Multi-position data collection and dynamic beam sizing: recent improvements to the automatic data-collection algorithms on MASSIF-1

CROSSMARK_Color_square_no_text.svg

aEuropean Synchrotron Radiation Facility, 71 Avenue des Martyrs, CS 40220, 38043 Grenoble, France, and bEuropean Molecular Biology Laboratory, Grenoble Outstation, 71 Avenue des Martyrs, CS 90181, 38042 Grenoble, France
*Correspondence e-mail: mbowler@embl.fr

Edited by S. Wakatsuki, Stanford University, USA (Received 20 December 2017; accepted 3 March 2018; online 24 April 2018)

Macromolecular crystallography is now a mature and widely used technique that is essential in the understanding of biology and medicine. Increases in computing power combined with robotics have not only enabled large numbers of samples to be screened and characterized but have also enabled better decisions to be taken on data collection itself. This led to the development of MASSIF-1 at the ESRF, the first beamline in the world to run fully automatically while making intelligent decisions taking user requirements into account. Since opening in late 2014, the beamline has processed over 42 000 samples. Improvements have been made to the speed of the sample-handling robotics and error management within the software routines. The workflows initially put into place, while highly innovative at the time, have been expanded to include increased complexity and additional intelligence using the information gathered during characterization; this includes adapting the beam diameter dynamically to match the diffraction volume within the crystal. Complex multi-position and multi-crystal data collections have now also been integrated into the selection of experiments available. This has led to increased data quality and throughput, allowing even the most challenging samples to be treated automatically.

1. Introduction

Automation is transforming the way that scientific data are collected, allowing large amounts of high-quality data to be gathered in a consistent manner (Quintana & Plätzer, 2015[Quintana, F. J. & Plätzer, K. (2015). Science, 349, 1567.]; Foster, 2005[Foster, I. (2005). Science, 308, 814-817.]). Advances in robotics and software have been key to these developments and have had a particular impact on structural biology, allowing multiple constructs to be screened and purified (Camper & Viola, 2009[Camper, D. V. & Viola, R. E. (2009). Anal. Biochem. 393, 176-181.]; Hart & Waldo, 2013[Hart, D. J. & Waldo, G. S. (2013). Curr. Opin. Struct. Biol. 23, 403-408.]; Vijayachandran et al., 2011[Vijayachandran, L. S., Viola, C., Garzoni, F., Trowitzsch, S., Bieniossek, C., Chaillet, M., Schaffitzel, C., Busso, D., Romier, C., Poterszman, A., Richmond, T. J. & Berger, I. (2011). J. Struct. Biol. 175, 198-208.]); huge numbers of crystallization experiments to be performed (Elsliger et al., 2010[Elsliger, M.-A., Deacon, A. M., Godzik, A., Lesley, S. A., Wooley, J., Wüthrich, K. & Wilson, I. A. (2010). Acta Cryst. F66, 1137-1142.]; Ferrer et al., 2013[Ferrer, J. L., Larive, N. A., Bowler, M. W. & Nurizzo, D. (2013). Exp. Opin. Drug. Discov. 8, 835-847.]; Heinemann et al., 2003[Heinemann, U., Büssow, K., Mueller, U. & Umbach, P. (2003). Acc. Chem. Res. 36, 157-163.]; Joachimiak, 2009[Joachimiak, A. (2009). Curr. Opin. Struct. Biol. 19, 573-584.]; Calero et al., 2014[Calero, G., Cohen, A. E., Luft, J. R., Newman, J. & Snell, E. H. (2014). Acta Cryst. F70, 993-1008.]); samples to be mounted at synchrotrons (Cipriani et al., 2006[Cipriani, F. et al. (2006). Acta Cryst. D62, 1251-1259.]; Cohen et al., 2002[Cohen, A. E., Ellis, P. J., Miller, M. D., Deacon, A. M. & Phizackerley, R. P. (2002). J. Appl. Cryst. 35, 720-726.]; Jacquamet et al., 2009[Jacquamet, L., Joly, J., Bertoni, A., Charrault, P., Pirocchi, M., Vernede, X., Bouis, F., Borel, F., Périn, J.-P., Denis, T., Rechatin, J.-L. & Ferrer, J.-L. (2009). J. Synchrotron Rad. 16, 14-21.]; Nurizzo et al., 2016[Nurizzo, D., Bowler, M. W., Caserotto, H., Dobias, F., Giraud, T., Surr, J., Guichard, N., Papp, G., Guijarro, M., Mueller-Dieckmann, C., Flot, D., McSweeney, S., Cipriani, F., Theveneau, P. & Leonard, G. A. (2016). Acta Cryst. D72, 966-975.]; Papp et al., 2017[Papp, G., Felisaz, F., Sorez, C., Lopez-Marrero, M., Janocha, R., Manjasetty, B., Gobbo, A., Belrhali, H., Bowler, M. W. & Cipriani, F. (2017). Acta Cryst. D73, 841-851.]; Snell et al., 2004[Snell, G., Cork, C., Nordmeyer, R., Cornell, E., Meigs, G., Yegian, D., Jaklevic, J., Jin, J., Stevens, R. C. & Earnest, T. (2004). Structure, 12, 537-545.]); data to be analysed and processed (Bourenkov & Popov, 2010[Bourenkov, G. P. & Popov, A. N. (2010). Acta Cryst. D66, 409-419.]; Holton & Alber, 2004[Holton, J. & Alber, T. (2004). Proc. Natl Acad. Sci. USA, 101, 1537-1542.]; Incardona et al., 2009[Incardona, M.-F., Bourenkov, G. P., Levik, K., Pieritz, R. A., Popov, A. N. & Svensson, O. (2009). J. Synchrotron Rad. 16, 872-879.]; Leslie et al., 2002[Leslie, A. G. W., Powell, H. R., Winter, G., Svensson, O., Spruce, D., McSweeney, S., Love, D., Kinder, S., Duke, E. & Nave, C. (2002). Acta Cryst. D58, 1924-1928.]; Monaco et al., 2013[Monaco, S., Gordon, E., Bowler, M. W., Delagenière, S., Guijarro, M., Spruce, D., Svensson, O., McSweeney, S. M., McCarthy, A. A., Leonard, G. & Nanao, M. H. (2013). J. Appl. Cryst. 46, 804-810.]; Winter, 2010[Winter, G. (2010). J. Appl. Cryst. 43, 186-190.]); and the entire PDB to be validated (Joosten et al., 2012[Joosten, R. P., Joosten, K., Murshudov, G. N. & Perrakis, A. (2012). Acta Cryst. D68, 484-496.]). The combination of robotic sample mounting and online data analysis has been particularly important in macromolecular crystallography (MX) as it has allowed time to be saved, large numbers of samples to be screened, and enabled the remote operation of beamlines. However, despite these advances, a human presence is still required to sequence actions. Pioneering beamlines that have fully automated the process, such as LRL-CAT at the Advanced Photon Source (Wasserman et al., 2012[Wasserman, S. R., Koss, J. W., Sojitra, S. T., Morisco, L. L. & Burley, S. K. (2012). Trends Pharm. Sci. 33, 261-267.]) and the Stanford Synchrotron Radiation Lightsource MX beamlines (Tsai et al., 2013[Tsai, Y., McPhillips, S. E., González, A., McPhillips, T. M., Zinn, D., Cohen, A. E., Feese, M. D., Bushnell, D., Tiefenbrunn, T., Stout, C. D., Ludaescher, B., Hedman, B., Hodgson, K. O. & Soltis, S. M. (2013). Acta Cryst. D69, 796-803.]), removed the need for a human presence, but as they rely on optical loop centring this means that restrictions have to be placed on the size of the crystals and they tend to be robust, well diffracting samples, generally those for proprietary research in the pharmaceutical industry.

