A white-X-ray (Bremsstrahlung) four-circle diffractometer with a solid-state detector [Si(Li)] has been constructed. The system is operated by a small computer with magnetic disk memories, and enables energy-dispersive intensity measurements to be carried out with the X-rays selected from white X-rays. The present paper describes (1) how the present system can be used for energy-dispersive diffractometry even with the white radiation from a standard X-ray tube, and (2) how this system allows the measurement of the integrated intensities with almost the same accuracy as in angle-dispersive diffractometry with characteristic X-rays.