Download citation
Download citation
link to html
Thermal neutron transmission measurements have been made as a function of wavelength on a pyrolytic graphite monochromator crystal that has been set to diffract a horizontal beam at different take-off angles. The major dips in the transmission caused by the various reflections have been identified. These results can be used for the design of a beamline on which more than one instrument is placed. The transmission data show that it is best for the monochromator with the greatest (horizontal) take-off angle to be placed upstream, with monochromators with decreasing take-off angles progressively further downstream. The order of instruments for which the wavelength is greater than 0.43 nm is unimportant.