photondiag2017 workshop\(\def\hfill{\hskip 5em}\def\hfil{\hskip 3em}\def\eqno#1{\hfil {#1}}\)

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775

Direct experimental observation of the gas density depression effect using a two-bunch X-ray FEL beam

crossmark logo

aSLAC National Accelerator Laboratory, 2575 Sand Hill Road, Menlo Park, CA 94025, USA
*Correspondence e-mail: yfeng@slac.stanford.edu

Edited by M. Zangrando, IOM-CNR and Elettra-Sincrotrone, Italy (Received 1 June 2017; accepted 3 October 2017)

The experimental observation of the depression effect in gas devices designed for X-ray free-electron lasers (FELs) is reported. The measurements were carried out at the Linac Coherent Light Source using a two-bunch FEL beam at 6.5 keV with 122.5 ns separation passing through an argon gas cell. The relative intensities of the two pulses of the two-bunch beam were measured, after and before the gas cell, from X-ray scattering off thin targets by using fast diodes with sufficient temporal resolution. At a cell pressure of 140 hPa, it was found that the after-to-before ratio of the intensities of the second pulse was about 17% ± 6% higher than that of the first pulse, revealing lower effective attenuation of the gas cell due to heating by the first pulse and subsequent gas density reduction in the beam path. This measurement is important in guiding the design and/or mitigating the adverse effects in gas devices for high-repetition-rate FELs such as the LCLS-II and the European XFEL or other future high-repetition-rate upgrades to existing FEL facilities.

