Download citation
Download citation
link to html
The successful use of cryogenically cooled silicon monochromators at third-generation synchrotron facilities is well documented. At the Advanced Photon Source (APS) it has been shown that, at 100 mA operation with the standard APS undulator A, the cryogenically cooled silicon monochromator performs very well with minimal (<2 arcsec) or no observable thermal distortions. However, to date there has not been any systematic experimental study on the performance limits of this approach. This paper presents experimental results on the performance limits of these directly cooled crystals. The results show that if the beam is limited to the size of the radiation central cone then, at the APS, the crystal will still perform well at twice the present 100 mA single 2.4 m-long 3.3 cm-period undulator heat load. However, the performance would degrade rapidly if a much larger incident white-beam size is utilized.

Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds