Download citation
Download citation
link to html
The quick-scanning XAFS (QXAFS) method is achieved at the BL14W1 XAFS beamline at the Shanghai Synchrotron Radiation Facility based on the EPICS and LabVIEW systems. This is realised by the unprecedented use of LabVIEW's data logging and supervisory control module for communication with EPICS in synchrotron radiation facilities. A fine QXAFS spectrum with an energy range of 1.2 keV at the Cu K-edge has been collected in 2 s with stable beam position and the data quality is comparable with that of the step-mode XAFS spectrum. Analog-to-digital converter and double-crystal monochromator set-ups have been optimized in order to acquire optimal parameters for the QXAFS experiments. Signal-to-noise ratios of these spectra have been calculated in order to estimate the importance of these parameters.
Keywords: QXAFS; beamline; EPICS; LabVIEW; DSC.