MXCuBE2: the dawn of MXCuBE Collaboration

CROSSMARK_Color_square_no_text.svg

aESRF - The European Synchrotron, 71 Avenue des Martyrs, 38000 Grenoble, France, bTXO Pedro de Alejandria 2-2C Pamplona, Spain, cSynchrotron SOLEIL, 91192 Gif-sur-Yvette Cedex, France, dMAX IV Laboratory, Lund University, SE-221 00 Lund, Sweden, eHelmholtz-Zentrum Berlin für Materialien und Energie, Macromolecular Crystallography, Albert-Einstein-Straße 15, D-12489 Berlin, Germany, fEuropean Molecular Biology Laboratory, Grenoble Outstation, 71 Avenue des Martyrs, 38042 Grenoble, France, gHamburg Unit c/o DESY, European Molecular Biology Laboratory (EMBL), Notkestrasse 85, 22603 Hamburg, Germany, hCELLS-ALBA Synchrotron Light Source, 08290 Cerdanyola del Vallès, Spain, and iGlobal Phasing Ltd, Sheraton House, Castle Park, Cambridge CB3 0AK, UK
*Correspondence e-mail: daniele.de_sanctis@esrf.fr

Edited by M. Yamamoto, RIKEN SPring-8 Center, Japan (Received 21 September 2018; accepted 23 January 2019; online 22 February 2019)

MXCuBE2 is the second-generation evolution of the MXCuBE beamline control software, initially developed and used at ESRF – the European Synchrotron. MXCuBE2 extends, in an intuitive graphical user interface (GUI), the functionalities and data collection methods available to users while keeping all previously available features and allowing for the straightforward incorporation of ongoing and future developments. MXCuBE2 introduces an extended abstraction layer that allows easy interfacing of any kind of macromolecular crystallography (MX) hardware component, whether this is a diffractometer, sample changer, detector or optical element. MXCuBE2 also works in strong synergy with the ISPyB Laboratory Information Management System, accessing the list of samples available for a particular experimental session and associating, either from instructions contained in ISPyB or from user input via the MXCuBE2 GUI, different data collection types to them. The development of MXCuBE2 forms the core of a fruitful collaboration which brings together several European synchrotrons and a software development factory and, as such, defines a new paradigm for the development of beamline control platforms for the European MX user community.

1. Introduction

The past 15 years or so have seen a dramatic increase in the productivity of synchrotron macromolecular crystallography (MX) beamlines, both in terms of the number of crystals analysed and the number of new crystal structures deposited in the Protein Data Bank (PDB) (Berman et al., 2000[Berman, H. M., Westbrook, J., Feng, Z., Gilliland, G., Bhat, T. N., Weissig, H., Shindyalov, I. N. & Bourne, P. E. (2000). Nucleic Acids Res. 28, 235-242.]). This evolution has, to a large extent, been driven by improvements in beamline hardware and software. In particular, the smaller X-ray beams provided by third-generation synchrotron sources have opened new avenues in MX, making it possible to collect diffraction data, using increasingly advanced experimental protocols, from smaller crystals of increasingly intricate and important structural biology targets. However, in order that experimenters can take best advantage of the increasingly complex data collection options available, a robust, user-friendly and adaptable beamline control graphical user interface (GUI) is essential.

In 2006, the ESRF and the Medical Research Council UK (MRC-UK) started to develop the MXCuBE beamline control GUI (Gabadinho et al., 2010[Gabadinho, J., Beteva, A., Guijarro, M., Rey-Bakaikoa, V., Spruce, D., Bowler, M. W., Brockhauser, S., Flot, D., Gordon, E. J., Hall, D. R., Lavault, B., McCarthy, A. A., McCarthy, J., Mitchell, E., Monaco, S., Mueller-Dieckmann, C., Nurizzo, D., Ravelli, R. B. G., Thibault, X., Walsh, M. A., Leonard, G. A. & McSweeney, S. M. (2010). J. Synchrotron Rad. 17, 700-707.]) with the aim of providing the users of its various MX facilities with an almost identical user-control environment regardless of any differences in beamline hardware. MXCuBE was initially deployed both at the ESRF MX beamlines and at the Collaborative Research Group (CRG) beamline BM14. This was quickly followed by installations on the MAX-lab beamline I911-3 and the Helmholtz-Zentrum Berlin (HZB, BESSY II) beamline BL14.1 (Mueller et al., 2012[Mueller, U., Darowski, N., Fuchs, M. R., Förster, R., Hellmig, M., Paithankar, K. S., Pühringer, S., Steffien, M., Zocher, G. & Weiss, M. S. (2012). J. Synchrotron Rad. 19, 442-449.]). However, installation at sites other than the ESRF was not always straightforward due to differences in the low-level beamline server and control software installed. Discussions concerning the development of the second-generation MXCuBE2, designed to be independent of both beamline hardware and lower-level server and control software, were thus instigated and the MXCuBE Collaboration was formed (§2). The MXCuBE2 GUI resulting from the collaboration was also simplified, permitting a closer interaction with the sample under study while, at the same time, including a growing number of new MX experimental methodologies. In particular and by design, MXCuBE2 is able to interact with external `experiment descriptors' (Brockhauser et al., 2012[Brockhauser, S., Svensson, O., Bowler, M. W., Nanao, M., Gordon, E., Leal, R. M. F., Popov, A., Gerring, M., McCarthy, A. A. & Gotz, A. (2012). Acta Cryst. D68, 975-984.]) which can iteratively define and adjust data collection parameters according to results obtained by connected analyses. Although there is no reciprocal dependency, user experience is greatly enhanced by an increased synergy between MXCuBE2 and the ISPyB Laboratory Information Management System (LIMS) (Delagenière et al., 2011[Delagenière, S., Brenchereau, P., Launer, L., Ashton, A. W., Leal, R., Veyrier, S., Gabadinho, J., Gordon, E. J., Jones, S. D., Levik, K. E., McSweeney, S. M., Monaco, S., Nanao, M., Spruce, D., Svensson, O., Walsh, M. A. & Leonard, G. A. (2011). Bioinformatics, 27, 3186-3192.]). To better interact with MXCuBE2, ISPyB has been expanded to include fuller sample descriptors and `diffraction plans' and now stores all the details of all the experiments (including the results of on-line data analyses) performed on a given sample.

The major aim of the MXCuBE Collaboration is the pooling of resources to provide a standardized and user-friendly beamline control GUI at the synchrotron sites involved and to provide a platform for the rapid implementation, at all sites where these are desired, of new functionality and experimental protocols. In this context, we present here details of the collaboration, of MXCuBE2 itself and of the contribution of the different partners making up the consortium.

2. The MXCuBE Collaboration

Experience gained at ESRF, first with ProDC (Arzt et al., 2005[Arzt, S., Beteva, A., Cipriani, F., Delageniere, S., Felisaz, F., Förstner, G., Gordon, E., Launer, L., Lavault, B., Leonard, G., Mairs, T., McCarthy, A., McCarthy, J., McSweeney, S., Meyer, J., Mitchell, E., Monaco, S., Nurizzo, D., Ravelli, R., Rey, V., Shepard, W., Spruce, D., Svensson, O. & Theveneau, P. (2005). Prog. Biophys. Mol. Biol. 89, 124-152.]), then with initial versions of MXCuBE (Gabadinho et al., 2010[Gabadinho, J., Beteva, A., Guijarro, M., Rey-Bakaikoa, V., Spruce, D., Bowler, M. W., Brockhauser, S., Flot, D., Gordon, E. J., Hall, D. R., Lavault, B., McCarthy, A. A., McCarthy, J., Mitchell, E., Monaco, S., Mueller-Dieckmann, C., Nurizzo, D., Ravelli, R. B. G., Thibault, X., Walsh, M. A., Leonard, G. A. & McSweeney, S. M. (2010). J. Synchrotron Rad. 17, 700-707.]), showed that the development, maintenance and continuous upgrade of GUIs for the control of experiments at synchrotron-based MX beamlines is an extremely labour-intensive endeavour requiring significant human resources from beamline scientists and software engineers. This is particularly the case when the beamline control GUI is linked to synchrotron-based LIMS such as PXWeb (Arzt et al., 2005[Arzt, S., Beteva, A., Cipriani, F., Delageniere, S., Felisaz, F., Förstner, G., Gordon, E., Launer, L., Lavault, B., Leonard, G., Mairs, T., McCarthy, A., McCarthy, J., McSweeney, S., Meyer, J., Mitchell, E., Monaco, S., Nurizzo, D., Ravelli, R., Rey, V., Shepard, W., Spruce, D., Svensson, O. & Theveneau, P. (2005). Prog. Biophys. Mol. Biol. 89, 124-152.]) or ISPyB (Delagenière et al., 2011[Delagenière, S., Brenchereau, P., Launer, L., Ashton, A. W., Leal, R., Veyrier, S., Gabadinho, J., Gordon, E. J., Jones, S. D., Levik, K. E., McSweeney, S. M., Monaco, S., Nanao, M., Spruce, D., Svensson, O., Walsh, M. A. & Leonard, G. A. (2011). Bioinformatics, 27, 3186-3192.]).

In Grenoble, the ProDC/PXWeb combination for synchrotron beamline control/experiment-tracking was developed by the ESRF–EMBL Joint Structural Biology Group (JSBG) while initial versions of the MXCuBE/ISPyB combination were developed in collaboration with the MRC-UK operated BM14 CRG. The JSBG–BM14 link-up had two aims: in tandem with funding from the BIOXHIT [Biocrystallography (X) on a Highly Integrated Technology Platform for European Structural Genomics] project, the collaboration helped ensure the availability of sufficient human resources to develop, maintain and upgrade MXCuBE/ISPyB and, at the same time, provided a standard beamline control GUI for experiments at eight ESRF-based MX beamlines. The MRC-UK BM14 CRG ceased operating in 2011. In addition to the ESRF, MXCuBE was also installed on MX beamlines at the MAX-II, HZB (BESSY II), EMBL-Hamburg and SOLEIL synchrotron sites who, along with the protein crystallography software company Global Phasing Ltd, continued to provide ideas and human resources for its development. This informal collaboration was officialized by the signing in 2012 of a Memorandum of Understanding (MoU), the aims of which were to define the organization of a collaboration, to oversee the installation, support, documentation and further development of MXCuBE at several European Synchrotron sites, to present a similar and familiar interface for MX experiments at all of the synchrotron sites involved, and to share the human resources costs of the development, support and maintenance of the software. The collaboration has continued to expand and the partners now comprise ESRF, EMBL, Synchrotron SOLEIL, HZB (BESSY II), ALBA, DESY, MAX IV Laboratory and Global Phasing Ltd, while Elettra and LNLS are in the process of joining the collaboration.

