laboratory notes
A bespoke capillary sample holder is described that attaches to the cold head of a commercially manufactured (PheniX) closed-cycle helium cryostat originally intended for flat-plate geometry. The new holder allows high-resolution synchrotron powder diffraction data to be collected from samples in Debye-Scherrer geometry over the temperature range 11-295 K.