In 2014 the ESRF beamline MASSIF-1 (Bowler et al., 2015[Bowler, M. W. et al. (2015). J. Synchrotron Rad. 22, 1540-1547.]) opened to users as the first beamline to fully automate MX data collection, including sample-location and complex decision-making algorithms (Svensson et al., 2015[Svensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757-1767.]). The combination of sample centring using X-ray diffraction quality as a metric (X-ray centring) and characterization allows even the smallest and most weakly diffracting samples to be treated automatically. This opened full automation to any sample presented in any mount and has provided a new tool to structural biologists, allowing the process of collecting hundreds of data sets or screening hundreds of crystals to be `outsourced', freeing their time and often leading to the collection of better data (Bowler et al., 2016[Bowler, M. W., Svensson, O. & Nurizzo, D. (2016). Crystallogr. Rev. 22, 233-249.]). At the time of writing, the beamline has processed more than 42 000 samples representing a wide range of projects, from those that require extensive screening to find the best diffracting crystal (Na et al., 2017[Na, Z., Yeo, S. P., Bharath, S. R., Bowler, M. W., Balıkçı, E., Wang, C.-I. & Song, H. (2017). Cell Res. 27, 147-150.]; Sorigué et al., 2017[Sorigué, D. et al. (2017). Science, 357, 903-907.]; Naschberger et al. 2017[Naschberger, A., Orry, A., Lechner, S., Bowler, M. W., Nurizzo, D., Novokmet, M., Keller, M. A., Oemer, G., Seppi, D., Haslbeck, M., Pansi, K., Dieplinger, H. & Rupp, B. (2017). Structure, 25, 1907-1915.e5.]) to small-molecule fragment screening (Cheeseman et al., 2017[Cheeseman, M. D. et al. (2017). J. Med. Chem. 60, 180-201.]; Hiruma et al., 2017[Hiruma, Y., Koch, A., Hazraty, N., Tsakou, F., Medema, R. H., Joosten, R. P. & Perrakis, A. (2017). J. Biol. Chem. 292, 14496-14504.]) and experimental phasing at high and low resolutions (Kharde et al., 2015[Kharde, S., Calviño, F. R., Gumiero, A., Wild, K. & Sinning, I. (2015). Nucleic Acids Res. 43, 7083-7095.]; Muir et al., 2016[Muir, K. W., Kschonsak, M., Li, Y., Metz, J., Haering, C. H. & Panne, D. (2016). Cell. Rep. 14, 2116-2126.]; Sabbadin et al., 2018[Sabbadin, F. et al. (2018). Nature Commun. 9, 756.]). The beamline is able to deal with a wide range of samples by combining parameters provided by the user with information gathered during processing. The workflows that were initially put into place have performed very well, but many enhancements remained possible.

Here, we describe how the hardware and algorithms have been improved to increase the amount and the quality of data collected on MASSIF-1. The speed of the robotics has been increased and additional subroutines have been added to monitor and correct errors in centring and to account for low-resolution data collection as well as dynamically adjusting the beam diameter to match homogenous diffraction volumes. In combination with new multiple-position and multiple-crystal data-collection workflows, fully automatic data collection is now possible for the most challenging samples.

2. Experimental details and results

2.1. Hardware improvements

One of the most time-consuming, and important, steps in the X-ray centring process is the initial mesh scan that locates and characterizes the crystal. When first implemented on MASSIF-1, a rotation of ω was included in the scan in order to avoid still images that can cause problems in processing. This implementation required that triggering of the acquisition of images was instigated by the ω axis, meaning that each line of the mesh was treated as a separate data collection. The preparation required between data collections led to additional time being taken for the scan. We have now implemented a scan that includes no rotation of the ω axis and only requires movements of the high-precision Y/Z stage beneath the RoboDiff (Nurizzo et al., 2016[Nurizzo, D., Bowler, M. W., Caserotto, H., Dobias, F., Giraud, T., Surr, J., Guichard, N., Papp, G., Guijarro, M., Mueller-Dieckmann, C., Flot, D., McSweeney, S., Cipriani, F., Theveneau, P. & Leonard, G. A. (2016). Acta Cryst. D72, 966-975.]). This allows the triggering of acquisition to be made by the motor position, allowing the whole mesh to be launched as a single data collection (Aishima et al., 2010[Aishima, J., Owen, R. L., Axford, D., Shepherd, E., Winter, G., Levik, K., Gibbons, P., Ashton, A. & Evans, G. (2010). Acta Cryst. D66, 1032-1035.]). This method was implemented in 2017, and analysis of data for two-month periods before and after its implementation showed that the time required for these scans has been reduced by an average of 1 min (Fig. 1[link]).

[Figure 1]
Figure 1
The decrease in the time taken to perform the mesh scan after hardware improvements. Log-normal distribution of the elapsed time for mesh scans using the former protocol (black) and the new fast mesh (blue) for the two months preceding and following the implementation of the new protocol.