Management of the collaboration is handled by three committees: the MXCuBE Steering Committee (STC), the MXCuBE Scientific Committee (SC) and the MXCuBE Developers Committee (DC). As a minimum condition for joining the collaboration each partner is required to assign a software engineer who should contribute at least 10% of working time to developments carried on within the collaboration. The role of the STC is to define the long-term strategy of the collaboration ensuring, in particular, adequate manpower and financial resources for the smooth functioning of the collaboration. The SC defines the scientific requirements and development priorities for MXCuBE while members of the DC are charged with implementing software developments requested by the SC. The DC also ensures that at any given time there is only a single official MXCuBE version, available through a publicly accessible website for download and collaboration on source code (https://github.com/mxcube; https://www.mxcube.org). The STC usually convenes only at biannual MXCuBE Collaboration meetings, organized by each of the partners in a round-robin fashion, at which progress is presented and strategic issues discussed. Meetings of the SC and the DC are held more frequently with, outside of the MXCuBE Collaboration meetings, sessions organized on an ad hoc basis as and when necessary. New members of the collaboration may join at any time. All that is required is the agreement of the STC and the subsequent signing by the new partner of the MXCuBE MoU.

The setting up of the MXCuBE Collaboration has provided, to the benefit of the European structural biology community, a stable, productive environment for the development of MXCuBE. Such has been the success of the collaboration that its MoU-based membership and management structure have now been adopted by the ISPyB Collaboration (https://github.com/ispyb) recently set up to oversee the joint development of the ISPyB LIMS.

3. The design of MXCuBE2

MXCuBE2 comprises two main components: a GUI and a data acquisition control layer. The MXCuBE2 GUI, similarly to its predecessor, is built using the Bliss framework (Guijarro et al., 2004[Guijarro, M., Berruyer, G., Klora, J. & Rey-Bakaikoa, V. (2004). Bliss Framework, NOBUGS 5, 18-20 October 2004, PSI, Switzerland.]; https://github.com/mxcube/BlissFramework), a tool developed at the ESRF for building graphical interfaces for beamline instrumentation control and based on Python and the Qt3 toolkit. Porting to more recent versions of Qt (versions 4 and 5, respectively) has been carried out within the MXCuBE Collaboration and has allowed the extension of the lifetime of the MXCuBE2 GUI beyond the official support provided by the Qt3 packages in different Linux distributions. The GUI is built by combining different graphical widgets named bricks. Bricks are individual graphical objects integrated in the GUI. They are connected to hardware objects that directly control and display information such as the position and the status of the beamline equipment (Fig. 1[link]).

[Figure 1]
Figure 1
MXCuBE2 architecture. The GUI gives access to the scientific abstraction layer (data collection methods or external `experiment descriptors') and manages the queue execution, centred positions and communicates with the LIMS. A beamline abstraction layer constituted by the hardware objects ensures the control of the hardware access layer independently from the control system used.

3.1. Beamline abstraction layer – hardware objects

Hardware objects are Python classes directly associated with a configuration eXtensible Markup Language (XML) file which make the link between beamline physical devices and the user interface graphical components. Hardware objects are based on abstract classes and are accessed through a common application programming interface (API) from MXCuBE2. This allows the straightforward adaptation of MXCuBE2 to the different hardware and beamline specificities of different synchrotron sites.

Hardware object XML configuration files can be made available either by a Hardware Repository Server or just read from a directory. The added benefit of the Hardware Repository Server is that it allows shared read and write access to hardware objects. Indeed, when the server modifies XML files it can send notifications to listening instances of MXCuBE2, which can, in turn, reload hardware objects accordingly.

Hardware objects can contain other hardware objects, thus enabling the more complete representation of complex hardware and equipment in a generic manner. The life cycle of Python hardware objects is as follows: at start-up, MXCuBE2 gets the list of hardware objects necessary to create the user interface components. The set of Python hardware objects is then instantiated from the information contained in hardware repository files. Finally, hardware objects are passed to the corresponding user interface components. Python hardware objects are singletons within a running MXCuBE2 instance, meaning that the objects are shared among all user interface components. Only one unique instance of each hardware object lives in a running MXCuBE2 application. When an MXCuBE2 session is terminated, the hardware objects are ended and destroyed.

Multiple MXCuBE2 applications (such as in remote access, §4.5[link]) create their own set of hardware objects. The current hardware objects model in MXCuBE2 assumes concurrent access to beamline devices from different processes and is managed at the beamline control system level.

3.2. The MXCuBE2 queue

The MXCuBE2 queue (Fig. 2[link]) holds the sequence of operations to be executed by a user carrying out an experiment at an MX beamline. One of the core ideas of MXCuBE2 is to represent all data acquisition sequences, including sample loading and unloading operations, as tasks in a tree structure. At the first level of the MXCuBE queue, tree nodes represent samples. At the second level, data collection group nodes contain logically grouped tasks; for example, inverse beam data collection tasks, or other interleaved data collection procedures (e.g. an absorption-edge scan followed by data collection at peak and inflection points etc.). Tasks are the leaf children nodes of the MXCuBE2 queue. In this way, different data collections on a single or multiple samples can be chained in a time-efficient manner. More complex data collection protocols can also be prepared and executed with minimal user intervention. The construction of a queue puts the focus on data collection rather than sample loading and unloading procedures, which is an intrinsic step for data collection on a sample.

[Figure 2]
Figure 2
The GUI representation of the MXCuBE2 queue. Each sample present in the sample changer dewar is represented as a node, while child nodes group data collection on that sample.

The queue is split into three main components: manager, queue entries and model objects. This allows isolating queue execution (QueueManager object), task logic (QueueEntry objects) and task description (QueueModel objects) from each other, ensuring a clean separation of the involvement of the different objects and securing a straightforward serialization of the queue either to disk or via the network (§3.3[link]). In fact, QueueEntry objects are composed of both QueueModel objects and hardware objects. The latter are live Python objects with connections to hardware devices (§3.1[link]) which cannot be self-described with basic data types. Model objects specify the kind of task to be executed and its associated data, whereas queue entries contain model objects (tasks) and the corresponding hardware objects to execute the task described in the model. QueueEntry objects define three main methods for implementing the execution logic for each task: `pre-execute', `execute' and `post-execute' (Fig. 3[link]). For example, a sample queue entry will operate the sample changer device to load the sample in the `pre-execute' method, and will unload the sample in the `post-execute' method. In addition to queue entry objects, the QueueManager is directly associated with the queue user interface component. Its role is to go through the tree of queue entries, and to execute them one by one as configured by the user, using queue controls from the user interface or from a third-party tool such as an `experiment descriptor'.

[Figure 3]
Figure 3
The three phases of the execution of a data collection on a selected sample. Unless already mounted, the pre-execute controls the sample changer. Once the sample is mounted, a centred position is selected, automatically or by the user, and the experiment details are then uploaded to the LIMS. After the data collection, the post-execute triggers the data analysis and updates the LIMS entry with the data collection status.

3.3. XML-RPC server

MXCuBE2 embeds an XML-RPC server allowing access to high-level MXCuBE2 queue methods by external, third party software. This concept is another key feature of MXCuBE2, permitting the creation of new data collection protocols that would be tedious to set up manually. As a result of this organization, an external `experiment descriptor' tool represented as an expert system can be very beneficial. The Remote Procedure Call service provided by the embedded XML-RPC server exposes the queue API to access core components of MXCuBE2, just like a single user would do using the queue control from the user interface. XML-RPC was chosen for its simplicity and its availability in many languages (C, C++, Java, Python and others). Indeed, an XML-RPC client for MXCuBE2 could be written in another language than that of MXCuBE2 itself. The data exchange between XML-RPC clients and the embedded XML-RPC server is performed using remote procedure calls, as specified by the XML-RPC standard and serialized MXCuBE2 QueueModel objects. The model objects are serialized in JSON (JavaScript Object Notation) format. The exposed API allows the creation and manipulation of queue entries and the assignment of these entries' model objects as tasks for execution. The queue manager object API is also exposed in order to be able to perform start and stop actions, and monitor queue execution.

4. GUI description

The MXCuBE2 GUI presents all of the most fundamental operations required during an experiment in a single window, allowing users to mount samples and perform various kinds of data collection in an intuitive way. The GUI is split into four main areas (Fig. 4[link]). The left column contains a login area (requiring site-specific credentials) and the sample list where the contents of the sample changer are permanently displayed and accessible. Users must login to the GUI in order to have access to, and to obtain control of, the sample changer and the sample list content. After logging in, a group identifier facilitates backing up and retrieval of the data by storing all data in a common folder named after the indicated group name. This option is particularly useful in the case of different groups sharing their beam time and proposal number, as often happens at the ESRF when scheduling users from the Block Allocation Group (BAG).

[Figure 4]
Figure 4
Overview of MXCuBE2 (using Qt3 framework). As described in the main text, four main areas constitute the GUI. The currently mounted sample is highlighted (while centred positions are tracked on screen) in green and yellow encircled crosses for active and inactive positions, respectively. The collection methods are organized in vertical tabs, each containing the most relevant data collection parameters.

The central part of the GUI accommodates widgets for controlling the goniometer motors and for the sample visualization microscope view. The collection methods area then presents six main groups: `Standard data collection', `Characterization', `Helical', `Energy scan', `XRF' and `Advanced' tabs can be accessed depending on the requirements of an experiment. The right column of the GUI groups a set of information widgets. The top part of this panel displays the information from the storage ring (such as stored ring current, filling mode, time to next injection), the photon flux at the sample position and the temperature of the cryo-system. Underneath this area, experimental parameters such as incident X-ray energy (or wavelength), resolution (or sample-to-detector distance) and transmission are displayed and their values can be changed in their respective motor or actuator bricks. Finally, widgets allowing the control and monitoring of beam shutters, the control and monitoring of the beam-stop position, and tools for remote access are displayed. A concise log at the bottom of the window displays information messages and beamline events. Three additional (and optional) tabs help users to access functionalities only occasionally used, i.e. the `System log' (containing a more verbose logging of the events), a `Feedback' widget for sending emails to the local staff and a `Chat' function allowing live communication with remotely logged users.

