The recent development of neutron collimators with rectangular transmission profiles (intensity versus angular divergence) extends hope of improved count rates on neutron scattering instruments. It is usually assumed that a more effective use of beam angular spread in these devices should increase count rates by about a factor of two. However, real beams have both angular and wavevector spread and both these spreads are governed by the allowed collimation. In this extended view, the gains from ideal rectangular-profile elements (angle filters) are shown to be much larger (about a factor of four). The mirror reflections used to achieve the rectangular profiles in real devices complicate the resolution effects. Specifically, the reflections disturb the wavevector-angular divergence correlation in the beams, leading to unusual peak shapes characterized by triple peaks on powder diffractometers. Thus, these reflecting collimators are likely to be universally useful only before the monochromator and immediately preceding the detector, where wavevector-angle correlations have no effect. This reduces the potential gains to a factor of two or so. Note that the gains are as previously expected but for quite different reasons than imagined. This remains a very significant gain in a field where most work is intensity-limited.