Herein, the successful use of a focused ion beam/scanning electron microscope to prepare microsamples of radioactive single crystals for X-ray diffraction analysis is reported. This technique was used to extract and analyze crystalline Pu-containing particles as small as 28 µm3 from Hanford soil taken from the 216-Z-9 waste crib, which were then crystallographically characterized using single-crystal X-ray diffraction to confirm the cubic structure of PuO2. As a systematic proof of concept, the technique was first tested using UO2 crystals milled into cubic shapes with approximate volumes of 4620, 1331, 125, 8 and 1 µm3, in order to empirically determine the crystal size limits for characterization by a laboratory-based diffractometer with a sealed tube Mo or Ag anode X-ray source and a charge-coupled device detector.