2.2. Dynamic beam sizing

One of the benefits of running a completely automated system is the ability to collect large amounts of data from samples and to use these data to improve strategies for data collection. We initially realized that the volumes of all crystals were determined during the X-ray centring routine, and this information was subsequently included in the strategy calculation, having the largest effect on the calculation of the maximum dose given to a crystal during data collection (Bowler et al., 2016[Bowler, M. W., Svensson, O. & Nurizzo, D. (2016). Crystallogr. Rev. 22, 233-249.]; Svensson et al., 2015[Svensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757-1767.]). Additionally, these measurements provided a distribution of crystal volumes, allowing us to use a default beam diameter of 50 µm, as this was the crystal dimension that was most frequently observed on the beamline. A specific beam diameter can be selected on a per-sample basis in the diffraction plan in ISPyB (Delagenière et al., 2011[Delagenière, S. et al. (2011). Bioinformatics, 27, 3186-3192.]); however, this option is usually used when users are sure that the crystal volumes are significantly smaller than the default beam diameter (Fig. 2[link]). Using the information gathered during the mesh scan, we can determine an optimized beam diameter. By accurately matching the beam diameter to the crystal, it has been shown that the background can be dramatically reduced (Moukhametzianov et al., 2008[Moukhametzianov, R., Burghammer, M., Edwards, P. C., Petitdemange, S., Popov, D., Fransen, M., McMullan, G., Schertler, G. F. X. & Riekel, C. (2008). Acta Cryst. D64, 158-166.]). This is particularly striking when the crystals are very small (Evans et al., 2011[Evans, G., Axford, D. & Owen, R. L. (2011). Acta Cryst. D67, 261-270.]), but if a crystal is large the additional diffraction power should not be wasted. This also has to be balanced with the degree of variability within each crystal (Bowler & Bowler, 2014[Bowler, M. G. & Bowler, M. W. (2014). Acta Cryst. F70, 127-132.]; Bowler et al., 2010[Bowler, M. W., Guijarro, M., Petitdemange, S., Baker, I., Svensson, O., Burghammer, M., Mueller-Dieckmann, C., Gordon, E. J., Flot, D., McSweeney, S. M. & Leonard, G. A. (2010). Acta Cryst. D66, 855-864.]; Pozharski, 2012[Pozharski, E. (2012). Acta Cryst. D68, 1077-1087.]).

[Figure 2]
Figure 2
Beam-diameter selections by users and the algorithm in 2017. The number of times a beam diameter was selected either by the user (black) or automatically (blue) is shown. 6877 data collections were performed with the beam diameter at the default value of 50 µm; as this is not changed by the software, the value is not shown.

We have now introduced a dynamic beam-diameter adjustment into all workflows running on MASSIF-1 where no value has been pre-selected by the user. The X-ray centring routine determines the crystal position relative to the beam and the crystal dimensions, and also determines the best homogenous volume within the crystal. The centre of mass of this volume is then used as the centring position and it is the dimensions of this volume that are used to select the beam diameter. There are five beam diameters available on MASSIF-1, 100, 50, 30, 15 and 10 µm (Bowler et al., 2015[Bowler, M. W. et al. (2015). J. Synchrotron Rad. 22, 1540-1547.]), and the smallest vertical volume dimension is used to select the aperture that matches most closely. In this way, the largest volume can be illuminated without increasing the background or `contaminating' the diffraction from variable areas. All steps during X-ray centring are performed with the 50 µm aperture. As the scans are performed with an overlap, the smallest dimension that can be measured is ∼20 µm, meaning that the 10 µm option is only used when selected by the user. Once X-ray centring is completed, the new aperture diameter is selected and characterization images are collected using this diameter. The strategy calculation will include the new flux for the aperture as well as the crystal volume determined during the centring procedure. Since introducing the adaptable beam diameter the system has selected the 30 µm beam size most frequently (Fig. 2[link]), followed by the 100 and 15 µm diameter apertures. Half of all data collections are performed with a diameter of 50 µm, reinforcing its choice as the default value.

Can the advantages of dynamic beam adjustment be demonstrated in a consistent manner? It is always problematic to clearly show that one data-collection method is better than another. However, here we show that dynamic beam sizing makes a significant difference for weakly diffracting samples.

We initially tested the adaptive beam-diameter protocol on crystals of the β1-adrenergic GPCR (Warne et al., 2008[Warne, T., Serrano-Vega, M. J., Baker, J. G., Moukhametzianov, R., Edwards, P. C., Henderson, R., Leslie, A. G. W, Tate, C. G. & Schertler, G. F. (2008). Nature (London), 454, 486-491.]). These crystals diffract weakly, exhibit considerable variation in diffraction quality and tend to form as thin plates or needles. A total of 30 crystals were run on MASSIF-1, first using the classic protocol (Svensson et al., 2015[Svensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757-1767.]), where a 50 µm beam diameter is the default, and then running a second protocol on the same crystal including the adaptive beam diameter. In many cases data sets were collected from the same crystal using both procedures; however, for some cases data sets were only processed where the beam diameter had been reduced to match the crystal size. Table 1[link] shows crystal dimensions and data-processing statistics for crystals where an automatically processed data set was produced (eight out of 30 crystals). Where crystals were of sufficient quality, data sets were mostly produced using both protocols, but it is for the smaller crystals that a difference is discernible. For crystals with a y dimension below 30 µm the data sets produced have a higher 〈I/σ(I)〉 or resolution limit (Table 1[link]; adrcpt-For42, adrcpt-For48, adrcpt-For59 and adrcpt-For67) even though the crystals had already been exposed. For one of these crystals, adrcpt-For42, the data set collected with the smaller beam is significantly better (Table 2[link]).

Table 1
Data-collection details for a β1-andrenergic GPCR crystal using standard and adapted beam-diameter protocols

The dimensions are x, the measured crystal length parallel to the spindle axis, y, the height orthogonal to the spindle axis, and z, the depth orthogonal to the spindle axis 90° away in ω.

    Fixed beam diameter Adaptable beam diameter
Crystal Crystal dimensions (x × y × z) (mm) Resolution limit (Å) I/σ(I)〉 Resolution limit (Å) I/σ(I)〉
adrcpt-For41 0.109 × 0.053 × 0.025 3.77 6.7 4.13 4.4
adrcpt-For42 0.084 × 0.025 × 0.025 4.22 4.3 3.53 10.6
adrcpt-For45 0.035 × 0.045 × 0.051 3.95 6.2
adrcpt-For47 0.105 × 0.061 × 0.051 3.74 4.7 3.72 5.4
adrcpt-For48 0.105 × 0.039 × 0.064 3.80 5.7
adrcpt-For58 0.169 × 0.050 × 0.061 3.88 6.6 4.11 4.5
adrcpt-For59 0.042 × 0.024 × 0.025 3.25 9.2 3.16 8.3
adrcpt-For67 0.064 × 0.026 × 0.031 3.80 5.6

Table 2
Comparison of data-collection strategies and processing statistics for standard and adaptive beam-diameter protocols using a β1-andrenergic GPCR crystal

Values in parentheses are for the outer shell.

