Download citation
Download citation
link to html
The aim of this work is to investigate the possibility of extracting correct structural parameters from fluorescence EXAFS data taken at high count rates with an energy-resolving detector. This situation is often encountered on third-generation synchrotron radiation sources which provide a high flux on the sample. Errors caused by pulse pile-up in the extraction of structural information have been quantified in a real experiment, and different approaches to the problem of data correction have been elaborated. The different approaches are discussed in a comparison of the ability of each kind of correction to recover the correct structural parameters. The result of our analysis is that it is possible to work in non-linear conditions and correct the data, if the response of the acquisition system is known. Reliable structural information can be obtained with data acquired up to a count rate equal to approximately 60% of the inverse of the dead time.