1. Introduction

The phase II project of the Linac Coherent Light Source (LCLS) (Emma et al., 2010[Emma, P. et al. (2010). Nat. Photon. 4, 641-647.]) will seek to upgrade the LCLS conventional accelerator to a superconducting linac, similar to that of the upcoming European XFEL, to generate X-ray free-electron laser (FEL) beams at much higher repetition rates up to 1 MHz. The high-repetition-rate operation of the European XFEL and LCLS-II is expected to provide additional capabilities to those already offered by the low-repetition-rate facilities that are either currently in user operation or being commissioned, including FLASH (Ackermann et al., 2007[Ackermann, W. et al. (2007). Nat. Photon. 1, 336-342.]) and its upgrade FLASH-II, LCLS, SACLA (Ishikawa et al., 2012[Ishikawa, T. et al. (2012). Nat. Photon. 6, 540-544.]), FERMI (Allaria et al., 2012[Allaria, E. et al. (2012). Nat. Photon. 6, 699-704.]), PAL-XFEL (Kang et al., 2013[Kang, H.-S., Han, J. H., Kang, T.-H., Kim, C., Kim, D. E., Kim, S. H., Ko, I. S., Lee, H.-S., Park, K.-H. & Park, S. J. (2013). Proceedings of the 2013 International Particle Accelerator Conference, Shanghai, China, 12-17 May 2013, p. 2074.]) and SwissFEL (Ganter et al., 2010[Ganter, R. (2010). SwissFEL-Conceptual Design Report. Paul Scherrer Institute, Villigen, Switzerland.]). At the same time, these new capabilities also bring about new challenges in conceiving, designing and implementing high-repetition-rate-compatible X-ray diagnostic, optics, beam regulation and safety devices, which have been proven to be very important to help fulfill FEL's great scientific potentials in the frontier research of physics, chemistry, life science, material, energy and earth sciences (Young et al., 2010[Young, L., Kanter, E., Krässig, B., Li, Y., March, A., Pratt, S., Santra, R., Southworth, S., Rohringer, N., Dimauro, L., Doumy, G., Roedig, C. A., Berrah, N., Fang, L., Hoener, M., Bucksbaum, P. H., Cryan, J. P., Ghimire, S., Glownia, J. M., Reis, D. A., Bozek, J. D., Bostedt, C. & Messerschmidt, M. (2010). Nature (London), 466, 56-61.]; Glover et al., 2012[Glover, T., Fritz, D., Cammarata, M., Allison, T., Coh, S., Feldkamp, J., Lemke, H., Zhu, D., Feng, Y., Coffee, R., Fuchs, M., Ghimire, S., Chen, J., Shwartz, S., Reis, D. A., Harris, S. E. & Hastings, J. B. (2012). Nature (London), 488, 603-608.]; Fuchs et al., 2015[Fuchs, M., Trigo, M., Chen, J., Ghimire, S., Shwartz, S., Kozina, M., Jiang, M., Henighan, T., Bray, C., Ndabashimiye, G., Bucksbaum, P. H., Feng, Y., Herrmann, S., Carini, G. A., Pines, J., Hart, P., Kenney, C., Guillet, S., Boutet, S., Williams, G. J., Messerschmidt, M., Seibert, M. M., Moeller, S., Hastings, J. B. & Reis, D. A. (2015). Nat. Phys. 11, 964-970.]; Minitti et al., 2015[Minitti, M., Budarz, J., Kirrander, A., Robinson, J., Ratner, D., Lane, T., Zhu, D., Glownia, J., Kozina, M., Lemke, H., Sikorski, M., Feng, Y., Nelson, S., Saita, K., Stankus, B., Northey, T., Hastings, J. B. & Weber, P. M. (2015). Phys. Rev. Lett. 114, 255501.]; Chapman et al., 2011[Chapman, H. N. et al. (2011). Nature (London), 470, 73-77.]; Seibert et al., 2011[Seibert, M. M. et al. (2011). Nature (London), 470, 78-81.]; Shwartz et al., 2014[Shwartz, S., Fuchs, M., Hastings, J. B., Inubushi, Y., Ishikawa, T., Katayama, T., Reis, D. A., Sato, T., Tono, K., Yabashi, M., Yudovich, S. & Harris, S. E. (2014). Phys. Rev. Lett. 112, 163901.]; Gerber et al., 2015[Gerber, S., Jang, H., Nojiri, H., Matsuzawa, S., Yasumura, H., Bonn, D., Liang, R., Hardy, W., Islam, Z., Mehta, A., Song, S., Sikorski, M., Stefanescu, D., Feng, Y., Kivelson, S. A., Devereaux, T. P., Shen, Z. X., Kao, C. C., Lee, W. S., Zhu, D. & Lee, J. S. (2015). Science, 350, 949-952.]; Harmand et al., 2015[Harmand, M., Ravasio, A., Mazevet, S., Bouchet, J., Denoeud, A., Dorchies, F., Feng, Y., Fourment, C., Galtier, E., Gaudin, J., Guyot, F., Kodama, R., Koenig, M., Lee, H. J., Miyanishi, K., Morard, G., Musella, R., Nagler, B., Nakatsutsumi, M., Ozaki, N., Recoules, V., Toleikis, S., Vinci, T., Zastrau, U., Zhu, D. & Benuzzi-Mounaix, A. (2015). Phys. Rev. B, 92, 024108.]; Yoneda et al., 2015[Yoneda, H., Inubushi, Y., Nagamine, K., Michine, Y., Ohashi, H., Yumoto, H., Yamauchi, K., Mimura, H., Kitamura, H., Katayama, T., Ishikawa, T. & Yabashi, M. (2015). Nature (London), 524, 446-449.]). The ever-so-important X-ray diagnostics for a FEL stems from the stochastic nature of its lasing mechanism, i.e. the self-amplified spontaneous emission (SASE) process (Kondratenko & Saldin, 1979[Kondratenko, A. & Saldin, E. (1979). Dokl. Akad. Nauk SSSR, 249, 843-847.]; Bonifacio et al., 1984[Bonifacio, R., Pellegrini, C. & Narducci, L. M. (1984). Opt. Commun. 50, 373-378.]), which gives rise to shot-to-shot fluctuations in all beam properties, requiring single-shot diagnostics in pulse energy, timing, spectrum, polarization and more. The successful construction of these diagnostics devices (Richter et al., 2003[Richter, M., Gottwald, A., Kroth, U., Sorokin, A., Bobashev, S., Shmaenok, L., Feldhaus, J., Gerth, C., Steeg, B., Tiedtke, K. & Treusch, R. (2003). Appl. Phys. Lett. 83, 2970-2972.]; Hau-Riege et al., 2008[Hau-Riege, S. P., Bionta, R., Ryutov, D. & Krzywinski, J. (2008). J. Appl. Phys. 103, 053306.]; Feng et al., 2011[Feng, Y., Feldkamp, J. M., Fritz, D. M., Cammarata, M., Aymeric, R., Caronna, C., Lemke, H. T., Zhu, D., Lee, S., Boutet, S., Williams, G., Tono, K., Yabashi, M. & Hastings, J. B. (2011). Proc. SPIE, 8104, 81040Q.]; Bionta et al., 2011[Bionta, M. R., Lemke, H., Cryan, J., Glownia, J., Bostedt, C., Cammarata, M., Castagna, J.-C., Ding, Y., Fritz, D., Fry, A., Krzywinski, J., Messerschmidt, M., Schorb, S., Swiggers, M. L. & Coffee, R. N. (2011). Opt. Express, 19, 21855-21865.]; Zhu et al., 2012[Zhu, D., Cammarata, M., Feldkamp, J. M., Fritz, D. M., Hastings, J. B., Lee, S., Lemke, H. T., Robert, A., Turner, J. L. & Feng, Y. (2012). Appl. Phys. Lett. 101, 034103.]; Allaria et al., 2014[Allaria, E. et al. (2014). Phys. Rev. X, 4, 041040.]; Braune et al., 2016[Braune, M., Brenner, G., Dziarzhytski, S., Juranić, P., Sorokin, A. & Tiedtke, K. (2016). J. Synchrotron Rad. 23, 10-20.]) have allowed users to gain sensitivity in signals which are otherwise difficult to discern, especially those in X-ray pump–probe experiments (Chollet et al., 2015[Chollet, M., Alonso-Mori, R., Cammarata, M., Damiani, D., Defever, J., Delor, J. T., Feng, Y., Glownia, J. M., Langton, J. B., Nelson, S., Ramsey, K., Robert, A., Sikorski, M., Song, S., Stefanescu, D., Srinivasan, V., Zhu, D., Lemke, H. T. & Fritz, D. M. (2015). J. Synchrotron Rad. 22, 503-507.]), as well as enabled FEL operators to tune up the FEL performance and develop new operating modes (Amann et al., 2012[Amann, J., Berg, W., Blank, V., Decker, F.-J., Ding, Y., Emma, P., Feng, Y., Frisch, J., Fritz, D., Hastings, J., Huang, Z., Krzywinski, J., Lindberg, R., Loos, H., Lutman, A., Nuhn, H.-D., Ratner, D., Rzepiela, J., Shu, D., Shvyd'ko, Y., Spampinati, S., Stoupin, S., Terentyev, S., Trakhtenberg, E., Walz, D., Welch, J., Wu, J., Zholents, A. & Zhu, D. (2012). Nat. Photon. 6, 693-698.]; Marinelli et al., 2013[Marinelli, A., Lutman, A. A., Wu, J., Ding, Y., Krzywinski, J., Nuhn, H.-D., Feng, Y., Coffee, R. N. & Pellegrini, C. (2013). Phys. Rev. Lett. 111, 134801.]; Lutman et al., 2014[Lutman, A., Decker, F.-J., Arthur, J., Chollet, M., Feng, Y., Hastings, J., Huang, Z., Lemke, H., Nuhn, H.-D., Marinelli, A., Turner, J. L., Wakatsuki, S., Welch, J. & Zhu, D. (2014). Phys. Rev. Lett. 113, 254801.]).