4.1. Sample changer and sample list

The sample list displays all the sample positions available within the sample changer installed on a given beamline. Once the sample changer has been loaded and the positions of the pucks noted then, provided the relevant information has been pre-recorded and stored in ISPyB, clicking the `Synchronize ISPyB' button will fill the sample list with a complete description of the contents (on a crystal level) of the sample changer (Fig. 5[link], left). This information is used to automatically create a folder and subfolder structure (/local/path/to/images/"protein"/"protein"-"sample"/) and diffraction image files with appropriate file names (Fig. 5[link], right). The sample currently mounted is highlighted in the sample list for immediate visualization/retrieval (Figs. 2[link] and 4[link]). Data collection methods for each crystal can be added and displayed as nodes of a tree directly at the selected sample location in the list (§3.2[link]). When an operation is performed, the background is then coloured either in green (successful), yellow (completed, but without results) or red (failure in the data collection process). Samples can be mounted directly from the sample list (by right clicking on the targeted sample). The execution of chain-loading procedures (i.e. tacit unloading of the current sample by directly loading the following one) is possible and recommended for sample changers requiring de-freezing cycles between explicit operations. If multiple samples are selected and data collection methods added to them, the queue will be executed in pipeline mode and the next sample automatically mounted when the list of tasks for the previous sample is successfully completed. The queue can be stopped or paused at any time by pressing a corresponding button if any modification is required. More advanced functionalities of the sample changer (motor homing, gripper changing, troubleshooting) are accessible by clicking on a dedicated button (`Show SC') that opens a new tab.

[Figure 5]
Figure 5
(Left) The MXCuBE2 sample list after synchronization with ISPyB (via the `Synch ISPyB' button) represents each sample by its name. (Right) Data are automatically saved in a unique folder and file name. When a space group and unit cell is specified in ISPyB, this information is used and transmitted to the auto-processing pipelines.

4.2. Sample viewing, saving position and mesh tool

The sample view displays a video of the sample retrieved through the camera of the sample microscope viewer. The video is always visible, centralizing the user interaction on the mounted sample. To facilitate the use of small X-ray beams to collect (partial) data sets at more than one position in a crystal, MXCuBE2 introduces the possibility to save multiple positions directly on screen and to relate each position in the laboratory and camera frame to the corresponding alignment and positions of the centring motors. Each centred position is tracked and displayed on the screen by a yellow encircled cross, while selected and active saved positions are shown in bold green. It is possible to activate positions sequentially and to add in the queue the same, or different, data collection protocols to different activated positions. The location of one or more small crystals on the same support, or the determination of the best diffracting volume of a large crystal, required the development of dedicated tools that allow rastering through a two-dimensional mesh drawn across the sample support. Originally a very time-consuming procedure, the advent of photon-counting detectors coupled to the synchronized movement of multiple motors (trajectories) led to the implementation of linear scans across the grid by horizontal lines (vertical for upward or downward spindles) while the detector is continuously acquiring data. Eventually, this further evolved in the implementation of a serpentine-like scan across the grid with the detector being triggered in correspondence of each mesh point. The grid drawing tool available in MXCuBE2 uses the current beam size as the default unit for calculating the size and number of grid points in a mesh scan, while it is also possible to specify spacing or overlaps between the measured positions (Fig. 6[link]).

[Figure 6]
Figure 6
(a) Examples of the different mesh tools, with vertical and horizontal spacing defined in the Grid tool dialogue box. (b) After data analysis, grid points are coloured according to the mesh analysis score.

4.3. Data collection methods

The sample list displays data collections that have been performed and those that are currently queued. Data collection operations are assigned to saved positions on a given crystal. Once an operation or a series of operations are assigned to a saved position, if these are still queued (i.e. not yet executed) it is possible to interactively delete or modify these by clicking on the data collection entry, then clicking on the `Remove' or `Details' sub-tab. In the latter, the parameters associated with a data collection can be edited. Some operations [i.e. `Characterization' (§4.3.2[link])] generate a new entry in the data collection tree. In such cases, the user has the choice of automatically launching the new entry or reviewing and/or modifying the experimental parameters generated (again via the `Details' sub-tab) before the data collection is launched.

4.3.1. Standard data collection

`Standard' data collection specifies a `normal' oscillation data collection (i.e. a contiguous total rotation range) operation performed at a single centred point on the crystal of interest. In addition to usual parameters (omega start, exposure time, oscillation, number of frames), it is possible to specify the resolution and the energy at which the diffraction measurements are to be performed. In case the energies for the peak (f′′ max), inflection point (f′ min) and remote were determined from an absorption-edge scan performed on the selected sample, it is possible to select them directly from a MAD dropdown menu within the same widget. Optionally, the space group and unit cell of the crystal can be indicated for use in downstream automatic data-processing pipelines [i.e. GrenADES (Monaco et al., 2013[Monaco, S., Gordon, E., Bowler, M. W., Delagenière, S., Guijarro, M., Spruce, D., Svensson, O., McSweeney, S. M., McCarthy, A. A., Leonard, G. & Nanao, M. H. (2013). J. Appl. Cryst. 46, 804-810.]); XDSAPP (Sparta et al., 2016[Sparta, K. M., Krug, M., Heinemann, U., Mueller, U. & Weiss, M. S. (2016). J. Appl. Cryst. 49, 1085-1092.]); autoPROC (Vonrhein et al., 2011[Vonrhein, C., Flensburg, C., Keller, P., Sharff, A., Smart, O., Paciorek, W., Womack, T. & Bricogne, G. (2011). Acta Cryst. D67, 293-302.]), etc.].

4.3.2. Characterization

Characterization of crystals has the scope of determining an optimal standard data collection strategy that fulfils certain requirements, most notably the taking into account of radiation damage to a sample during a data collection, defined by the user. The MXCuBE2 characterization pipeline is embedded into the EDNA framework (Incardona et al., 2009[Incardona, M.-F., Bourenkov, G. P., Levik, K., Pieritz, R. A., Popov, A. N. & Svensson, O. (2009). J. Synchrotron Rad. 16, 872-879.]) and links software for the indexing and/or integration of diffraction images (MOSFLM, LABELIT, XDS) (Leslie & Powell, 2007[Leslie, A. G. W. & Powell, H. R. (2007). Evolving Methods for Macromolecular Crystallography, pp. 41-51. Springer: The Netherlands.]; Sauter & Poon, 2010[Sauter, N. K. & Poon, B. K. (2010). J. Appl. Cryst. 43, 611-616.]; Kabsch, 2010[Kabsch, W. (2010). Acta Cryst. D66, 125-132.]), absorbed dose calculation/prediction (RADDOSE) (Paithankar & Garman, 2010[Paithankar, K. S. & Garman, E. F. (2010). Acta Cryst. D66, 381-388.]) and data collection strategy calculation (BEST) (Bourenkov & Popov, 2010[Bourenkov, G. P. & Popov, A. N. (2010). Acta Cryst. D66, 409-419.]). The program XOalign (Legrand, unpublished; https://github.com/legrandp/xdsme/tree/master/XOalign) calculates possible crystal reorientation using kappa goniometry that can be exploited further for a new strategy calculation.

The EDNA characterization module relies on various inputs, such as crystal and beam size, and the photon flux impinging on the sample. In most cases this is either stored as metadata in ISPyB and retrieved for analysis or provided directly by MXCuBE2. A successful characterization produces a `diffraction plan', automatically added to the data collection queue. If required, the parameters for the diffraction plan data collection can be edited by the users via the `Details' sub-tab before execution (§4.3[link]).

4.3.3. Helical data collection/characterization

Helical (also known as vector) data collection is performed by continuously rotating a crystal while, at the same time, translating it along a defined line or vector (Flot et al., 2010[Flot, D., Mairs, T., Giraud, T., Guijarro, M., Lesourd, M., Rey, V., van Brussel, D., Morawe, C., Borel, C., Hignette, O., Chavanne, J., Nurizzo, D., McSweeney, S. & Mitchell, E. (2010). J. Synchrotron Rad. 17, 107-118.]). Thus, an undamaged part of the crystal is continuously exposed to the X-ray beam, allowing for the use of higher absorbed dose limits than would be the case for a standard data collection. In MXCuBE2, the line along which the crystal is translated is defined by two previous centred and saved positions on the crystal. Once these are chosen, the user then defines data collection parameters as for a standard data collection. At the ESRF, to help ensure that the absorbed dose is optimized as a function of beam size and vector length, users have the option, found in the `Advanced' data collection tab (§4.3.5[link]), of carrying out a `Helical Characterization'. Upon addition of this option to the data collection queue and its subsequent execution, a data collection `wizard' guides the user through the setting up of the experiment, including the choice of centred positions to be used to define the vector. Once these are defined and saved, MXCuBE2 automatically carries out a characterization at the mid-point between the two positions and calculates an optimal diffraction plan (§4.3.2[link]).

4.3.4. Absorption-edge scan and X-ray fluorescence spectra

In MXCuBE2, the measurement of absorption-edge (XANES; X-ray absorption near-edge structure) scans and X-ray fluorescence (XRF) spectra (Leonard et al., 2009[Leonard, G., Solé, V. A., Beteva, A., Gabadinho, J., Guijarro, M., McCarthy, J., Marrocchelli, D., Nurizzo, D., McSweeney, S. & Mueller-Dieckmann, C. (2009). J. Appl. Cryst. 42, 333-335.]) are treated as any other data collection (i.e. they are added to the queue for subsequent execution) (Fig. 7[link]). Absorption-edge scans can be set up by choosing the relevant element from a periodic table and, if applicable, defining which of the element's L-edges within the reach of the beamline energy range should be targeted. The scan itself is then performed with the lowest photon flux that produces a fluorescence signal above a predefined threshold as measured on the beamline-specific fluorescence detector. The results of the scan are displayed in the corresponding `Details' sub-tab, and the energies for the peak and inflection point obtained after data fitting by CHOOCH (Evans & Pettifer, 2001[Evans, G. & Pettifer, R. F. (2001). J. Appl. Cryst. 34, 82-86.]) are extracted for potential use in ensuing data collections applied to the selected sample. The energy for any `remote' data set is taken as 50 eV above the energy of the peak. It is thus possible, for example, to construct a data collection queue comprising the calculation of a data collection strategy using characterization, the measurement and processing of an absorption edge scan and, finally, standard data collections using the photon energies (peak, inflection point and remote) derived from this. Peak, inflection point and remote energies are automatically stored for the selected sample after each energy scan.

[Figure 7]
Figure 7
Energy scan and X-ray fluorescence spectra are treated as two data collection methods that are added to queue for execution.

X-ray fluorescence spectra are recorded and displayed in a similar way and automatically fitted using the PyMCA package (Solé et al., 2007[Solé, V. A., Papillon, E., Cotte, M., Walter, P. & Susini, J. (2007). At. Spectrosc. 62, 63-68.]). As for any other data collection, results are displayed in the `Details' window. Here, moving of the cursor highlights the energy of emission lines and shows the element most likely to have produced it. Absorption-edge scans, X-ray fluorescence spectra and their results are also stored and displayed in the ISPyB database.

