Download citation
Download citation
link to html
To bridge the gap between traditional multilayer and crystal optics a high-resolution multilayer monochromator with a bandwidth of 0.22% has been designed and installed on a bending-magnet beamline (F3) at the Cornell High Energy Synchrotron Source (CHESS) to provide an unfocused monochromatic X-ray beam for protein crystallography experiments. Crystallographic data of excellent quality from a medium-sized protein, Concanavalin A, were collected and processed using standard crystallographic programs. The data were successfully used for a structure solution and refinement. The flux from the multilayer monochromator is enhanced, relative to that from a flat Si(111) monochromator, by a factor of 5; consequently, data collection is faster and/or smaller samples may be used. At the same time, the bandwidth is narrow enough to avoid streaked spots. This experiment suggests that multilayer optics may play a valuable role in satisfying the demands of the structural biology community for rapid X-ray data collection, particularly at under-utilized bending-magnet beamlines.

Follow J. Synchrotron Rad.
Sign up for e-alerts
Follow J. Synchrotron Rad. on Twitter
Follow us on facebook
Sign up for RSS feeds