At soft X-ray energies especially, FEL diagnostics and components are often limited to those based on a gas medium to circumvent single-shot damage risks posed by the enormous peak power of a FEL beam. For example, gas intensity monitors and attenuators have been implemented at various facilities including FLASH (Richter et al., 2003[Richter, M., Gottwald, A., Kroth, U., Sorokin, A., Bobashev, S., Shmaenok, L., Feldhaus, J., Gerth, C., Steeg, B., Tiedtke, K. & Treusch, R. (2003). Appl. Phys. Lett. 83, 2970-2972.]; Hahn & Tiedtke, 2007[Hahn, U. & Tiedtke, K. (2007). AIP Conf. Proc. 879, 276-282.]), FLASH-II and LCLS (Hau-Riege et al., 2008[Hau-Riege, S. P., Bionta, R., Ryutov, D. & Krzywinski, J. (2008). J. Appl. Phys. 103, 053306.]; Ryutov et al., 2009[Ryutov, D. D., et al. (2009). Technical Report LLNL-TR-421318, Lawrence Livermore National Laboratory, Livermore, CA, USA.]), and are being planned for LCLS-II. The LCLS has the highest operating frequency of all low-repetition-rate facilities, in user operation now or otherwise, at 120 Hz, leaving a time ΔT of ∼8.3 ms for the gas system to return to its starting density after the passage of each pulse. A fast energy dissipation has always been implicitly assumed for longer ΔT, but this assumption will start to break down as ΔT will be reduced by about four orders of magnitude to 222 ns for the European XFEL and 1 µs for LCLS-II. The impact of the very short ΔT on the performance of these gas diagnostics and devices has been studied extensively using thermodynamic and hydro­dynamic simulations (Feng et al., 2015a[Feng, Y., Krzywinski, J., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2015a). Proc. SPIE, 9589, 958910.],b[Feng, Y., Krzywinski, J., Campell, M. L., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2015b). Proceedings of the 37th International Free Electron Laser Conference (FEL2015), Daejeon, Korea, 23-28 August 2015. TUP026.], 2016[Feng, Y., Krzywinski, J., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2016). J. Synchrotron Rad. 23, 21-28.]; Yang et al., 2017[Yang, B., Wu, J., Raubenheimer, T. O. & Feng, Y. (2017). J. Synchrotron Rad. 24, 547-559.]; Feng & Raubenheimer, 2017[Feng, Y. & Raubenheimer, T. O. (2017). Proc. SPIE, 10237, 1023714.]), revealing nonlinear effects in the precision of intensity measurements for gas monitors operating at relatively high pressures (Hau-Riege et al., 2008[Hau-Riege, S. P., Bionta, R., Ryutov, D. & Krzywinski, J. (2008). J. Appl. Phys. 103, 053306.]) as well as the effective attenuation for gas attenuators for trailing pulses, stemming from the density depression or depression phenomenon induced by the energy deposition of the preceding pulses. Parallel efforts have been put forth to experimentally confirm and quantify the depression effect, first by using an optical pump and optical probe technique similar to that used in a previous experiment (Cheng et al., 2013[Cheng, Y.-H., Wahlstrand, J. K., Jhajj, N. & Milchberg, H. (2013). Opt. Express, 21, 4740-4751.]), and then an X-ray pump and optical probe measurement (Galtier & Schafer, 2017[Galtier, E. & Schafer, D. (2017). Unpublished.]). It should be noted that for gas monitors operating at extremely low pressures (Richter et al., 2003[Richter, M., Gottwald, A., Kroth, U., Sorokin, A., Bobashev, S., Shmaenok, L., Feldhaus, J., Gerth, C., Steeg, B., Tiedtke, K. & Treusch, R. (2003). Appl. Phys. Lett. 83, 2970-2972.]) the amount of energy deposited into the medium is very low, and more importantly the absorbed energy is dissipated via the detection mechanisms themselves, i.e. photoelectrons at the anode and ions neutralized at the cathode in the form of currents.

In this report, we present the first direct experimental observation of the gas depression effect by an ultrafast X-ray FEL beam using an X-ray pump and X-ray probe technique. The measurements were carried out at the LCLS on the X-ray Correlation Spectroscopy (XCS) instrument (Alonso-Mori et al., 2015[Alonso-Mori, R., Caronna, C., Chollet, M., Curtis, R., Damiani, D. S., Defever, J., Feng, Y., Flath, D. L., Glownia, J. M., Lee, S., Lemke, H. T., Nelson, S., Bong, E., Sikorski, M., Song, S., Srinivasan, V., Stefanescu, D., Zhu, D. & Robert, A. (2015). J. Synchrotron Rad. 22, 508-513.]) at an X-ray energy of 6.5 keV. Using a special operating mode of the LCLS, a two-bunch FEL beam having two similar intensity femtosecond pulses, which were spatially collinear but separated by over 100 ns in time, was generated. It was sent through the argon-filled gas cell, and the intensities of the two pulses were measured independently, before and after the gas cell. It was found that the transmission of the second pulse was consistently higher than that of the first pulse, revealing a lower effective attenuation consistent with the gas depression effect predicted by thermodynamic and hydro­dynamic calculations (Feng et al., 2015a[Feng, Y., Krzywinski, J., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2015a). Proc. SPIE, 9589, 958910.],b[Feng, Y., Krzywinski, J., Campell, M. L., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2015b). Proceedings of the 37th International Free Electron Laser Conference (FEL2015), Daejeon, Korea, 23-28 August 2015. TUP026.], 2016[Feng, Y., Krzywinski, J., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2016). J. Synchrotron Rad. 23, 21-28.]; Yang et al., 2017[Yang, B., Wu, J., Raubenheimer, T. O. & Feng, Y. (2017). J. Synchrotron Rad. 24, 547-559.]; Feng & Raubenheimer, 2017[Feng, Y. & Raubenheimer, T. O. (2017). Proc. SPIE, 10237, 1023714.]).

