Cryst. 19, 1014–1018.  CrossRef IUCr Journals Web of Science Google Scholar
First citationMinor, W., Cymborowski, M., Otwinowski, Z. & Chruszcz, M. (2006). Acta Cryst. D62, 859–866.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMonaco, H. L. & Artioli, G. (2002). Fundamentals of Crystallography, 2nd ed., edited by H. Giacovazzo, ch. 5, pp. 376–388. Oxford University Press.  Google Scholar
First citationMurshudov, G. N., Vagin, A. A. & Dodson, E. J. (1997). Acta Cryst. D53, 240–255.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationNorth, A. C. T., Phillips, D. C. & Mathews, F. S. (1968). Acta Cryst. A24, 351–359.  CrossRef IUCr Journals Web of Science Google Scholar
First citationSchaffner, I., Mlynek, G., Flego, N., Pühringer, D., Libiseller-Egger, J., Coates, L., Hofbauer, S., Bellei, M., Furtmüller, P. G., Battistuzzi, G., Smulevich, G., Djinović-Carugo, K. & Obinger, C. (2017). ACS Catal. 7, 7962–7976.  Web of Science CrossRef CAS PubMed Google Scholar
First citationSheldrick, G. M. (1996). SADABS. University of Göttingen, Germany.  Google Scholar
First citationSkubák, P. & Pannu, N. S. (2013). Nat. Commun. 4, 2777.  Web of Science PubMed Google Scholar
First citationStrutz, T. (2011). IEEE/ACM Trans. Comput. Biol. Bioinf. 8, 797–807.  Web of Science CrossRef Google Scholar
First citationThorn, A. & Sheldrick, G. M. (2011). J. Appl. Cryst. 44, 1285–1287.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationVo, N. T., Atwood, R. C. & Drakopoulos, M. (2018). Opt. Express, 26, 28396–28412.  Web of Science CrossRef PubMed Google Scholar
First citationWagner, A., Duman, R., Henderson, K. & Mykhaylyk, V. (2016). Acta Cryst. D72, 430–439.  Web of Science CrossRef IUCr Journals Google Scholar
First citationWalker, N. & Stuart, D. (1983). Acta Cryst. A39, 158–166.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationWarren, A. J., Armour, W., Axford, D., Basham, M., Connolley, T., Hall, D. R., Horrell, S., McAuley, K. E., Mykhaylyk, V., Wagner, A. & Evans, G. (2013). Acta Cryst. D69, 1252–1259.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWeiss, M. S. (2001). J. Appl. Cryst. 34, 130–135.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWeiss, M. S. & Hilgenfeld, R. (1997). J. Appl. Cryst. 30, 203–205.  CrossRef CAS Web of Science IUCr Journals Google Scholar
First citationWinter, G., Beilsten–Edmands, J., Devenish, N., Gerstel, M., Gildea, R. J., McDonagh, D., Pascal, E., Waterman, D. G., Williams, B. H. & Evans, G. (2022). Protein Sci. 31, 232–250.  Web of Science CrossRef CAS PubMed Google Scholar
First citationWinter, G., Waterman, D. G., Parkhurst, J. M., Brewster, A. S., Gildea, R. J., Gerstel, M., Fuentes-Montero, L., Vollmar, M., Michels-Clark, T., Young, I. D., Sauter, N. K. & Evans, G. (2018). Acta Cryst. D74, 85–97.  Web of Science CrossRef IUCr Journals Google Scholar
First citationWong, J. L., Romano, M., Kerry, L. E., Kwong, H.-S., Low, W.-W., Brett, S. J., Clements, A., Beis, K. & Frankel, G. (2019). Nat. Commun. 10, 1–10.  PubMed Google Scholar
First citationYang, C., Pflugrath, J. W., Courville, D. A., Stence, C. N. & Ferrara, J. D. (2003). Acta Cryst. D59, 1943–1957.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationZeldin, O. B., Gerstel, M. & Garman, E. F. (2013). J. Appl. Cryst. 46, 1225–1230.  Web of Science CrossRef CAS IUCr Journals Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.