Crystal adrcpt-For42 adrcpt-For42
Beam diameter (µm) 50 30
Space group P212121 [No. 19] P212121 [No. 19]
Unit-cell parameters (Å, °) a = 116.8, b = 121.1, c = 129.5, α = β = γ = 90 a = 116.5, b = 120.78, c = 128.7 α = β = γ = 90
Flux (photons s−1) 5.7 × 1011 2.2 × 1011
Rotation width (°) 0.15 0.05
Total oscillation range (°) 149.1 124.0
Total dose (MGy) 5.25 5.92
Detector resolution 4.05 3.9
Wilson B factor (Å2) 94.0 97.0
Resolution (Å) 48.2–3.95 (4.09–3.95) 48.7–3.53 (3.66–3.53)
Completeness (%) 98.2 (90.9) 81.5 (33.0)
Observed reflections 15940 (1413) 18785 (721)
Average multiplicity 4.1 (2.7) 4.4 (1.6)
I/σ(I)〉 4.3 (0.8) 10.6 (0.6)
Rmeas 0.23 (1.45) 0.1 (1.64)
Rmerge 0.18 (1.05) 0.077 (1.17)
CC1/2 0.99 (0.43) 1 (0.33)

What effect has the protocol had on overall data collection? In order to analyse the difference, we looked at the average signal-to-noise ratios, 〈I/σ(I)〉, for all data sets processed automatically (Monaco et al., 2013[Monaco, S., Gordon, E., Bowler, M. W., Delagenière, S., Guijarro, M., Spruce, D., Svensson, O., McSweeney, S. M., McCarthy, A. A., Leonard, G. & Nanao, M. H. (2013). J. Appl. Cryst. 46, 804-810.]; Vonrhein et al., 2011[Vonrhein, C., Flensburg, C., Keller, P., Sharff, A., Smart, O., Paciorek, W., Womack, T. & Bricogne, G. (2011). Acta Cryst. D67, 293-302.]; Sparta et al., 2016[Sparta, K. M., Krug, M., Heinemann, U., Mueller, U. & Weiss, M. S. (2016). J. Appl. Cryst. 49, 1085-1092.]; Winter, 2010[Winter, G. (2010). J. Appl. Cryst. 43, 186-190.]) in the year preceding and following the introduction of the protocol. This amounts to data for approximately 22 000 samples. Fig. 3[link] shows the distributions of overall 〈I/σ(I)〉 for the data sets. While the distributions are similar for high 〈I/σ(I)〉 they diverge at lower values, with a significant shift lower for the adaptive beam-diameter data sets. The average before the procedure was put in place was 14.4 and it decreased to 12.2 after, with modal values of 8.55 before and 5.9 after (Fig. 3[link]). We initially found this surprising as we had expected a general increase in 〈I/σ(I)〉. However, given the effect seen on the GPCR crystals, the distribution change is understandable. While the beam diameter is increased or decreased to match the diffraction volume, the 〈I/σ(I)〉 values for strongly diffracting crystals will remain the same given the dose to achieve a certain resolution without radiation damage, taking the crystal volume and changed flux into account. However, it was for the weakly diffracting crystals that the adaptive beam diameter had the most significant effect. Whether the diameter was increased or decreased, there was a large shift in the number of data sets that were processed that had rather low 〈I/σ(I)〉 after the adaptive beam-diameter protocol was implemented. This implies that by introducing this routine into the regular data-collection workflow, the beamline is able to increase the number of data sets processed from these samples by reducing the background noise.

[Figure 3]
Figure 3
Distribution of overall 〈I/σ(I)〉 for data sets processed in the year preceding implementation of the dynamic beam aperture (black) and the year after implementation of the dynamic beam aperture (blue). There was a significant shift in the number of data sets processed with lower 〈I/σ(I)〉 after dynamic beam-diameter adjustment was introduced.

2.3. Improved error handling

The correct handling of errors is paramount in an automated system. We initially introduced many error-handling routines at both a high level, such as the collection of a data set with default settings when indexing fails (Svensson et al., 2015[Svensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757-1767.]), and a low level, such as escaping from small robotic errors (Nurizzo et al., 2016[Nurizzo, D., Bowler, M. W., Caserotto, H., Dobias, F., Giraud, T., Surr, J., Guichard, N., Papp, G., Guijarro, M., Mueller-Dieckmann, C., Flot, D., McSweeney, S., Cipriani, F., Theveneau, P. & Leonard, G. A. (2016). Acta Cryst. D72, 966-975.]). After processing more than 42 000 samples we have now been able to observe most errors that may be encountered and have extended the processes to catch them.

2.3.1. Centring errors

We have occasionally observed that after the X-ray centring routine the crystal was still not correctly aligned over the full rotation range. This may be due to movements of the support after the routine has completed or to errors in the routine arising from multiple peaks being selected for centring. Whatever the reason, it can lead to a data set being lost. We therefore introduced a check in the characterization step that ensures that the four images have a diffraction signal. If one image has a signal that is below 10% of the top signal, a recovery routine is launched. This involves three short line scans (50 µm above and below the current centred position) being launched over the currently centred position. In most cases it corrects the error. In 2017, 13 776 samples were processed on MASSIF-1 and centring recovery was launched 221 times. This represents a centring-error rate of 1.6%, which should be reduced further by being able to detect and recover incorrectly centred samples.

2.3.2. Low-resolution data collection

Calculating an optimized strategy for low-resolution (>4 Å) data collection can be problematic, but is very important as useful information can be extracted from carefully collected data at these resolutions. Here, we used information from low-resolution data collections that had already been performed to improve how the routines deal with these samples. Unless a resolution is specified by the user, all mesh and characterization images are collected at 2 Å. If the predicted resolution extends beyond the corners of the detector (1.42 Å) the detector is moved to the new resolution and a further characterization is launched. This allows the highest possible resolution to be obtained and ensures that characterization is performed at an optimal detector distance (Svensson et al., 2015[Svensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757-1767.]). However, for low resolution, the resolution determined by BEST is used and data collection proceeds according to the determined strategy. It would seem sensible that if very low diffraction resolution is determined, the characterization images should also be re-collected at this resolution. We therefore introduced a routine to re-collect the characterization images at 4 Å resolution for all samples where the determined resolution is below this value. This allows the distribution of intensity to be better estimated and should lead to better strategy calculations (Popov & Bourenkov, 2003[Popov, A. N. & Bourenkov, G. P. (2003). Acta Cryst. D59, 1145-1153.]).