2. Experimental setup

To perform a direct X-ray pump and X-ray probe of the gas depression phenomenon at LCLS before the arrival of LCLS-II, or the availability of a hard X-ray split-and-delay device, the two X-ray pulses were generated using an accelerator-based approach. In a specially developed operating procedure, or the so-called `two-bunch operating mode', two spatially collinear and similar intensity pulses p1 and p2 could be produced with p2 delayed by time Δt amounting to multiples of 0.35 ns (Decker et al., 2010[Decker, F.-J., Akre, R., Brachmann, A., Ding, Y., Dowell, D., Emma, P. J., Fisher, A., Frisch, J., Gilevich, S., Hering, Ph., Huang, Z., Iverson, R. H., Loos, H., Messerschmidt, M., Nuhn, H.-D., Ratner, D. F., Schlotter, W. F., Smith, T. J., Turner, J. L., Welch, J. J., White, W. E. & Wu, J. (2010). Proceedings of the 32nd International Free Electron Laser Conference (FEL2010), Malmo, Sweden, 23-27 August 2010. WEPB33.]). For reasons that will become clear later, an optimal delay of order 1 µs would have been ideal for the current experiment, but rather a delay of only 122.5 ns was used due to a limitation of the two-bunch mode. The average intensity of the first pulse p1 was estimated to be about 1 mJ entering into the gas cell and was purposely tuned to be greater than that of p2 to maximize the pump and yet to retain sufficient signal-to-noise ratio for the probe.

The two-bunch p1 and p2 FEL beam was directed through an argon-filled gas cell depicted schematically in Fig. 1[link], and attenuated. The gas cell had a diameter 2R = 22.1 mm and was sealed off by two 50 µm-thick Kapton windows W1 and W2, each inclined at 45° to the direction of the FEL beam and separated by a nominal distance of L = 165 ± 8 mm depending on the exact beam path through the cell. The gas pressure P was measured via an MKS Baratron gauge (Model No. 722B23TGA2FA), with a reading precision of 0.5%, and was varied from P = 0 to P = 257 hPa. To minimize gas motions not related to the underlying physics to be studied, the gas cell was valved off, but a small (pinhole) leak in the Kapton window developed towards the end of the experiment, leading to a slow creep-up of the gas pressure on a time scale of many tens of minutes. Care was taken to account for the gas leakage by estimating the true effective pressure of the Ar/air mixture by the transmission of a first pulse of the two-bunch beam.

[Figure 1]
Figure 1
Schematic of the experimental setup, consisting of a gas cell of diameter 2R and effective length L, filled with argon gas to a pressure P. The cell was capped off by two 50 µm-thick Kapton windows W1 and W2 each inclined by 45° to the direction of an X-ray two-bunch FEL beam coming from the left and being attenuated. The X-ray scattering off the Kapton windows was measured as the intensities of p1 and p2 by two fast diodes D1 and D2 positioned perpendicular to the beam and each connected to a different channel of a fast digitizer.

The total X-ray scattering, including both the coherent Thomson, incoherent Compton and other processes, off the Kapton windows was measured via two Hamamatsu diodes (Model No. MSM Photodetector G4176-03), D1 and D2, positioned perpendicular to the beam. The sides of the diodes were shielded from scattering, but no thin metal foils were used to block any ambient light. The rise time of the diodes was estimated to be of the order of 1 ns, and each diode was connected directly to a different input channel (via a 50 Ω coupling resistor) of a fast digitizer (Aquiris 10-bit High-Speed cPCI digitizer, Model No. U1065A). The sampling time was set to 0.25 ns and there was a total of 4000 sampling channels, generating a 1 µs range of digitization. As such, the contribution in the dark noise of the digitizer by ambient light was completely negligible, although any concurrent optical emission upon the passage of the X-ray pulses within the digitization window cannot be separated from the signal. The scattering intensities of p1 and p2 were used as measurements of the pulse intensity I1i and I2i, where i = 1 for upstream measurement and i = 2 for downstream measurement.

The nominal photon energy of the two-bunch FEL beam was 6.5 keV, with each bunch (pulse) produced independently, and exhibited typical SASE characteristics, including large fluctuations in intensity. The FWHM beam size was measured to be 800 µm in width and 1 mm in height. The two pulses could be steered independently, and care was taken to ensure their spatial overlap over the length of the gas cell by better than 80% of the FWHM size in either transverse direction using a downstream high-resolution beam imager with a spatial resolution of 8 µm. Note that a complete overlap is in principle not achievable due to the SASE nature of FEL beams, which can produce approximately 10% relative spatial jitter pulse-to-pulse and independent of the absolute beam size. The time delay was fixed at 122.5 ns, the maximum achievable with the LCLS accelerator configuration at the time of the measurement. As such, the rise time of the diodes (∼1 ns) was sufficient to resolve the twin pulses, making the data analysis very straightforward without having to be concerned about temporally overlapping signals.

