Convenient off-line error quantification and characterization of concentricity of two circles of rotation for diffractometer alignment
L. M. D. Cranswick
,
R. Donaberger
,
I. P. Swainson
and
Z. Tun
To ensure concentricity of the θ–2θ rotation stages, but with a very limited 2θ range of rotation, an alignment technique using two dial gauges mounted at right angles to each other (
X
and
Y
axes) is described.