By analysing the relationship between the predicted and the determined resolution from all data sets collected so far we can also try to improve the quality of the data collected on MASSIF-1 (Fig. 4[link]). The distribution shows that the agreement is excellent, usually slightly underestimating the achievable resolution. This may well be due to the difference in the criterion for resolution-limit determination, which is 〈I/σ(I)〉 for characterization and CC1/2 for complete data sets. A clear trend is that the strategy tends to underestimate the resolution for weakly diffracting crystals (Fig. 4[link]). This is due to the difficulty in estimating a B factor at very low resolution. In addition to re-collecting the characterization images, the procedure now always sets the detector resolution to 4 Å for all data collections where the predicted resolution is lower than this value. In this way, we hope that higher resolution data will not be missed, if possible, as complete data to 4 Å resolution are more important than suboptimal data collection at 7 Å resolution.

[Figure 4]
Figure 4
Scatter plot of predicted resolutions for data collections against the resolutions determined by autoprocessing for all crystals processed so far on MASSIF-1 that resulted in a strategy and an automatically processed data set. The red line shows perfect agreement between predicted and achieved resolution and the gradient shows the density of data points.

2.4. Multiple-crystal and multiple-position data-collection strategies

The possibility to input a number of positions from which to collect data was introduced into the diffraction plan early in the operation of MASSIF-1 (Bowler et al., 2016[Bowler, M. W., Svensson, O. & Nurizzo, D. (2016). Crystallogr. Rev. 22, 233-249.]; Svensson et al., 2015[Svensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757-1767.]). This has allowed complete data sets to be collected either from separate crystals contained on a sample support or from multiple positions within a single crystal and has proved to be a popular option, with many samples received on MASSIF-1 having between two and 12 positions requested (Fig. 5[link]). While extremely useful, this protocol does not cover the scenario where radiation-sensitive samples can benefit from a large dose being spread over multiple partial data sets, a procedure known generally as helical data collection (Flot et al., 2010[Flot, D., Mairs, T., Giraud, T., Guijarro, M., Lesourd, M., Rey, V., van Brussel, D., Morawe, C., Borel, C., Hignette, O., Chavanne, J., Nurizzo, D., McSweeney, S. & Mitchell, E. (2010). J. Synchrotron Rad. 17, 107-118.]) and that has been shown to be beneficial in many cases (Polsinelli et al., 2017[Polsinelli, I., Savko, M., Rouanet-Mehouas, C., Ciccone, L., Nencetti, S., Orlandini, E., Stura, E. A. & Shepard, W. (2017). J. Synchrotron Rad. 24, 42-52.]). Radiation damage can often make it difficult to collect complete data, or data with sufficient anomalous signal, from a single crystal or a single position within a crystal. A new experiment type is now available on MASSIF-1 that will automatically collect multiple partial data sets from positions within a homogenous volume of a crystal. This can lead to improved data quality, increased resolution and higher anomalous peaks. This is the first fully automated helical data-collection protocol that also accounts for the heterogeneity of crystal diffraction quality.

[Figure 5]
Figure 5
Number of positions selected by users for multiple-crystal and multiple-position data collections in 2017. Multiple-position data collections were requested for 9% of samples.

Multiple-position/crystal experiments are selected either by specifying a number of positions in the diffraction plan for the sample or by requesting MXPressP or MXPressP_SAD (for a pseudohelical strategy or a pseudohelical with SAD strategy, respectively) in the `experiment type' field for the required samples in ISPyB (Delagenière et al., 2011[Delagenière, S. et al. (2011). Bioinformatics, 27, 3186-3192.]). These new experiments operate in much the same way as the usual automated workflows on MASSIF-1 in that all of the current features are retained, such as resolution selection, strategy input, diffraction-volume calculation and smart beam sizing. If multiple positions are selected, the automesh algorithm, which determines the area to scan to locate the crystal (Svensson et al., 2015[Svensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757-1767.]), uses the widest orientation of the sample support, rather than the smallest, in order to avoid overlapping crystals or positions in ω (Fig. 6a[link]). After the mesh scan is complete, the map is analysed either for the number of peaks requested or, if no positions are specified for MXPressP, a default value of five. In the original procedure, a diffraction volume was selected by applying a threshold (50% of the maximum) to the diffraction signal and determining a centre of mass over connected regions (Svensson et al., 2015[Svensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757-1767.]). For multi-position experiments the same thresholding is applied but is iterated over the whole crystal volume using the next highest point to start after each volume has been defined (Fig. 6[link]). The determined volumes must be within 10% of the value of the peak for multiple-crystal data collection or 70% of the second highest peak for pseudohelical data collection. Additionally, any peaks that are closer than a beam diameter together or that will overlap in ω are eliminated. The number of allowed detected volumes is then specified by a comment in ISPyB. If multiple crystals have been selected, each point is then centred as usual and a complete data set is collected according to user input requirements. If MXPressP is selected, the top volume is centred and four characterization images are collected. A strategy is then calculated for a complete data set and the data are collected; for MXPressP_SAD the strategy is calculated for structure solution by SAD (Svensson et al., 2015[Svensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757-1767.]). As usual, in the case of indexing failure a default data collection of 180° is collected (240° for triclinic and 360° for SAD data collection). Once completed, a strategy is then calculated to collect a complete data set from the N positions determined in the mesh scan that are within 70% of the value of position 2. The strategy, as usual, accounts for the volume of the positions, beam diameter etc. Again, in the case of a failure in indexing default data collections are performed at each position using a full dose and a rotation range determined by 180°/N (240°/N for triclinic and 360°/N for SAD). Each partial data set has a 5° overlap with the next to assist with scaling.

[Figure 6]
Figure 6
Multiple-crystal and multiple-position data collection. (a) automesh scan of a CrystalDirect (Zander, Hoffmann et al., 2016[Zander, U., Hoffmann, G. et al. (2016). Acta Cryst. D72, 454-466.]) support that has three crystals mounted. The widest orientation of the mount was selected. (b) Mesh scan of the mount shown in (a). Three positions were requested and three were detected. (c) Mesh scan for a βPGM crystal where a native pseudohelical data collection was requested, five positions were detected and a beam diameter of 30 µm was selected. (d) Mesh scan for an FAE crystal where a SAD pseudohelical data collection was requested, five positions were detected and a beam diameter of 100 µm was selected.