The two-bunch FEL beam size was about 1 mm FWHM before focusing. The formation of the density depression upon passage of a FEL beam takes place over a time scale set by the speed of the shock wave vshock (Yang et al., 2017[Yang, B., Wu, J., Raubenheimer, T. O. & Feng, Y. (2017). J. Synchrotron Rad. 24, 547-559.]) at 200 m s−1, which is approximately on the same scale as the speed of sound in an ideal gas (vsound ≃ 300 m s−1). As such, the edge of the density depression only reaches about 24 µm after 122.5 ns and is much smaller than the half-width of the beam at 500 µm, making it much more difficult to observe using the transmission measurement described next. By focusing the beam down to about 100 µm on average over the length of the cell (the spatial jitter is still 10% or 10 µm), clear evidence of an enhanced transmission of the second pulse as induced by the first pulse was found, in qualitative agreement with simulations.

3. Data analysis and results

3.1. Transmissions at zero attenuation

The average traces of the pulses p1 and p2 from the digitizer of the two-bunch pulses at P = 0 hPa are shown in Fig. 2[link], with those measured by diode D2 somewhat larger and at the same time narrower by about 16 ± 1% thus summing up to the same integrated intensity. This is due to the slight difference in the rise time of the two diodes. The average was performed over more than 32000 pulses. There were two contributions to the background, one of random nature and the other periodic and alternating exactly at the first subharmonic of the sampling frequency of 0.25 ns of the digitizer, indicating a digital artifact arising from the clocking scheme. This periodic contribution was simply removed by applying a filter at half of the sampling frequency over the entire sampling range, thus is not visible in Fig. 2[link].

[Figure 2]
Figure 2
Average digitizer traces of the two-bunch beam measured by (a) the upstream diode D1 and (b) the downstream diode D2 at an argon pressure P = 0 hPa. The digitizer was set to sample at 0.25 ns per channel. The integrated intensities of the first pulse p1 are denoted as I11 and I12 on D1 and D2, respectively, and those of the second pulse p2 as I21 and I22 on D1 and D2, respectively.

The first pulse p1 and second pulse p2 were separated by exactly 122.5 ns, and the rise time of the upstream diode D1 was about 2 ns, whereas for the downstream diode D2 it was about 1.75 ns, consistent with the 16% difference in their respective amplitudes. The amplitude of the undershoots in traces was also proportional to the main peak, and was caused by a slight impedance mismatch or ringing. The relative intensities of the first pulse I1i and second pulse I2i on different diodes Di, for i = 1 or 2, were obtained by integrating the digitizer traces over the main peak (not including the undershoot), and the average transmission coefficients of the two pulses Tj = Ij2/Ij1, for j = 1 and 2, were then calculated, obtaining 〈T1〉 = 1.01 ± 0.02 and 〈T2〉 = 0.99 ± 0.02, respectively. The near equality of 〈T1〉 and 〈T2〉 within the experimental uncertainty was expected as there was no attenuation at P = 0 hPa.

The Hamamatsu G4176-03 diode was chosen to achieve sufficient time resolution, but at the expense of having a relatively small dynamic range for detecting hard X-rays. At 6.5 keV, a large number of electron–hole pairs, approximately 1800, are created by a single photon, and the dynamic range was limited to approximately 100 6.5 keV X-rays photons. To avoid saturation, the number of scattered photons into the diodes was purposely limited to less than 100 by adjusting the distances of the diodes to the Kapton windows. As such, large uncertainties were observed in the single-pulse measurement of Iji for i, j = 1 and 2, as shown in Figs. 3(a) and 3(b)[link] and were attributed to Poisson noise arising from the small number of detected X-ray photons (<100 or a minimum of 10% shot noise). At a given measured relative intensity I11 or I21, there is a very wide range of values for I12 or I22, with the standard deviation reaching as much as 25%. However, the correlations such as in Fig. 3[link] still provided consistent transmission measurements with those obtained from the average traces in Fig. 2[link], as given by the slope of a linear fit, at 〈T1〉 = 0.975 ± 0.003 and 〈T2〉 = 0.976 ± 0.001 for the first and second pulses, respectively.

[Figure 3]
Figure 3
Correlations between the intensities measured by the upstream and downstream diodes of (a) the first pulse and (b) the second pulse at argon gas pressure P = 0 hPa. The slope is a measure of the transmission, amounting to 〈T1〉 = 97.5% for the first pulse and nearly identical 〈T2〉 = 97.6% for the second pulse as expected. The very slight difference from unity is considered as a systematic error of the detection scheme. The red line for the first pulse was overlaid in (b) to stress the similarities between the transmissions of the two pulses.

3.2. Transmissions at finite attenuation

Additional measurements similar to those at zero attenuation were repeated at finite attenuations. The average traces of the pulses p1 and p2 from the digitizer of the two-bunch pulses at an effective argon pressure of P = 140 hPa are shown in Fig. 4[link]. During the measurement at P = 0 hPa, the FEL beam was focused from about 1 mm in size to 100 µm at the location of the gas cell to gain sensitivity. The reading on the pressure gauge at 132 hPa was lower than the actual transmission would indicate; the discrepancy was not completely understood. Again, the systematic periodic background was first removed. Similar to the zero-pressure case, the trace average was performed over more than 35000 pulses. The relative intensities on the downstream diode were lower as expected than those on the upstream but by different amounts in relative terms, indicating different effective transmissions for the two pulses.

[Figure 4]
Figure 4
Average digitizer traces of the two-bunch beam measured by (a) the upstream diode D1 and (b) the downstream diode D2 at an effective argon pressure P = 140 hPa. The various intensities Iij are defined previously in Fig. 2[link].