4.3.5. Advanced data collection

The data collection methods included in this area represent more complex data collection protocols that may require input from external data analyses and/or `experiment descriptors' to define the actions to be carried out in subsequent steps of the protocol selected. This can be performed using different decision-making tools. The access to these methods is achieved via the XML-RPC server (§3.3[link]). Installation of the optional methods available in this tab diverges at the different sites where MXCuBE2 is installed, matching local needs and the different configuration of the beamlines.

4.4. Site-specific installations and partner contributions

4.4.1. ESRF

The ESRF Structural Biology Group MX beamlines [ID23-1 (Nurizzo et al., 2006[Nurizzo, D., Mairs, T., Guijarro, M., Rey, V., Meyer, J., Fajardo, P., Chavanne, J., Biasci, J.-C., McSweeney, S. & Mitchell, E. (2006). J. Synchrotron Rad. 13, 227-238.]), ID23-2 (Flot et al., 2010[Flot, D., Mairs, T., Giraud, T., Guijarro, M., Lesourd, M., Rey, V., van Brussel, D., Morawe, C., Borel, C., Hignette, O., Chavanne, J., Nurizzo, D., McSweeney, S. & Mitchell, E. (2010). J. Synchrotron Rad. 17, 107-118.]), ID29 (de Sanctis et al., 2012[Sanctis, D. de, Beteva, A., Caserotto, H., Dobias, F., Gabadinho, J., Giraud, T., Gobbo, A., Guijarro, M., Lentini, M., Lavault, B., Mairs, T., McSweeney, S., Petitdemange, S., Rey-Bakaikoa, V., Surr, J., Theveneau, P., Leonard, G. A. & Mueller-Dieckmann, C. (2012). J. Synchrotron Rad. 19, 455-461.]), ID30A-1 (MASSIF-1) (Bowler et al., 2015[Bowler, M. W., Nurizzo, D., Barrett, R., Beteva, A., Bodin, M., Caserotto, H., Delagenière, S., Dobias, F., Flot, D., Giraud, T., Guichard, N., Guijarro, M., Lentini, M., Leonard, G. A., McSweeney, S., Oskarsson, M., Schmidt, W., Snigirev, A., von Stetten, D., Surr, J., Svensson, O., Theveneau, P. & Mueller-Dieckmann, C. (2015). J. Synchrotron Rad. 22, 1540-1547.]), ID30A-3 (MASSIF-3) (Mueller-Dieckmann et al., 2015[Mueller-Dieckmann, C., Bowler, M. W., Carpentier, P., Flot, D., McCarthy, A. A., Nanao, M. H., Nurizzo, D., Pernot, P., Popov, A., Round, A., Royant, A., de Sanctis, D., von Stetten, D. & Leonard, G. A. (2015). Eur. Phys. J. Plus, 130, 70.]), ID30B (McCarthy et al., 2018[McCarthy, A. A., Barrett, R., Beteva, A., Caserotto, H., Dobias, F., Felisaz, F., Giraud, T., Guijarro, M., Janocha, R., Khadrouche, A., Lentini, M., Leonard, G. A., Lopez Marrero, M., Malbet-Monaco, S., McSweeney, S., Nurizzo, D., Papp, G., Rossi, C., Sinoir, J., Sorez, C., Surr, J., Svensson, O., Zander, U., Cipriani, F., Theveneau, P. & Mueller-Dieckmann, C. (2018). J. Synchrotron Rad. 25, 1249-1260.])] run MXCuBE2 in the Qt3 framework interfaced with the Passerelle workflow engine (https://supportsquare.io/) which manages complex experiments and runs on the Beamline Expert System (BES) server. The BES is a customized version of the Passerelle Enterprise Decision Manager (EDM) that runs the requested workflow on a separate computing cluster. The workflow communicates with MXCuBE2 via the XML-RPC protocol (§3.3[link]) and requests data collections or motor movements (for example, goniometer motors). The different steps of a workflow execution are consecutively executed after either a decision point following the analysis of the data recorded in the previous step or a request for user intervention. Thanks to the ease of the design and the versatility of the XML-RPC communication protocol, the available workflows at the ESRF's MX beamlines are constantly evolving in order to include novel methods and accommodate requests from the user community. The most popular workflows include mesh scans and X-ray centring methods, which are mainly used in different flavours of hands-off data collection protocols (Svensson et al., 2015[Svensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757-1767.]) and in micro-crystallography experiments such as Mesh&Collect (Zander et al., 2015[Zander, U., Bourenkov, G., Popov, A. N., de Sanctis, D., Svensson, O., McCarthy, A. A., Round, E., Gordeliy, V., Mueller-Dieckmann, C. & Leonard, G. A. (2015). Acta Cryst. D71, 2328-2343.]). Dedicated workflows also facilitate use of multi-axis goniometers, particularly in the calculation of the MK3 mini-κ goniometer angles required to adjust the diffraction geometry needed to optimize the total rotation angle to minimize spot overlap or to record Friedel pairs on the same diffraction image (de Sanctis et al., 2016[Sanctis, D. de, Oscarsson, M., Popov, A., Svensson, O. & Leonard, G. (2016). Acta Cryst. D72, 413-420.]; Brockhauser et al., 2013[Brockhauser, S., Ravelli, R. B. G. & McCarthy, A. A. (2013). Acta Cryst. D69, 1241-1251.]). Workflows are also very important in the design and execution of elaborate data collection strategies (i.e. inverse beam, interleaved MAD/SAD) for experimental phasing protocols exploiting anomalous dispersion. MXCuBE2 is also used to navigate and collect diffraction data from in situ crystals using a crystallization plate holder (Arinax, France) on ID30B (McCarthy et al., 2018[McCarthy, A. A., Barrett, R., Beteva, A., Caserotto, H., Dobias, F., Felisaz, F., Giraud, T., Guijarro, M., Janocha, R., Khadrouche, A., Lentini, M., Leonard, G. A., Lopez Marrero, M., Malbet-Monaco, S., McSweeney, S., Nurizzo, D., Papp, G., Rossi, C., Sinoir, J., Sorez, C., Surr, J., Svensson, O., Zander, U., Cipriani, F., Theveneau, P. & Mueller-Dieckmann, C. (2018). J. Synchrotron Rad. 25, 1249-1260.]). Together with the MAX IV team (§4.4.4[link]), the ESRF has developed the latest version of the user interface MXCuBE3 (Mueller et al., 2017[Mueller, U., Thunnissen, M., Nan, J., Eguiraun, M., Bolmsten, F., Milàn-Otero, A., Guijarro, M., Oscarsson, M., de Sanctis, D. & Leonard, G. (2017). Synchrotron Radiat. News. 30, 22-27.]), which has already been proposed to the users of ID29 and ID23-2 and can be used for remote-access experiments by using a recent web browser. The installation on the other ESRF beamlines was completed by the end of 2018.

4.4.2. SOLEIL – PROXIMA-1 and PROXIMA-2A

In line with the implementation of MXCuBE2 at the ESRF, PROXIMA-1 (PX1) and PROXIMA-2 A (PX2-A) at SOLEIL operate the GUI in the Qt3 framework. Although highly similar overall, slight differences exist at the versions of MXCuBE2 installed at the two beamlines due to a difference in beamline hardware and a variability in the scientific specializations. At PX1 and PX2-A, the overall integrity of the diffractometer motors is independently controlled via TANGO devices (Chaize et al., 1999[Chaize, J-M., Gotz, A., Klotz, W-D., Meyer, J., Perez, M. & Taurel, E. (1999). Proceedings of the 7th International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS1999), 4-8 October 1999, Trieste, Italy, pp. 475-479.]; https://www.tango-controls.org/), which are, for the majority, linked to and displayed within MXCuBE2. One major hardware difference on PX1 with other beamlines lies in the specific implementation of a three-axis goniometer based on a κ-geometry. To properly operate the diffractometer while keeping the entire hardware safe from collisions, protocols have been developed to rapidly control the experiments while considering a sphere of collision and safety location of all instruments around the sample position. Consequently, users are provided with only accessible positions for each component based on the positions and input of other diffractometer components, including the beam-stop position, the presence or absence of a pre-sample capillary, the κ-angle of the goniometer, the targeted energy for the data collection and the targeted resolution (both of which have an influence on the sample-to-detector and sample-to-beam-stop distances).

4.4.3. HZB (BESSY II)

At the BESSY II electron storage ring, two energy-tuneable MX beamlines and one fixed-energy station (Mueller et al., 2015[Mueller, U., Förster, R., Hellmig, M., Huschmann, F. U., Kastner, A., Malecki, P., Pühringer, S., Röwer, M., Sparta, K., Steffien, M., Ühlein, M., Wilk, P. & Weiss, M. S. (2015). Eur. Phys. J. Plus, 130, 141.], 2012[Mueller, U., Darowski, N., Fuchs, M. R., Förster, R., Hellmig, M., Paithankar, K. S., Pühringer, S., Steffien, M., Zocher, G. & Weiss, M. S. (2012). J. Synchrotron Rad. 19, 442-449.]) are currently operated by the HZB-MX group. Both energy tuneable beamlines, BL14.1 and BL14.2, are controlled by MXCuBE2 featuring the Qt4-based user interface. A further installation of MXCuBE2 is anticipated in the scope of the experimental-station upgrade of the fixed-energy beamline BL14.3 scheduled for early 2019. Although the end-station hardware (including the main components such as the diffractometer and the sample-transfer robot) differ significantly between beamlines, an almost identical user interface can be offered to the beamline users thanks to the MXCuBE2 hardware abstraction layer. Beamline BL14.1 features a MD2-microdiffractometer (Arinax, France), a CATS sample changer (IRELEC, France) and a PILATUS 6M detector (DECTRIS, Switzerland). In order to integrate the CATS sample changer, a new hardware object had to be developed. Recently, a plate manipulator (Arinax, France) has been installed, which will soon be integrated and will allow in situ crystal screening and data collection. In order to support the GROB sample-changing robot (NatX-ray, France) as well as the piezo-controlled beam-shaping devices and the fast air-bearing axis of the Nanodiff-diffractometer on HZB's most recent MX beamline BL14.2, new hardware object software modules were added to the pool of supported control devices. MXCuBE2 at HZB (BESSY II) currently supports crystal characterization by EDNA and standard data collection workflows with full sample-changer integration well suited for routine data acquisition, e.g. in fragment-screening campaigns (Huschmann et al., 2016[Huschmann, F. U., Linnik, J., Sparta, K., Ühlein, M., Wang, X., Metz, A., Schiebel, J., Heine, A., Klebe, G., Weiss, M. S. & Mueller, U. (2016). Acta Cryst. F72, 346-355.]). These workflows are complemented with XANES scans and energy-dispersive spectra using an Amptek silicon-drift fluorescence detector for de novo structure solution applications.