We are, for the moment, remaining cautious with pseudohelical data collection by collecting a full single-position complete data set from the best volume. The reason for this is twofold: (i) we have observed that crystal heterogeneity can often lead to a number of the partial data sets being of varying quality despite the stringent quality threshold that we have implemented and (ii) we are eager to compile a large amount of data on how and when helical data collection is superior to single-position data collection. This is extremely important as, so far, the few studies on helical data collection have not considered crystal heterogeneity (Bowler & Bowler, 2014[Bowler, M. G. & Bowler, M. W. (2014). Acta Cryst. F70, 127-132.]; Bowler et al., 2010[Bowler, M. W., Guijarro, M., Petitdemange, S., Baker, I., Svensson, O., Burghammer, M., Mueller-Dieckmann, C., Gordon, E. J., Flot, D., McSweeney, S. M. & Leonard, G. A. (2010). Acta Cryst. D66, 855-864.]). Strategy parameters and data-processing statistics for two example systems using the pseudohelical routines for native and SAD data collections are shown in Table 3[link]. Two proteins that tend to form crystals with a needle morphology were selected: β-phosphoglucomutase (βPGM) in an open conformation (Baxter et al., 2010[Baxter, N. J., Bowler, M. W., Alizadeh, T., Cliff, M. J., Hounslow, A. M., Wu, B., Berkowitz, D. B., Williams, N. H., Blackburn, G. M. & Waltho, J. P. (2010). Proc. Natl Acad. Sci. USA, 107, 4555-4560.]) (Fig. 6c[link]) and ferulic acid esterase (FAE, Fig. 6d[link]), which contains eight Se atoms and five Cd2+ ions (Prates et al., 2001[Prates, J. A. M., Tarbouriech, N., Charnock, S. J., Fontes, C. M. G. A., Ferreira, L. & Davies, G. J. (2001). Structure, 9, 1183-1190.]) with a significant anomalous signal at the MASSIF-1 wavelength of 0.966 Å. Comparing the single-position data collection with the merged multiple-position data sets shows that in these cases there is not a significant increase in data quality. However, in the SAD case the helical data set has considerably higher 〈I/σ(I)〉, anomalous correlation co­efficients and mid-slope of anomalous probability than the single-position data set. For the native data sets, the single position is slightly better. This may reflect the heterogeneity within the crystal and highlights the importance of this parameter in whether to select helical versus single position for a certain project. The ability to automatically run clustering algorithms (Giordano et al., 2012[Giordano, R., Leal, R. M. F., Bourenkov, G. P., McSweeney, S. & Popov, A. N. (2012). Acta Cryst. D68, 649-658.]; Zander, Cianci et al., 2016[Zander, U., Cianci, M., Foos, N., Silva, C. S., Mazzei, L., Zubieta, C., de Maria, A. & Nanao, M. H. (2016). Acta Cryst. D72, 1026-1035.]) on these partial data sets may also improve the quality of the final data. We hope that by being able to analyse the variation in diffraction quality, and compare single-position data with multi-position data from the same crystal, a more general strategy for these types of data collection may emerge.

Table 3
Pseudohelical data collection

Comparison of data sets collected from a single position and from multiple positions for native and SAD strategies. Values in parentheses are for the outer shell.

  MXPressP MXPressP_SAD
  Single position Multi-position Single position Multi-position
Protein βPGM βPGM FAE FAE
Crystal dimensions (x × y × z) (µm) 211 × 45 × 61 211 × 45 × 61 485 × 117 × 91 485 × 117 × 91
Space group P212121 P212121 P41212 P41212
Unit-cell parameters (Å, °) a = 52.7, b = 53.9, c = 81.5, α = β = γ = 90 a = 52.7, b = 53.9, c = 81.5, α = β = γ = 90 a = b = 111.4, c = 65.5, α = β = γ = 90 a = b = 111.4, c = 65.5, α = β = γ = 90
No. of positions 1 4 1 4
Beam diameter (µm) 30 30 100 100
Flux (photons s−1) 3.3 × 1011 3.3 × 1011 1.2 × 1012 1.2 × 1012
Transmission (%) 100 100 100 100
Dose per position (MGy) 8.23 5.8 16.53 5.85
Total dose (MGy) 8.23 23.2 16.53 23.40
Total exposure time (s) 115.8 326.8 275 390
Rotation width (°) 0.1 0.1 0.1 0.1
Total oscillation range (°) 125 120 [30 × 4] 360 360 [90 × 4]
Detector resolution (Å) 1.7 1.61 2.62 2.08
Resolution range (Å) 44.5–1.80 44.8–1.80 42.5–2.50 42.5–2.50
Rmeas 0.092 (0.94) 0.104 (0.74) 0.103 (0.974) 0.071 (0.522)
I/σ(I)〉 11.0 (1.6) 9.8 (1.3) 26.2 (5.2) 35.2 (8.6)
CC1/2 0.99 (0.57) 0.99 (0.77) 0.99 (0.93) 0.99 (0.98)
Multiplicity 4.2 (4.1) 4.3 (4.5) 24.7 (26.3) 26.2 (28.2)
Mid-slope of anomalous normal probability 1.49 1.80
Anomalous completeness (%) 100 (99.3) 100 (100)
CCano 0.69 (0.07) 0.82 (0.13)

3. Discussion

The results presented here demonstrate not only the increase in the speed and reliability of automatic data collections but also that more complex strategies can be brought into the arena of autonomous experiments. Automation is often seen as a way to deal with mundane experiments that require little human input. The autonomous system presented here is different in that in addition to automating mounting and centring, it also uses data gathered during the process to improve data-collection strategies. We have already demonstrated that MASSIF-1 collects, on average, better quality data than humans are able to (Bowler et al., 2016[Bowler, M. W., Svensson, O. & Nurizzo, D. (2016). Crystallogr. Rev. 22, 233-249.]). The additional routines presented here add even more expert knowledge into the system that should further enhance its ability to extract the best possible data from every sample. This built-in expert knowledge means that the system is excellent not only for robust and routine data collections but also for challenging systems that diffract weakly. We have demonstrated that adapting the beam diameter can increase the number of data sets that can be processed from these types of sample. We hope that by providing more data on more samples we can improve feedback into experiment cycles and increase the amount of useful data produced.