The different effective transmissions for the two pulses are best shown by the linear fits of the intensity correlation plots in Fig. 5[link], equalling 〈T1〉 = 0.427 for the first pulse and a higher value 〈T2〉 = 0.498 for the second pulse. This amounts to a 17% enhancement in the transmission of the second pulse induced by the first pulse. This difference is further highlighted by overlaying the calculated transmissions in Fig. 6(a)[link] sorted by the intensity of the probe pulse I11, and the transmission histograms in Fig. 6(b)[link], both giving an average of 〈T1〉 = 43% and 〈T2〉 = 50% for the first and second pulse, respectively. The large uncertainty in the measurements due to the Poisson noise was again apparent in both figures, with the second pulse having roughly 50% greater noise, consistent with its reduced intensity at 43%, which should scale like [1/\sqrt{0.43}] ≃ 1.52.

[Figure 5]
Figure 5
Correlations between the intensities measured by the upstream and downstream diodes of (a) the first pulse and (b) the second pulse at an equivalent argon gas pressure P = 140 hPa. The slope is a measure of the transmission, amounting to 42.7% for the first pulse and an enhanced 49.8% for the second pulse.
[Figure 6]
Figure 6
(a) Transmission of the first and second pulse sorted based on the intensity of the first or probe pulse I11, at an effective argon gas pressure P = 140 hPa, with an average of 〈T1〉 = 43% ± 3% and 〈T2〉 = 50% ± 5% for the first and second pulse, respectively. (b) Histogram of the transmission of the first and second pulse. The average transmissions are again 〈T1〉 = 43% ± 3% and 〈T2〉 = 50% ± 5% for the first and second pulse, respectively.

The enhanced transmission from 〈T1〉 = 43% to 〈T2〉 = 50% is much smaller than the 〈T2〉 = 75% value calculated based on the methodology developed in the previous thermodynamic studies (Feng et al., 2015a[Feng, Y., Krzywinski, J., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2015a). Proc. SPIE, 9589, 958910.]) and the current experimental geometries. There is a main reason for this discrepancy. The delay time Δt at 122.5 ns was too short so the density depression was not fully developed, or the density depression was not completely formed according to the recent hydro­dynamic studies (Yang et al., 2017[Yang, B., Wu, J., Raubenheimer, T. O. & Feng, Y. (2017). J. Synchrotron Rad. 24, 547-559.]). If using the sound speed of 200 m s−1 obtained in the hydro­dynamic simulation and the focused beam size of 100 µm, the largest transmission would occur at approximately 500 ns. As a possible explanation for the smaller than calculated transmission enhancement, one can simply extrapolate linearly the development of the density depression in time, and would arrive at an expected enhanced transmission of 〈T2〉 = (75% − 43%) × 122.5/500 + 43% = 51%, which is in very good agreement with the measurement. This consistency also supports the time-dependent thermodynamic simulations (Feng et al., 2015a[Feng, Y., Krzywinski, J., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2015a). Proc. SPIE, 9589, 958910.][Feng, Y., Krzywinski, J., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2015a). Proc. SPIE, 9589, 958910.]) for time scales beyond the fully opening of the density depression, which depends on the beam size and sound speed. For shorter time scales down to that of the thermalization (Feng et al., 2016[Feng, Y., Krzywinski, J., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2016). J. Synchrotron Rad. 23, 21-28.]), hydro­dynamic simulations are required to predict the exact transmission. Furthermore, the current experimental measurement also suggests that the main energy transfer mechanism is hydro­dynamic and then thermodynamic, and any radiative losses or other processes may be secondary. More experimental and theoretical studies should be pursued to gain further understanding of the performance of gas devices for high-repetition-rate X-ray FELs.

4. Summary

We have observed experimentally the depression effect in a gas cell induced by an ultrashort X-ray FEL beam using a two-bunch scheme. The observed enhanced transmission of the delayed second pulse in the wake of the first pulse was in good quantitative agreement with the estimate, which was based on thermodynamic simulations and proper extrapolation of the hydro­dynamic simulation results. No enhancement was found as expected when the cell was not filled with an attenuating gas. Our finding is important in guiding the design and/or possibly mitigating the adverse effect in gas devices operating at relatively high pressure for high-repetition-rate FELs such as the LCLS-II, the European XFEL or other future high-repetition-rate upgrades to the existing FEL facilities, and have broader implications not only on gas-based applications but also on liquid- and solid-based devices and experiments.

Acknowledgements

Use of the Linac Coherent Light Source (LCLS), SLAC National Accelerator Laboratory, is supported by the US Department of Energy, Office of Science, Office of Basic Energy Sciences. The authors would also like to thank T. Raubenheimer, M. Rowen, E. Ortiz, E. Galtier and Z. Huang for useful discussions and the LCLS XCS technical staff for their support of the experimental effort.

Funding information

The following funding is acknowledged: US Department of Energy, Office of Science (award No. DE-AC02-76SF00515).