In order to enhance the locally available experimental options for users, the integration of the helical data collection protocol as well as the two-dimensional grid scans has started and the release of this for standard user experiments is expected in early 2019. It is also planned to link the XDS data processing expert system XDSAPP (Sparta et al., 2016[Sparta, K. M., Krug, M., Heinemann, U., Mueller, U. & Weiss, M. S. (2016). J. Appl. Cryst. 49, 1085-1092.]), developed at HZB (BESSY II) directly to data acquisition via MXCuBE2.

4.4.4. MAX IV

At the old MAX-lab (MAX II), MXCuBE2 was used to control the experimental setup of the former energy-tuneable MX-beamline 911-3 until its decommissioning at the end of 2015. This included the control of the main beamline functions and support of the experimental station environment including: an MD2 micro-diffractometer, a CATS sample changer (IRELEC, France) and a Rayonix 225 CCD detector (Rayonix, USA). As a result of the successful MXCuBE collaboration, full support of the CATS sample changer using an ALBA developed TANGO device server and a HZB developed CATS hardware object was implemented and utilized. Furthermore, a cryo-shutter designed by HZB was integrated into MXCuBE2 for crystal annealing. With the setup of remote access in MXCuBE2, the MAX IV team together with SOLEIL could perform different data collection experiments remotely. Data characterization by EDNA and data evaluation using the GrenADeS [developed by ESRF (Monaco et al., 2013[Monaco, S., Gordon, E., Bowler, M. W., Delagenière, S., Guijarro, M., Spruce, D., Svensson, O., McSweeney, S. M., McCarthy, A. A., Leonard, G. & Nanao, M. H. (2013). J. Appl. Cryst. 46, 804-810.])] and EDNA processing pipelines (Incardona et al., 2009[Incardona, M.-F., Bourenkov, G. P., Levik, K., Pieritz, R. A., Popov, A. N. & Svensson, O. (2009). J. Synchrotron Rad. 16, 872-879.]) were used during user operations on a daily basis. The MAX IV team developed their own hardware object and evaluation environment for carrying out XRF scans using an Amptek silicon drift detector. Finally, all samples and related data acquisition and processing results were stored in the ISPyB LIMS database, and were accessible to users after their experiments.

At the new MAX IV synchrotron (3 GeV storage ring), the MX team has been developing the next generation interface, MXCuBE3, in close collaboration with the ESRF (Mueller et al., 2017[Mueller, U., Thunnissen, M., Nan, J., Eguiraun, M., Bolmsten, F., Milàn-Otero, A., Guijarro, M., Oscarsson, M., de Sanctis, D. & Leonard, G. (2017). Synchrotron Radiat. News. 30, 22-27.]). The new MXCuBE3 has been deployed at the first MX-beamline (BioMAX) of the MAX IV Laboratory (Thunnissen et al., 2013[Thunnissen, M. M. G. M., Sondhauss, P., Wallén, E., Theodor, K., Logan, D. T., Labrador, A., Unge, J., Appio, R., Fredslund, F. & Ursby, T. (2013). J. Phys. Conf. Ser. 425, 072012.]) and is used as the beamline control GUI during user experiments. While the hardware object layer of MXCuBE3 has been inherited from MXCuBE2, preserving a high downstream compatibility, a completely new and user-friendly interface for MXCuBE3 has been conceived as a web application. Within its current implementation, MXCuBE3 supports the MD3 micro-diffractometer (Arinax, France) and the EIGER 16M hybrid pixel detector (DECTRIS, Switzerland), providing standard data collections, including helical scans and the basic beamline functions. Control of the ISARA (IRELEC, France) sample changer and additional features, including raster scans, XRF scans, XANES, workflows and remote access are currently being developed and will be implemented in the near future.

4.4.5. EMBL-Hamburg

Two MX beamlines, P13 and P14, run by the European Molecular Biology Laboratory (EMBL) at the PETRA III storage ring are currently using MXCuBE2 featuring a graphics layer built with Qt4 libraries. The migration to Qt4 of the core Qt-components of the BlissFramework (such as configuration tools, GUI designer and GUI supervisor) was contributed to the collaboration by EMBL-Hamburg. The core endstation components are an MD2 diffractometer (Arinax, France) and a PILATUS 6M-F detector (DECTRIS, Switzerland) at P13, and an MD3 diffractometer (Arinax, France) and an EIGER 16M detector (DECTRIS, Switzerland) at P14. Both beamlines operate cryogenic `MARVIN' sample changers developed at EMBL-Hamburg (Fiedler et al., unpublished). On both diffractometers, mini-κ goniostats (Brockhauser et al., 2013[Brockhauser, S., Ravelli, R. B. G. & McCarthy, A. A. (2013). Acta Cryst. D69, 1241-1251.]) are installed and supported by MXCuBE2. At P14, MXCuBE2 provides additional support to the operation of the CrystalDirectTM (Cipriani et al., 2012[Cipriani, F., Röwer, M., Landret, C., Zander, U., Felisaz, F. & Márquez, J. A. (2012). Acta Cryst. D68, 1393-1399.]) plates with the plate manipulator when this is mounted on the MD3.

In contrast to other sites involved in the MXCuBE Collaboration, at P13 and P14 MXCuBE2 communicates with beamline devices via the TINE (Bartkiewicz & Duval, 2007[Bartkiewicz, P. & Duval, P. (2007). Meas. Sci. Technol. 18, 2379-2386.]) control system. Furthermore, data acquisition sequences at the `Start/Execute' step are delegated to middleware TINE services which tightly coordinate the actions of a multitude of devices, as briefly described in Karpics et al. (2016[Karpics, I., Bourenkov, G., Nikolova, M. & Schneider, T. R. (2016). Enhancing the MxCuBE User Interface by a Finite State Machine (FSM) Model, NOBUGS, 8-13 Oct 2017. Barcelona, Spain. TUPHA187.]). For the `Collection methods', MXCuBE2 at P13/P14 implements standard, helical and mesh data collections, sample characterization, and XRF and energy scans. In these methods, automated data analysis is triggered and the results are deposited in ISPyB. For rotation data processing, EDNAproc (Monaco et al., 2013[Monaco, S., Gordon, E., Bowler, M. W., Delagenière, S., Guijarro, M., Spruce, D., Svensson, O., McSweeney, S. M., McCarthy, A. A., Leonard, G. & Nanao, M. H. (2013). J. Appl. Cryst. 46, 804-810.]), autoPROC (Vonrhein et al., 2011[Vonrhein, C., Flensburg, C., Keller, P., Sharff, A., Smart, O., Paciorek, W., Womack, T. & Bricogne, G. (2011). Acta Cryst. D67, 293-302.]) and XIA2 (Winter, 2010[Winter, G. (2010). J. Appl. Cryst. 43, 186-190.]) are used.

When executing queues consisting of standard or helical collections, MXCuBE2 provides an option for interleaved – as opposed to sequential – execution of queue elements. Thus, interleaved collections at different energies, rotation angles, κ-diffractometer settings, sample centrings, or any other collection setting are achieved via a straightforward and transparent interface.

Under the `Advanced' collection method, mesh and serial helical-scan data collections [mesh scans with larger rotation angles per line (Gati et al., 2014[Gati, C., Bourenkov, G., Klinge, M., Rehders, D., Stellato, F., Oberthür, D., Yefanov, O., Sommer, B. P., Mogk, S., Duszenko, M., Betzel, C., Schneider, T. R., Chapman, H. N. & Redecke, L. (2014). IUCrJ, 1, 87-94.])] are offered. With small and intense beams, in combination with a high-frame-rate detector at P14, such collections often involve many tens of thousands of frames, requiring the customization of the representation of meshes. After delegating the scan execution to middleware TINE services (and further to the low-level hardware control), MXCuBE2 triggers automatic diffraction detection using the program DOZOR (Popov & Bourenkov, 2016[Popov, A. N. & Bourenkov, G. (2016). DOZOR. ESRF, Grenoble, France.]) encapsulated in the EDNA plugin (Incardona et al., 2009[Incardona, M.-F., Bourenkov, G. P., Levik, K., Pieritz, R. A., Popov, A. N. & Svensson, O. (2009). J. Synchrotron Rad. 16, 872-879.]). MXCuBE2 receives the results of the DOZOR analysis via XML-RPC and constructs and presents a diffraction heat map on-the-fly. The heat map is linked to the diffraction image display and provides a broad range of interactive functionalities for analysis and definition of centring points for subsequent data collections.

At the EMBL-Hamburg beamlines, MXCuBE2 implements a number of customized features for automatic re-configuration of the beamline optics and for automatic beam centring. For instance, when P14 operates in a collimated regime, the shape of the (top-hat) beam at the sample is defined by dragging a rectangle over the sample on the video display.

Remote access to the complete data collection environment at P13/P14, including the single instance of MXCuBE2 running at each beamline, is provided via the TeamViewer (https://www.teamviewer.com) remote desktop software.

4.4.6. ALBA

At ALBA, MXCuBE2 is the user interface for the MX beamlines, XALOC (Juanhuix et al., 2014[Juanhuix, J., Gil-Ortiz, F., Cuní, G., Colldelram, C., Nicolás, J., Lidón, J., Boter, E., Ruget, C., Ferrer, S. & Benach, J. (2014). J. Synchrotron Rad. 21, 679-689.]) and XAIRA at the ALBA synchrotron. MXCuBE2 has been recently implemented for regular use at the operational beamline XALOC after an extended commissioning phase with external users. XAIRA, which is the microfocus beamline currently under construction at ALBA, will also benefit from the implementation of MXCuBE2, even though the hardware for this beamline differs substantially from that installed at XALOC.