All of the developments described here have been exported to the human-operated ESRF beamlines (Mueller-Dieckmann et al., 2015[Mueller-Dieckmann, C., Bowler, M., Carpentier, P., Flot, D., McCarthy, A., Nanao, M., Nurizzo, D., Pernot, P., Popov, A., Round, A., Royant, A., de Sanctis, D., von Stetten, D. & Leonard, G. (2015). Eur. Phys. J. Plus, 130, 70.]). As structural biologists now turn to an ever wider variety of tech­niques, we hope that fully automatic data collection will become the standard data-collection method for MX as the best possible data can be collected from samples, be they large and robust or small and weakly diffracting. In combination with developments in the robotic mounting and soaking of crystals (Zander, Hoffmann et al., 2016[Zander, U., Hoffmann, G. et al. (2016). Acta Cryst. D72, 454-466.]), we envision that the future of macromolecular crystallo­graphy is the provision of a fully automated high-throughput service that is able to rapidly produce high-quality structural models and to screen for potential therapeutic and probe molecules.

Acknowledgements

We thank Tony Warne (MRC-LMB, Cambridge, England) for the gift of β1-adrenergic GPCR crystals; Josan Marquez and Guillaume Hoffman (EMBL, Grenoble, France) for providing CrystalDirect supports with multiple crystals; and the ESRF Automation Task Force (ATF) and the ESRF–EMBL Joint Structural Biology Group (JSBG) for supporting the developments. We are also grateful to Gordon Leonard (ESRF) for comments on the manuscript.

References

First citationAishima, J., Owen, R. L., Axford, D., Shepherd, E., Winter, G., Levik, K., Gibbons, P., Ashton, A. & Evans, G. (2010). Acta Cryst. D66, 1032–1035.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBaxter, N. J., Bowler, M. W., Alizadeh, T., Cliff, M. J., Hounslow, A. M., Wu, B., Berkowitz, D. B., Williams, N. H., Blackburn, G. M. & Waltho, J. P. (2010). Proc. Natl Acad. Sci. USA, 107, 4555–4560.  Web of Science CrossRef CAS PubMed Google Scholar
First citationBourenkov, G. P. & Popov, A. N. (2010). Acta Cryst. D66, 409–419.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBowler, M. G. & Bowler, M. W. (2014). Acta Cryst. F70, 127–132.  Web of Science CrossRef IUCr Journals Google Scholar
First citationBowler, M. W., Guijarro, M., Petitdemange, S., Baker, I., Svensson, O., Burghammer, M., Mueller-Dieckmann, C., Gordon, E. J., Flot, D., McSweeney, S. M. & Leonard, G. A. (2010). Acta Cryst. D66, 855–864.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationBowler, M. W. et al. (2015). J. Synchrotron Rad. 22, 1540–1547.  Web of Science CrossRef IUCr Journals Google Scholar
First citationBowler, M. W., Svensson, O. & Nurizzo, D. (2016). Crystallogr. Rev. 22, 233–249.  Web of Science CrossRef CAS Google Scholar
First citationCalero, G., Cohen, A. E., Luft, J. R., Newman, J. & Snell, E. H. (2014). Acta Cryst. F70, 993–1008.  Web of Science CrossRef IUCr Journals Google Scholar
First citationCamper, D. V. & Viola, R. E. (2009). Anal. Biochem. 393, 176–181.  CrossRef CAS Google Scholar
First citationCheeseman, M. D. et al. (2017). J. Med. Chem. 60, 180–201.  CrossRef CAS Google Scholar
First citationCipriani, F. et al. (2006). Acta Cryst. D62, 1251–1259.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationCohen, A. E., Ellis, P. J., Miller, M. D., Deacon, A. M. & Phizackerley, R. P. (2002). J. Appl. Cryst. 35, 720–726.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationDelagenière, S. et al. (2011). Bioinformatics, 27, 3186–3192.  Web of Science PubMed Google Scholar
First citationElsliger, M.-A., Deacon, A. M., Godzik, A., Lesley, S. A., Wooley, J., Wüthrich, K. & Wilson, I. A. (2010). Acta Cryst. F66, 1137–1142.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationEvans, G., Axford, D. & Owen, R. L. (2011). Acta Cryst. D67, 261–270.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationFerrer, J. L., Larive, N. A., Bowler, M. W. & Nurizzo, D. (2013). Exp. Opin. Drug. Discov. 8, 835–847.  Web of Science CrossRef CAS Google Scholar
First citationFlot, D., Mairs, T., Giraud, T., Guijarro, M., Lesourd, M., Rey, V., van Brussel, D., Morawe, C., Borel, C., Hignette, O., Chavanne, J., Nurizzo, D., McSweeney, S. & Mitchell, E. (2010). J. Synchrotron Rad. 17, 107–118.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationFoster, I. (2005). Science, 308, 814–817.  Web of Science CrossRef PubMed CAS Google Scholar
First citationGiordano, R., Leal, R. M. F., Bourenkov, G. P., McSweeney, S. & Popov, A. N. (2012). Acta Cryst. D68, 649–658.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationHart, D. J. & Waldo, G. S. (2013). Curr. Opin. Struct. Biol. 23, 403–408.  Web of Science CrossRef CAS PubMed Google Scholar
First citationHeinemann, U., Büssow, K., Mueller, U. & Umbach, P. (2003). Acc. Chem. Res. 36, 157–163.  Web of Science CrossRef PubMed CAS Google Scholar
First citationHiruma, Y., Koch, A., Hazraty, N., Tsakou, F., Medema, R. H., Joosten, R. P. & Perrakis, A. (2017). J. Biol. Chem. 292, 14496–14504.  CrossRef CAS Google Scholar
First citationHolton, J. & Alber, T. (2004). Proc. Natl Acad. Sci. USA, 101, 1537–1542.  Web of Science CrossRef PubMed CAS Google Scholar
First citationIncardona, M.-F., Bourenkov, G. P., Levik, K., Pieritz, R. A., Popov, A. N. & Svensson, O. (2009). J. Synchrotron Rad. 16, 872–879.  Web of Science CrossRef IUCr Journals Google Scholar
First citationJacquamet, L., Joly, J., Bertoni, A., Charrault, P., Pirocchi, M., Vernede, X., Bouis, F., Borel, F., Périn, J.-P., Denis, T., Rechatin, J.-L. & Ferrer, J.-L. (2009). J. Synchrotron Rad. 16, 14–21.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationJoachimiak, A. (2009). Curr. Opin. Struct. Biol. 19, 573–584.  Web of Science CrossRef PubMed CAS Google Scholar
First citationJoosten, R. P., Joosten, K., Murshudov, G. N. & Perrakis, A. (2012). Acta Cryst. D68, 484–496.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationKharde, S., Calviño, F. R., Gumiero, A., Wild, K. & Sinning, I. (2015). Nucleic Acids Res. 43, 7083–7095.  Web of Science CrossRef CAS PubMed Google Scholar
First citationLeslie, A. G. W., Powell, H. R., Winter, G., Svensson, O., Spruce, D., McSweeney, S., Love, D., Kinder, S., Duke, E. & Nave, C. (2002). Acta Cryst. D58, 1924–1928.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMonaco, S., Gordon, E., Bowler, M. W., Delagenière, S., Guijarro, M., Spruce, D., Svensson, O., McSweeney, S. M., McCarthy, A. A., Leonard, G. & Nanao, M. H. (2013). J. Appl. Cryst. 46, 804–810.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMoukhametzianov, R., Burghammer, M., Edwards, P. C., Petitdemange, S., Popov, D., Fransen, M., McMullan, G., Schertler, G. F. X. & Riekel, C. (2008). Acta Cryst. D64, 158–166.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationMueller-Dieckmann, C., Bowler, M., Carpentier, P., Flot, D., McCarthy, A., Nanao, M., Nurizzo, D., Pernot, P., Popov, A., Round, A., Royant, A., de Sanctis, D., von Stetten, D. & Leonard, G. (2015). Eur. Phys. J. Plus, 130, 70.  Google Scholar
First citationMuir, K. W., Kschonsak, M., Li, Y., Metz, J., Haering, C. H. & Panne, D. (2016). Cell. Rep. 14, 2116–2126.  Web of Science CrossRef CAS PubMed Google Scholar
First citationNa, Z., Yeo, S. P., Bharath, S. R., Bowler, M. W., Balıkçı, E., Wang, C.-I. & Song, H. (2017). Cell Res. 27, 147–150.  CrossRef CAS Google Scholar
First citationNaschberger, A., Orry, A., Lechner, S., Bowler, M. W., Nurizzo, D., Novokmet, M., Keller, M. A., Oemer, G., Seppi, D., Haslbeck, M., Pansi, K., Dieplinger, H. & Rupp, B. (2017). Structure, 25, 1907–1915.e5.  CrossRef CAS Google Scholar
First citationNurizzo, D., Bowler, M. W., Caserotto, H., Dobias, F., Giraud, T., Surr, J., Guichard, N., Papp, G., Guijarro, M., Mueller-Dieckmann, C., Flot, D., McSweeney, S., Cipriani, F., Theveneau, P. & Leonard, G. A. (2016). Acta Cryst. D72, 966–975.  Web of Science CrossRef IUCr Journals Google Scholar
First citationPapp, G., Felisaz, F., Sorez, C., Lopez-Marrero, M., Janocha, R., Manjasetty, B., Gobbo, A., Belrhali, H., Bowler, M. W. & Cipriani, F. (2017). Acta Cryst. D73, 841–851.  CrossRef IUCr Journals Google Scholar
First citationPolsinelli, I., Savko, M., Rouanet-Mehouas, C., Ciccone, L., Nencetti, S., Orlandini, E., Stura, E. A. & Shepard, W. (2017). J. Synchrotron Rad. 24, 42–52.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationPopov, A. N. & Bourenkov, G. P. (2003). Acta Cryst. D59, 1145–1153.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationPozharski, E. (2012). Acta Cryst. D68, 1077–1087.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationPrates, J. A. M., Tarbouriech, N., Charnock, S. J., Fontes, C. M. G. A., Ferreira, L. & Davies, G. J. (2001). Structure, 9, 1183–1190.  CrossRef CAS Google Scholar
First citationQuintana, F. J. & Plätzer, K. (2015). Science, 349, 1567.  CrossRef Google Scholar
First citationSabbadin, F. et al. (2018). Nature Commun. 9, 756.  CrossRef Google Scholar
First citationSnell, G., Cork, C., Nordmeyer, R., Cornell, E., Meigs, G., Yegian, D., Jaklevic, J., Jin, J., Stevens, R. C. & Earnest, T. (2004). Structure, 12, 537–545.  Web of Science CrossRef PubMed CAS Google Scholar
First citationSorigué, D. et al. (2017). Science, 357, 903–907.  Google Scholar
First citationSparta, K. M., Krug, M., Heinemann, U., Mueller, U. & Weiss, M. S. (2016). J. Appl. Cryst. 49, 1085–1092.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationSvensson, O., Malbet-Monaco, S., Popov, A., Nurizzo, D. & Bowler, M. W. (2015). Acta Cryst. D71, 1757–1767.  Web of Science CrossRef IUCr Journals Google Scholar
First citationTsai, Y., McPhillips, S. E., González, A., McPhillips, T. M., Zinn, D., Cohen, A. E., Feese, M. D., Bushnell, D., Tiefenbrunn, T., Stout, C. D., Ludaescher, B., Hedman, B., Hodgson, K. O. & Soltis, S. M. (2013). Acta Cryst. D69, 796–803.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationVijayachandran, L. S., Viola, C., Garzoni, F., Trowitzsch, S., Bieniossek, C., Chaillet, M., Schaffitzel, C., Busso, D., Romier, C., Poterszman, A., Richmond, T. J. & Berger, I. (2011). J. Struct. Biol. 175, 198–208.  CrossRef CAS Google Scholar
First citationVonrhein, C., Flensburg, C., Keller, P., Sharff, A., Smart, O., Paciorek, W., Womack, T. & Bricogne, G. (2011). Acta Cryst. D67, 293–302.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWarne, T., Serrano-Vega, M. J., Baker, J. G., Moukhametzianov, R., Edwards, P. C., Henderson, R., Leslie, A. G. W, Tate, C. G. & Schertler, G. F. (2008). Nature (London), 454, 486–491.  CrossRef CAS Google Scholar
First citationWasserman, S. R., Koss, J. W., Sojitra, S. T., Morisco, L. L. & Burley, S. K. (2012). Trends Pharm. Sci. 33, 261–267.  CrossRef CAS Google Scholar
First citationWinter, G. (2010). J. Appl. Cryst. 43, 186–190.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationZander, U., Cianci, M., Foos, N., Silva, C. S., Mazzei, L., Zubieta, C., de Maria, A. & Nanao, M. H. (2016). Acta Cryst. D72, 1026–1035.  Web of Science CrossRef IUCr Journals Google Scholar
First citationZander, U., Hoffmann, G. et al. (2016). Acta Cryst. D72, 454–466.  CrossRef IUCr Journals Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoSTRUCTURAL
BIOLOGY
ISSN: 2059-7983
Follow Acta Cryst. D
Sign up for e-alerts
Follow Acta Cryst. on Twitter
Follow us on facebook
Sign up for RSS feeds