References

First citationAckermann, W. et al. (2007). Nat. Photon. 1, 336–342.  Web of Science CrossRef Google Scholar
First citationAllaria, E. et al. (2012). Nat. Photon. 6, 699–704.  Web of Science CrossRef CAS Google Scholar
First citationAllaria, E. et al. (2014). Phys. Rev. X, 4, 041040.  Google Scholar
First citationAlonso-Mori, R., Caronna, C., Chollet, M., Curtis, R., Damiani, D. S., Defever, J., Feng, Y., Flath, D. L., Glownia, J. M., Lee, S., Lemke, H. T., Nelson, S., Bong, E., Sikorski, M., Song, S., Srinivasan, V., Stefanescu, D., Zhu, D. & Robert, A. (2015). J. Synchrotron Rad. 22, 508–513.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationAmann, J., Berg, W., Blank, V., Decker, F.-J., Ding, Y., Emma, P., Feng, Y., Frisch, J., Fritz, D., Hastings, J., Huang, Z., Krzywinski, J., Lindberg, R., Loos, H., Lutman, A., Nuhn, H.-D., Ratner, D., Rzepiela, J., Shu, D., Shvyd'ko, Y., Spampinati, S., Stoupin, S., Terentyev, S., Trakhtenberg, E., Walz, D., Welch, J., Wu, J., Zholents, A. & Zhu, D. (2012). Nat. Photon. 6, 693–698.  Web of Science CrossRef CAS Google Scholar
First citationBionta, M. R., Lemke, H., Cryan, J., Glownia, J., Bostedt, C., Cammarata, M., Castagna, J.-C., Ding, Y., Fritz, D., Fry, A., Krzywinski, J., Messerschmidt, M., Schorb, S., Swiggers, M. L. & Coffee, R. N. (2011). Opt. Express, 19, 21855–21865.  Web of Science CrossRef CAS PubMed Google Scholar
First citationBonifacio, R., Pellegrini, C. & Narducci, L. M. (1984). Opt. Commun. 50, 373–378.  CrossRef CAS Web of Science Google Scholar
First citationBraune, M., Brenner, G., Dziarzhytski, S., Juranić, P., Sorokin, A. & Tiedtke, K. (2016). J. Synchrotron Rad. 23, 10–20.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationChapman, H. N. et al. (2011). Nature (London), 470, 73–77.  Web of Science CrossRef CAS PubMed Google Scholar
First citationCheng, Y.-H., Wahlstrand, J. K., Jhajj, N. & Milchberg, H. (2013). Opt. Express, 21, 4740–4751.  CrossRef PubMed Google Scholar
First citationChollet, M., Alonso-Mori, R., Cammarata, M., Damiani, D., Defever, J., Delor, J. T., Feng, Y., Glownia, J. M., Langton, J. B., Nelson, S., Ramsey, K., Robert, A., Sikorski, M., Song, S., Stefanescu, D., Srinivasan, V., Zhu, D., Lemke, H. T. & Fritz, D. M. (2015). J. Synchrotron Rad. 22, 503–507.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationDecker, F.-J., Akre, R., Brachmann, A., Ding, Y., Dowell, D., Emma, P. J., Fisher, A., Frisch, J., Gilevich, S., Hering, Ph., Huang, Z., Iverson, R. H., Loos, H., Messerschmidt, M., Nuhn, H.-D., Ratner, D. F., Schlotter, W. F., Smith, T. J., Turner, J. L., Welch, J. J., White, W. E. & Wu, J. (2010). Proceedings of the 32nd International Free Electron Laser Conference (FEL2010), Malmo, Sweden, 23–27 August 2010. WEPB33.  Google Scholar
First citationEmma, P. et al. (2010). Nat. Photon. 4, 641–647.  Web of Science CrossRef CAS Google Scholar
First citationFeng, Y., Feldkamp, J. M., Fritz, D. M., Cammarata, M., Aymeric, R., Caronna, C., Lemke, H. T., Zhu, D., Lee, S., Boutet, S., Williams, G., Tono, K., Yabashi, M. & Hastings, J. B. (2011). Proc. SPIE, 8104, 81040Q.  Google Scholar
First citationFeng, Y., Krzywinski, J., Campell, M. L., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2015b). Proceedings of the 37th International Free Electron Laser Conference (FEL2015), Daejeon, Korea, 23–28 August 2015. TUP026.  Google Scholar
First citationFeng, Y., Krzywinski, J., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2015a). Proc. SPIE, 9589, 958910.  Google Scholar
First citationFeng, Y., Krzywinski, J., Schafer, D. W., Ortiz, E., Rowen, M. & Raubenheimer, T. O. (2016). J. Synchrotron Rad. 23, 21–28.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationFeng, Y. & Raubenheimer, T. O. (2017). Proc. SPIE, 10237, 1023714.  Google Scholar
First citationFuchs, M., Trigo, M., Chen, J., Ghimire, S., Shwartz, S., Kozina, M., Jiang, M., Henighan, T., Bray, C., Ndabashimiye, G., Bucksbaum, P. H., Feng, Y., Herrmann, S., Carini, G. A., Pines, J., Hart, P., Kenney, C., Guillet, S., Boutet, S., Williams, G. J., Messerschmidt, M., Seibert, M. M., Moeller, S., Hastings, J. B. & Reis, D. A. (2015). Nat. Phys. 11, 964–970.  Web of Science CrossRef CAS Google Scholar
First citationGaltier, E. & Schafer, D. (2017). Unpublished.  Google Scholar
First citationGanter, R. (2010). SwissFEL-Conceptual Design Report. Paul Scherrer Institute, Villigen, Switzerland.  Google Scholar