The current implementation of MXCuBE2 at XALOC is built on top of the Sardana/Taurus control system and TANGO (Reszela et al., 2014[Reszela, Z., Cuní, G., Fernández-Carreiras, D., Klora, J. & Pascual-Izarra, C. (2014). Proceedings of the 10th International Workshop on Personal Computer and Particle Accelerator Controls (PCaPAC 2014), 14-17 October 2014, Karlsruhe, Germany. WCO206.], 2017[Reszela, Z., Becheri, F., Cuní, G., Falcon-Torres, C., Fernández-Carreiras, D., Homs-Puron, R., Moldes, J., Pascual-Izarra, C., Pastor Ortiz, R., Roldán, D. & Rosanes Siscart, M. (2017). Proceedings of the 16th International Conference on Accelerator and Large Experimental Control Systems (ICALEPCS2017). 8-13 October 2017, Barcelona, Spain, pp. 1067-1072. WEBPL06.]; Fernández-Carreiras et al., 2011[Fernández-Carreiras, D., Beltran, D., Coutinho, T., Cuní, G., Klora, J., Matilla, O., Montaño, R., Pascual-Izarra, C., Pusó, S., Ranz, R., Rubio, A., Rubio-Manrique, S. & Suñé, R. (2011). Proceedings of the 13th International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS2011), 10-14 October 2011, WTC Grenoble, France, pp. 1318-1321.]; Pascual-Izarra et al., 2015[Pascual-Izarra, C., Coutinho, T., Cuní, G., Falcón Torres, C., Fernández-Carreiras, D., Reszela, Z. & Rosanes Siscart, M. (2015). Proceedings of the 15th International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS 2015), 17-23 October 2015, Melbourne, Australia. THHC3O03.]). The hardware installed on the beamline is accessed through the Sardana layer or the TANGO Device Server (DS) and linked to the MXCuBE2 user interface. Different generic and specific hardware objects for the current hardware are used as connectors between the native control system and the MXCuBE2/Qt4 interface, while standard collection methods are implemented using the Sardana Scan Framework. The overall status of the beamline is managed with a specific Beamline Supervisor TANGO DS. This supervisor defines a set of beamline configurations (or phases) corresponding to different operation procedures such as `sample change', `diffraction' or `beam visualization', providing a robust machine state and safe operation. Furthermore, this implementation allows for a smooth future transition from MXCuBE2 to the web-based MXCuBE3.

Test images are processed using EDNA running on a computing cluster at ALBA, and the results are uploaded to the ISPyB database and made available to users. Collected data are processed on the cluster using the autoPROC program from Global Phasing Ltd (Cambridge, UK) (Von­rhein et al., 2011[Vonrhein, C., Flensburg, C., Keller, P., Sharff, A., Smart, O., Paciorek, W., Womack, T. & Bricogne, G. (2011). Acta Cryst. D67, 293-302.]). The default data processing is performed using standard parameters, with default resolution cutoffs based on CC1/2. The user can request a change of any of the processing parameters. The data can conveniently be reprocessed and phased using molecular replacement [Phaser, (McCoy et al., 2007[McCoy, A. J., Grosse-Kunstleve, R. W., Adams, P. D., Winn, M. D., Storoni, L. C. & Read, R. J. (2007). J. Appl. Cryst. 40, 658-674.])] or ARCIMBOLDO (Rodríguez et al., 2009[Rodríguez, D. D., Grosse, C., Himmel, S., González, C., de Ilarduya, I. M., Becker, S., Sheldrick, G. M. & Usón, I. (2009). Nat. Methods, 6, 651-653.]) using Xamurai (unpublished), an in-house-developed program.

In the near future, the implementation of MXCuBE2 at XALOC will be expanded to include automated beam-conditioning options such as the adjustment of slits and apertures, and to offer more complex and tailored methods of data collection, including interleaved data collections. XAIRA will implement these developments while also including options specific to its scientific case and final hardware configuration.

4.4.7. Global Phasing Ltd

Global Phasing Ltd (GPhL) is the only non-synchrotron member of the MXCuBE Collaboration. Its involvement in joint developments with synchrotrons dates back to its participation in the BIOXHIT project, with a special interest in the data processing aspects of the deployment of multi-axis goniometry for experimental phasing that was taking place at that time. Since then, as part of its activities in developing improved and better-integrated software for using MX as a tool for structure-based drug discovery, with support from the pharmaceutical industry for the past 20+ years, GPhL has pursued sustained efforts towards finding ways of increasing the quality of X-ray data collected for high-throughput ligand screening by crystallography. High degrees of automation have been achieved towards this goal, e.g. on the MASSIF-1 (ESRF ID30A-1) (Bowler et al., 2015[Bowler, M. W., Nurizzo, D., Barrett, R., Beteva, A., Bodin, M., Caserotto, H., Delagenière, S., Dobias, F., Flot, D., Giraud, T., Guichard, N., Guijarro, M., Lentini, M., Leonard, G. A., McSweeney, S., Oskarsson, M., Schmidt, W., Snigirev, A., von Stetten, D., Surr, J., Svensson, O., Theveneau, P. & Mueller-Dieckmann, C. (2015). J. Synchrotron Rad. 22, 1540-1547.]) and XChem (Diamond I04-1) beamlines, resulting in protocols that are very time-efficient by being `minimalistic'. This entails skipping steps such as the design of strategies using multiple orientations adapted to the symmetry and orientation of each individual crystal. The rationale for this choice is that although such steps are known to have the potential of significantly boosting the quality of the final datasets, they would, in the current state of affairs, require human intervention and thus disrupt automation. GPhL's work on the implementation of fully automated expert protocols was initiated in July 2013 as part of a collaboration with Diamond focused on the I23 beamline, on which the curved geometry and the resulting aspect ratio of the PILATUS 12M detector made it absolutely necessary to use multiple-orientation strategies and to make their design completely automatic, with built-in avoidance of collisions as well as anticipation and remediation of shadowing by the goniometer. For reasons both methodological (viz. aiming at a maximally generic and transferable implementation) and practical (the GPhL funding model, emphasizing the need for maximum ease of deployment on a variety of beamlines at multiple synchrotrons) it was decided that the software should take the form of a unique workflow, embodying all the logic and expertise needed to design optimal multi-orientation strategies on-the-fly and to supervise their execution by directly communicating with the beamline control software (BCS), including, crucially, the automatic re-centring of the sample after each change of orientation. A major milestone of this collaboration was reached in September 2016, when the GPhL workflow steered a completely automated three-wavelength interleaved MAD experiment on a lysozyme crystal (including the alignment of its fourfold axis and the use of a cusp-filling orientation) on the Diamond I04-1 beamline. This steering took place through a message bus connecting the workflow with GDA (the Diamond BCS), based on an abstract beamline interface intended to make the task of writing similar connections with other BCSs as simple as possible.

Work aimed at creating such a connection with MXCuBE2 began in the summer of 2017 and was able to progress rapidly thanks to access to the MXCuBE2 source code and to its developers, provided by membership of the MXCuBE collaboration. In particular, the need to enlarge the internal execution queue of the old MXCuBE to accommodate the `third-party design and control' paradigm proposed by GPhL was recognized and assigned a high priority by the developers, and that enlargement was a major structural component of the transition to MXCuBE2. The first live tests of this connection (with the Qt3 version of MXCuBE2) took place on the ESRF ID30B beamline in October 2017, and by May 2018 the GPhL team was able to successfully run remotely a similar (fully automated three-wavelength interleaved MAD) experiment to the one previously performed at Diamond. Work is underway to extend the scope of this connection so as to include the Qt4 version of MXCuBE2 (on beamlines at ALBA and SOLEIL) and to MXCuBE3.

4.5. Remote access

Both MXCuBE and MXCuBE2 offer integrated remote access features (Gabadinho et al., 2010[Gabadinho, J., Beteva, A., Guijarro, M., Rey-Bakaikoa, V., Spruce, D., Bowler, M. W., Brockhauser, S., Flot, D., Gordon, E. J., Hall, D. R., Lavault, B., McCarthy, A. A., McCarthy, J., Mitchell, E., Monaco, S., Mueller-Dieckmann, C., Nurizzo, D., Ravelli, R. B. G., Thibault, X., Walsh, M. A., Leonard, G. A. & McSweeney, S. M. (2010). J. Synchrotron Rad. 17, 700-707.]) to the beamlines where they are installed. In the current remote access set-up, access to the local control computer is performed via dedicated software such as NX NoMachine, which connects to a dedicated machine inside the synchrotron site firewall and from there to the local control machine. Although remote control of MX beamlines is increasing, performance can suffer from deterioration in connection speed. In order to facilitate and improve remote access, MXCuBE is currently being redesigned as a web application called MXCuBE3.

MXCuBE3 simplifies and speeds up remote-access procedures as it will not require any installation on the client (remote) side other than a recent version of a web browser and the ability of the client to connect to the beamline via a predefined port. Currently available at the BioMAX beamline of the MAX IV Laboratory (§4.4.4[link]) and at the ID29 and ID23-2 beamlines at the ESRF (§4.4.1[link]), MXCuBE3 is platform independent and uses the same hardware repository as MXCuBE2, making it possible, without any change in configuration, to run either MXCuBE2 or the new web-based application on the same control computer. While a preliminary report on the design of MXCuBE3 has already been published (Mueller et al., 2017[Mueller, U., Thunnissen, M., Nan, J., Eguiraun, M., Bolmsten, F., Milàn-Otero, A., Guijarro, M., Oscarsson, M., de Sanctis, D. & Leonard, G. (2017). Synchrotron Radiat. News. 30, 22-27.]), further details of this new avenue for the MXCuBE Collaboration will be given in due course.

5. Discussion

The MXCuBE Collaboration brings together different partners aiming for the development of a common beamline control environment for macromolecular crystallography experiments. The European MX user-community takes advantage of finding a very similar and intuitive graphical interface when performing their experiments at different beamlines. Currently the collaboration is developing a new interface (MXCuBE3) that can ensure the future sustainability of MXCuBE and permit the evolution of the graphical user interface over time. The software architecture is explicitly designed to facilitate the deployment on any hardware controller layer at different synchrotron sites and the necessary development for the implementation of new hardware and new data collection methods can easily be adopted by all MXCuBE Collaboration partner sites. This agility allows MXCuBE to be ready for the future challenges in MX, and of serial crystallography at synchrotrons in particular.

Starting from the early implementations at X-ray free-electron laser sources (Chapman et al., 2011[Chapman, H. N., Fromme, P., Barty, A., White, T. A., Kirian, R. A., Aquila, A., Hunter, M. S., Schulz, J., DePonte, D. P., Weierstall, U., Doak, R. B., Maia, F. R. N. C., Martin, A. V., Schlichting, I., Lomb, L., Coppola, N., Shoeman, R. L., Epp, S. W., Hartmann, R., Rolles, D., Rudenko, A., Foucar, L., Kimmel, N., Weidenspointner, G., Holl, P., Liang, M., Barthelmess, M., Caleman, C., Boutet, S., Bogan, M. J., Krzywinski, J., Bostedt, C., Bajt, S., Gumprecht, L., Rudek, B., Erk, B., Schmidt, C., Hömke, A., Reich, C., Pietschner, D., Strüder, L., Hauser, G., Gorke, H., Ullrich, J., Herrmann, S., Schaller, G., Schopper, F., Soltau, H., Kühnel, K.-U., Messerschmidt, M., Bozek, J. D., Hau-Riege, S. P., Frank, M., Hampton, C. Y., Sierra, R. G., Starodub, D., Williams, G. J., Hajdu, J., Timneanu, N., Seibert, M. M., Andreasson, J., Rocker, A., Jönsson, O., Svenda, M., Stern, S., Nass, K., Andritschke, R., Schröter, C.-D., Krasniqi, F., Bott, M., Schmidt, K. E., Wang, X., Grotjohann, I., Holton, J. M., Barends, T. R. M., Neutze, R., Marchesini, S., Fromme, R., Schorb, S., Rupp, D., Adolph, M., Gorkhover, T., Andersson, I., Hirsemann, H., Potdevin, G., Graafsma, H., Nilsson, B. & Spence, J. C. H. (2011). Nature, 470, 73-77.]), MX is experiencing a paradigm shift in approaches to data collection, where the traditional dogma of `one crystal, one data collection' is being replaced by more complex, yet more efficient, strategies using multiple crystal orientations of a single crystal (Weinert et al., 2015[Weinert, T., Olieric, V., Waltersperger, S., Panepucci, E., Chen, L., Zhang, H., Zhou, D., Rose, J., Ebihara, A., Kuramitsu, S., Li, D., Howe, N., Schnapp, G., Pautsch, A., Bargsten, K., Prota, A. E., Surana, P., Kottur, J., Nair, D. T., Basilico, F., Cecatiello, V., Pasqualato, S., Boland, A., Weichenrieder, O., Wang, B.-C., Steinmetz, M. O., Caffrey, M. & Wang, M. (2015). Nat. Methods, 12, 131-133.]; Klinke et al., 2015[Klinke, S., Foos, N., Rinaldi, J. J., Paris, G., Goldbaum, F. A., Legrand, P., Guimarães, B. G. & Thompson, A. (2015). Acta Cryst. D71, 1433-1443.]), multiple crystals to increase the multiplicity (and hence statistical reliability) in anomalous data collection (Liu et al., 2012[Liu, Q., Dahmane, T., Zhang, Z., Assur, Z., Brasch, J., Shapiro, L., Mancia, F. & Hendrickson, W. A. (2012). Science, 336, 1033-1037.]), right through to serial crystallography methods (Standfuss & Spence, 2017[Standfuss, J. & Spence, J. (2017). IUCrJ, 4, 100-101.]). These serial methods, at least for the present, can be viewed as complementary to (and not a replacement for) more traditional data collection, offering methods of extracting useful information from crystals of micrometre dimensions or smaller, being able to mitigate the effects of radiation damage by sharing dose over many crystals (which exhibit presumably identical internal molecular structures – the ultimate case of isomorphism), and even permitting the extraction of usable anomalous signal for phasing (Zander et al., 2015[Zander, U., Bourenkov, G., Popov, A. N., de Sanctis, D., Svensson, O., McCarthy, A. A., Round, E., Gordeliy, V., Mueller-Dieckmann, C. & Leonard, G. A. (2015). Acta Cryst. D71, 2328-2343.]; Schlichting, 2017[Schlichting, I. (2017). IUCrJ, 4, 516-517.]). Although the extent of the role of such methods in the crystallographers' armoury is as yet unclear, it is certain that the `book-keeping' of highly multiple-crystal data collection represents a major challenge. For example, the automatic presentation of many expected crystal positions (from a grid or plate) along with the registration of the necessary parameters (rotation angle, oscillation range if applicable and their link to an automatically assigned image number etc.) for many hundreds or indeed thousands of positions will be essential in order to pave the way for future automatic data collections and subsequent processing. MXCuBE and the different software programs that work in association with it (ISPyB, different automatic data analysis pipelines such as XIA2, autoPROC, XDSME, XDSAPP, GrenADeS etc.) will require major changes based on a generic data model which includes the possibility of interleaving multiple data collections (energy, orientation) with multiple samples (from several up to many thousands) coupled with rapidity and fluidity of a user interface and the correct level of information about the progress of the different types of data collection, permitting decision making by the user.

Footnotes

Current address: Energy Sciences Directorate, NSLS II, Brookhaven National Laboratory, Upton, NY 11973-5000, USA.

§Current address: Paul Scherrer Institut, 5232 Villigen PSI, Switzerland.

Funding information

The ESRF–EMBL Joint Structural Biology Group (JSBG) and the ESRF Automation Task Force (ATF) are gratefully acknowledged for supporting the development. Parts of the work presented here were supported by European Commission 6th Framework Programme `Life Sciences, Genomics and Biotechnology for Health' (Integrated Research project BIOXHIT, Contract No. LHSG-CT-2003–503420), by iNEXT, project number 653706, funded by the Horizon 2020 programme of the European Union and by Vetenskapsrådet (SwedSTRUCT and MAX IV – SOLEIL).

References

First citationArzt, S., Beteva, A., Cipriani, F., Delageniere, S., Felisaz, F., Förstner, G., Gordon, E., Launer, L., Lavault, B., Leonard, G., Mairs, T., McCarthy, A., McCarthy, J., McSweeney, S., Meyer, J., Mitchell, E., Monaco, S., Nurizzo, D., Ravelli, R., Rey, V., Shepard, W., Spruce, D., Svensson, O. & Theveneau, P. (2005). Prog. Biophys. Mol. Biol. 89, 124–152.  Web of Science CrossRef PubMed CAS Google Scholar
First citationBartkiewicz, P. & Duval, P. (2007). Meas. Sci. Technol. 18, 2379–2386.  Web of Science CrossRef CAS Google Scholar
First citationBerman, H. M., Westbrook, J., Feng, Z., Gilliland, G., Bhat, T. N., Weissig, H., Shindyalov, I. N. & Bourne, P. E. (2000). Nucleic Acids Res. 28, 235–242.  Web of Science CrossRef PubMed CAS Google Scholar
First citationBourenkov, G. P. & Popov, A. N. (2010). Acta Cryst. D66, 409–419.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBowler, M. W., Nurizzo, D., Barrett, R., Beteva, A., Bodin, M., Caserotto, H., Delagenière, S., Dobias, F., Flot, D., Giraud, T., Guichard, N., Guijarro, M., Lentini, M., Leonard, G. A., McSweeney, S., Oskarsson, M., Schmidt, W., Snigirev, A., von Stetten, D., Surr, J., Svensson, O., Theveneau, P. & Mueller-Dieckmann, C. (2015). J. Synchrotron Rad. 22, 1540–1547.  Web of Science CrossRef IUCr Journals Google Scholar
First citationBrockhauser, S., Ravelli, R. B. G. & McCarthy, A. A. (2013). Acta Cryst. D69, 1241–1251.  Web of Science CrossRef IUCr Journals Google Scholar
First citationBrockhauser, S., Svensson, O., Bowler, M. W., Nanao, M., Gordon, E., Leal, R. M. F., Popov, A., Gerring, M., McCarthy, A. A. & Gotz, A. (2012). Acta Cryst. D68, 975–984.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationChaize, J-M., Gotz, A., Klotz, W-D., Meyer, J., Perez, M. & Taurel, E. (1999). Proceedings of the 7th International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS1999), 4–8 October 1999, Trieste, Italy, pp. 475–479.  Google Scholar
First citationChapman, H. N., Fromme, P., Barty, A., White, T. A., Kirian, R. A., Aquila, A., Hunter, M. S., Schulz, J., DePonte, D. P., Weierstall, U., Doak, R. B., Maia, F. R. N. C., Martin, A. V., Schlichting, I., Lomb, L., Coppola, N., Shoeman, R. L., Epp, S. W., Hartmann, R., Rolles, D., Rudenko, A., Foucar, L., Kimmel, N., Weidenspointner, G., Holl, P., Liang, M., Barthelmess, M., Caleman, C., Boutet, S., Bogan, M. J., Krzywinski, J., Bostedt, C., Bajt, S., Gumprecht, L., Rudek, B., Erk, B., Schmidt, C., Hömke, A., Reich, C., Pietschner, D., Strüder, L., Hauser, G., Gorke, H., Ullrich, J., Herrmann, S., Schaller, G., Schopper, F., Soltau, H., Kühnel, K.-U., Messerschmidt, M., Bozek, J. D., Hau-Riege, S. P., Frank, M., Hampton, C. Y., Sierra, R. G., Starodub, D., Williams, G. J., Hajdu, J., Timneanu, N., Seibert, M. M., Andreasson, J., Rocker, A., Jönsson, O., Svenda, M., Stern, S., Nass, K., Andritschke, R., Schröter, C.-D., Krasniqi, F., Bott, M., Schmidt, K. E., Wang, X., Grotjohann, I., Holton, J. M., Barends, T. R. M., Neutze, R., Marchesini, S., Fromme, R., Schorb, S., Rupp, D., Adolph, M., Gorkhover, T., Andersson, I., Hirsemann, H., Potdevin, G., Graafsma, H., Nilsson, B. & Spence, J. C. H. (2011). Nature, 470, 73–77.  Web of Science CrossRef CAS PubMed Google Scholar
First citationCipriani, F., Röwer, M., Landret, C., Zander, U., Felisaz, F. & Márquez, J. A. (2012). Acta Cryst. D68, 1393–1399.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationDelagenière, S., Brenchereau, P., Launer, L., Ashton, A. W., Leal, R., Veyrier, S., Gabadinho, J., Gordon, E. J., Jones, S. D., Levik, K. E., McSweeney, S. M., Monaco, S., Nanao, M., Spruce, D., Svensson, O., Walsh, M. A. & Leonard, G. A. (2011). Bioinformatics, 27, 3186–3192.  Web of Science PubMed Google Scholar
First citationEvans, G. & Pettifer, R. F. (2001). J. Appl. Cryst. 34, 82–86.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationFernández-Carreiras, D., Beltran, D., Coutinho, T., Cuní, G., Klora, J., Matilla, O., Montaño, R., Pascual-Izarra, C., Pusó, S., Ranz, R., Rubio, A., Rubio-Manrique, S. & Suñé, R. (2011). Proceedings of the 13th International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS2011), 10–14 October 2011, WTC Grenoble, France, pp. 1318–1321.  Google Scholar
First citationFlot, D., Mairs, T., Giraud, T., Guijarro, M., Lesourd, M., Rey, V., van Brussel, D., Morawe, C., Borel, C., Hignette, O., Chavanne, J., Nurizzo, D., McSweeney, S. & Mitchell, E. (2010). J. Synchrotron Rad. 17, 107–118.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationGabadinho, J., Beteva, A., Guijarro, M., Rey-Bakaikoa, V., Spruce, D., Bowler, M. W., Brockhauser, S., Flot, D., Gordon, E. J., Hall, D. R., Lavault, B., McCarthy, A. A., McCarthy, J., Mitchell, E., Monaco, S., Mueller-Dieckmann, C., Nurizzo, D., Ravelli, R. B. G., Thibault, X., Walsh, M. A., Leonard, G. A. & McSweeney, S. M. (2010). J. Synchrotron Rad. 17, 700–707.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationGati, C., Bourenkov, G., Klinge, M., Rehders, D., Stellato, F., Oberthür, D., Yefanov, O., Sommer, B. P., Mogk, S., Duszenko, M., Betzel, C., Schneider, T. R., Chapman, H. N. & Redecke, L. (2014). IUCrJ, 1, 87–94.  Web of Science CrossRef CAS PubMed IUCr Journals Google Scholar
First citationGuijarro, M., Berruyer, G., Klora, J. & Rey-Bakaikoa, V. (2004). Bliss Framework, NOBUGS 5, 18–20 October 2004, PSI, Switzerland.  Google Scholar
First citationHuschmann, F. U., Linnik, J., Sparta, K., Ühlein, M., Wang, X., Metz, A., Schiebel, J., Heine, A., Klebe, G., Weiss, M. S. & Mueller, U. (2016). Acta Cryst. F72, 346–355.  CrossRef IUCr Journals Google Scholar
First citationIncardona, M.-F., Bourenkov, G. P., Levik, K., Pieritz, R. A., Popov, A. N. & Svensson, O. (2009). J. Synchrotron Rad. 16, 872–879.  Web of Science CrossRef IUCr Journals Google Scholar
First citationJuanhuix, J., Gil-Ortiz, F., Cuní, G., Colldelram, C., Nicolás, J., Lidón, J., Boter, E., Ruget, C., Ferrer, S. & Benach, J. (2014). J. Synchrotron Rad. 21, 679–689.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationKabsch, W. (2010). Acta Cryst. D66, 125–132.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationKarpics, I., Bourenkov, G., Nikolova, M. & Schneider, T. R. (2016). Enhancing the MxCuBE User Interface by a Finite State Machine (FSM) Model, NOBUGS, 8–13 Oct 2017. Barcelona, Spain. TUPHA187.  Google Scholar
First citationKlinke, S., Foos, N., Rinaldi, J. J., Paris, G., Goldbaum, F. A., Legrand, P., Guimarães, B. G. & Thompson, A. (2015). Acta Cryst. D71, 1433–1443.  CrossRef IUCr Journals Google Scholar
First citationLeonard, G., Solé, V. A., Beteva, A., Gabadinho, J., Guijarro, M., McCarthy, J., Marrocchelli, D., Nurizzo, D., McSweeney, S. & Mueller-Dieckmann, C. (2009). J. Appl. Cryst. 42, 333–335.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationLeslie, A. G. W. & Powell, H. R. (2007). Evolving Methods for Macromolecular Crystallography, pp. 41–51. Springer: The Netherlands.  Google Scholar
First citationLiu, Q., Dahmane, T., Zhang, Z., Assur, Z., Brasch, J., Shapiro, L., Mancia, F. & Hendrickson, W. A. (2012). Science, 336, 1033–1037.  Web of Science CrossRef CAS PubMed Google Scholar
First citationMcCarthy, A. A., Barrett, R., Beteva, A., Caserotto, H., Dobias, F., Felisaz, F., Giraud, T., Guijarro, M., Janocha, R., Khadrouche, A., Lentini, M., Leonard, G. A., Lopez Marrero, M., Malbet-Monaco, S., McSweeney, S., Nurizzo, D., Papp, G., Rossi, C., Sinoir, J., Sorez, C., Surr, J., Svensson, O., Zander, U., Cipriani, F., Theveneau, P. & Mueller-Dieckmann, C. (2018). J. Synchrotron Rad. 25, 1249–1260.  CrossRef CAS IUCr Journals Google Scholar
First citationMcCoy, A. J., Grosse-Kunstleve, R. W., Adams, P. D., Winn, M. D., Storoni, L. C. & Read, R. J. (2007). J. Appl. Cryst. 40, 658–674.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMonaco, S., Gordon, E., Bowler, M. W., Delagenière, S., Guijarro, M., Spruce, D., Svensson, O., McSweeney, S. M., McCarthy, A. A., Leonard, G. & Nanao, M. H. (2013). J. Appl. Cryst. 46, 804–810.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMueller, U., Darowski, N., Fuchs, M. R., Förster, R., Hellmig, M., Paithankar, K. S., Pühringer, S., Steffien, M., Zocher, G. & Weiss, M. S. (2012). J. Synchrotron Rad. 19, 442–449.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMueller, U., Förster, R., Hellmig, M., Huschmann, F. U., Kastner, A., Malecki, P., Pühringer, S., Röwer, M., Sparta, K., Steffien, M., Ühlein, M., Wilk, P. & Weiss, M. S. (2015). Eur. Phys. J. Plus, 130, 141.  Web of Science CrossRef Google Scholar
First citationMueller, U., Thunnissen, M., Nan, J., Eguiraun, M., Bolmsten, F., Milàn-Otero, A., Guijarro, M., Oscarsson, M., de Sanctis, D. & Leonard, G. (2017). Synchrotron Radiat. News. 30, 22–27.  CrossRef Google Scholar
First citationMueller-Dieckmann, C., Bowler, M. W., Carpentier, P., Flot, D., McCarthy, A. A., Nanao, M. H., Nurizzo, D., Pernot, P., Popov, A., Round, A., Royant, A., de Sanctis, D., von Stetten, D. & Leonard, G. A. (2015). Eur. Phys. J. Plus, 130, 70.  Google Scholar
First citationNurizzo, D., Mairs, T., Guijarro, M., Rey, V., Meyer, J., Fajardo, P., Chavanne, J., Biasci, J.-C., McSweeney, S. & Mitchell, E. (2006). J. Synchrotron Rad. 13, 227–238.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationPaithankar, K. S. & Garman, E. F. (2010). Acta Cryst. D66, 381–388.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationPascual-Izarra, C., Coutinho, T., Cuní, G., Falcón Torres, C., Fernández-Carreiras, D., Reszela, Z. & Rosanes Siscart, M. (2015). Proceedings of the 15th International Conference on Accelerator and Large Experimental Physics Control Systems (ICALEPCS 2015), 17–23 October 2015, Melbourne, Australia. THHC3O03.  Google Scholar
First citationPopov, A. N. & Bourenkov, G. (2016). DOZOR. ESRF, Grenoble, France.  Google Scholar
First citationReszela, Z., Becheri, F., Cuní, G., Falcon-Torres, C., Fernández-Carreiras, D., Homs-Puron, R., Moldes, J., Pascual-Izarra, C., Pastor Ortiz, R., Roldán, D. & Rosanes Siscart, M. (2017). Proceedings of the 16th International Conference on Accelerator and Large Experimental Control Systems (ICALEPCS2017). 8–13 October 2017, Barcelona, Spain, pp. 1067–1072. WEBPL06.  Google Scholar
First citationReszela, Z., Cuní, G., Fernández-Carreiras, D., Klora, J. & Pascual-Izarra, C. (2014). Proceedings of the 10th International Workshop on Personal Computer and Particle Accelerator Controls (PCaPAC 2014), 14–17 October 2014, Karlsruhe, Germany. WCO206.  Google Scholar
First citationRodríguez, D. D., Grosse, C., Himmel, S., González, C., de Ilarduya, I. M., Becker, S., Sheldrick, G. M. & Usón, I. (2009). Nat. Methods, 6, 651–653.  Web of Science PubMed Google Scholar
First citationSanctis, D. de, Beteva, A., Caserotto, H., Dobias, F., Gabadinho, J., Giraud, T., Gobbo, A., Guijarro, M., Lentini, M., Lavault, B., Mairs, T., McSweeney, S., Petitdemange, S., Rey-Bakaikoa, V., Surr, J., Theveneau, P., Leonard, G. A. & Mueller-Dieckmann, C. (2012). J. Synchrotron Rad. 19, 455–461.  Web of Science CrossRef IUCr Journals Google Scholar
First citationSanctis, D. de, Oscarsson, M., Popov, A., Svensson, O. & Leonard, G. (2016). Acta Cryst. D72, 413–420.  CrossRef IUCr Journals Google Scholar
First citationSauter, N. K. & Poon, B. K. (2010). J. Appl. Cryst. 43, 611–616.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationSchlichting, I. (2017). IUCrJ, 4, 516–517.  CrossRef CAS IUCr Journals Google Scholar
First citationSolé, V. A., Papillon, E., Cotte, M., Walter, P. & Susini, J. (2007). At. Spectrosc. 62, 63–68.  Google Scholar
First citationSparta, K. M., Krug, M., Heinemann, U., Mueller, U. & Weiss, M. S. (2016). J. Appl. Cryst. 49, 1085–1092.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationStandfuss, J. & Spence, J. (2017). IUCrJ, 4, 100–101.  Web of Science CrossRef CAS PubMed IUCr Journals Google Scholar
First citationSvensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757–1767.  Web of Science CrossRef IUCr Journals Google Scholar
First citationThunnissen, M. M. G. M., Sondhauss, P., Wallén, E., Theodor, K., Logan, D. T., Labrador, A., Unge, J., Appio, R., Fredslund, F. & Ursby, T. (2013). J. Phys. Conf. Ser. 425, 072012.  CrossRef Google Scholar
First citationVonrhein, C., Flensburg, C., Keller, P., Sharff, A., Smart, O., Paciorek, W., Womack, T. & Bricogne, G. (2011). Acta Cryst. D67, 293–302.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWeinert, T., Olieric, V., Waltersperger, S., Panepucci, E., Chen, L., Zhang, H., Zhou, D., Rose, J., Ebihara, A., Kuramitsu, S., Li, D., Howe, N., Schnapp, G., Pautsch, A., Bargsten, K., Prota, A. E., Surana, P., Kottur, J., Nair, D. T., Basilico, F., Cecatiello, V., Pasqualato, S., Boland, A., Weichenrieder, O., Wang, B.-C., Steinmetz, M. O., Caffrey, M. & Wang, M. (2015). Nat. Methods, 12, 131–133.  Web of Science CrossRef CAS PubMed Google Scholar
First citationWinter, G. (2010). J. Appl. Cryst. 43, 186–190.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationZander, U., Bourenkov, G., Popov, A. N., de Sanctis, D., Svensson, O., McCarthy, A. A., Round, E., Gordeliy, V., Mueller-Dieckmann, C. & Leonard, G. A. (2015). Acta Cryst. D71, 2328–2343.  Web of Science CrossRef IUCr Journals Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.