First citationGerber, S., Jang, H., Nojiri, H., Matsuzawa, S., Yasumura, H., Bonn, D., Liang, R., Hardy, W., Islam, Z., Mehta, A., Song, S., Sikorski, M., Stefanescu, D., Feng, Y., Kivelson, S. A., Devereaux, T. P., Shen, Z. X., Kao, C. C., Lee, W. S., Zhu, D. & Lee, J. S. (2015). Science, 350, 949–952.  CrossRef CAS PubMed Google Scholar
First citationGlover, T., Fritz, D., Cammarata, M., Allison, T., Coh, S., Feldkamp, J., Lemke, H., Zhu, D., Feng, Y., Coffee, R., Fuchs, M., Ghimire, S., Chen, J., Shwartz, S., Reis, D. A., Harris, S. E. & Hastings, J. B. (2012). Nature (London), 488, 603–608.  CrossRef CAS PubMed Google Scholar
First citationHahn, U. & Tiedtke, K. (2007). AIP Conf. Proc. 879, 276–282.  CrossRef CAS Google Scholar
First citationHarmand, M., Ravasio, A., Mazevet, S., Bouchet, J., Denoeud, A., Dorchies, F., Feng, Y., Fourment, C., Galtier, E., Gaudin, J., Guyot, F., Kodama, R., Koenig, M., Lee, H. J., Miyanishi, K., Morard, G., Musella, R., Nagler, B., Nakatsutsumi, M., Ozaki, N., Recoules, V., Toleikis, S., Vinci, T., Zastrau, U., Zhu, D. & Benuzzi-Mounaix, A. (2015). Phys. Rev. B, 92, 024108.  Web of Science CrossRef Google Scholar
First citationHau-Riege, S. P., Bionta, R., Ryutov, D. & Krzywinski, J. (2008). J. Appl. Phys. 103, 053306.  Google Scholar
First citationIshikawa, T. et al. (2012). Nat. Photon. 6, 540–544.  Web of Science CrossRef CAS Google Scholar
First citationKang, H.-S., Han, J. H., Kang, T.-H., Kim, C., Kim, D. E., Kim, S. H., Ko, I. S., Lee, H.-S., Park, K.-H. & Park, S. J. (2013). Proceedings of the 2013 International Particle Accelerator Conference, Shanghai, China, 12–17 May 2013, p. 2074.  Google Scholar
First citationKondratenko, A. & Saldin, E. (1979). Dokl. Akad. Nauk SSSR, 249, 843–847.  Google Scholar
First citationLutman, A., Decker, F.-J., Arthur, J., Chollet, M., Feng, Y., Hastings, J., Huang, Z., Lemke, H., Nuhn, H.-D., Marinelli, A., Turner, J. L., Wakatsuki, S., Welch, J. & Zhu, D. (2014). Phys. Rev. Lett. 113, 254801.  Web of Science CrossRef PubMed Google Scholar
First citationMarinelli, A., Lutman, A. A., Wu, J., Ding, Y., Krzywinski, J., Nuhn, H.-D., Feng, Y., Coffee, R. N. & Pellegrini, C. (2013). Phys. Rev. Lett. 111, 134801.  Web of Science CrossRef PubMed Google Scholar
First citationMinitti, M., Budarz, J., Kirrander, A., Robinson, J., Ratner, D., Lane, T., Zhu, D., Glownia, J., Kozina, M., Lemke, H., Sikorski, M., Feng, Y., Nelson, S., Saita, K., Stankus, B., Northey, T., Hastings, J. B. & Weber, P. M. (2015). Phys. Rev. Lett. 114, 255501.  CrossRef PubMed Google Scholar
First citationRichter, M., Gottwald, A., Kroth, U., Sorokin, A., Bobashev, S., Shmaenok, L., Feldhaus, J., Gerth, C., Steeg, B., Tiedtke, K. & Treusch, R. (2003). Appl. Phys. Lett. 83, 2970–2972.  CrossRef CAS Google Scholar
First citationRyutov, D. D., et al. (2009). Technical Report LLNL-TR-421318, Lawrence Livermore National Laboratory, Livermore, CA, USA.  Google Scholar
First citationSeibert, M. M. et al. (2011). Nature (London), 470, 78–81.  Web of Science CrossRef CAS PubMed Google Scholar
First citationShwartz, S., Fuchs, M., Hastings, J. B., Inubushi, Y., Ishikawa, T., Katayama, T., Reis, D. A., Sato, T., Tono, K., Yabashi, M., Yudovich, S. & Harris, S. E. (2014). Phys. Rev. Lett. 112, 163901.  Web of Science CrossRef PubMed Google Scholar
First citationYang, B., Wu, J., Raubenheimer, T. O. & Feng, Y. (2017). J. Synchrotron Rad. 24, 547–559.  CrossRef CAS IUCr Journals Google Scholar
First citationYoneda, H., Inubushi, Y., Nagamine, K., Michine, Y., Ohashi, H., Yumoto, H., Yamauchi, K., Mimura, H., Kitamura, H., Katayama, T., Ishikawa, T. & Yabashi, M. (2015). Nature (London), 524, 446–449.  Web of Science CrossRef CAS PubMed Google Scholar
First citationYoung, L., Kanter, E., Krässig, B., Li, Y., March, A., Pratt, S., Santra, R., Southworth, S., Rohringer, N., Dimauro, L., Doumy, G., Roedig, C. A., Berrah, N., Fang, L., Hoener, M., Bucksbaum, P. H., Cryan, J. P., Ghimire, S., Glownia, J. M., Reis, D. A., Bozek, J. D., Bostedt, C. & Messerschmidt, M. (2010). Nature (London), 466, 56–61.  CrossRef CAS PubMed Google Scholar
First citationZhu, D., Cammarata, M., Feldkamp, J. M., Fritz, D. M., Hastings, J. B., Lee, S., Lemke, H. T., Robert, A., Turner, J. L. & Feng, Y. (2012). Appl. Phys. Lett. 101, 034103.  Web of Science CrossRef Google Scholar

This is an open-access article distributed under the terms of the Creative Commons Attribution (CC-BY) Licence, which permits unrestricted use, distribution, and reproduction in any medium, provided the original authors and source are cited.

Journal logoJOURNAL OF
SYNCHROTRON
RADIATION
ISSN: 1600-5775
Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds