Depth capabilities of neutron and synchrotron diffraction strain measurement instruments. I. The maximum feasible path length

aManchester Materials Science Centre, Grosvenor Street, Manchester M1 7HS, UK
*Correspondence e-mail: philip.withers@man.ac.uk

(Received 18 July 2003; accepted 25 May 2004)

In this paper an algorithm is presented for estimating the maximum feasible penetration path length for neutron and synchrotron X-ray strain measurement instruments. This reflects the attenuation and scattering capability of the material under examination, the incident flux and detector arrangement, the likely background signal, the required strain measurement accuracy, the sampling volume and the diffracting geometry. Its validity and generality is examined through a consideration of data collected using a number of instruments on a variety of materials. Two criteria for the maximum feasible path length are examined: one based on the maximum acquisition time, the other based on the minimum acceptable peak height to background ratio. As demonstrated in the companion paper [part II: Withers (2004). J. Appl. Cryst. 37, 607–612], the algorithm can be used to delineate those conditions under which neutron and synchrotron X-ray radiations can provide useful information and to identify which is most suited to any particular measurement task.

1. Introduction

Since access to neutron and synchrotron X-ray facilities is limited, one needs to consider the advantages of these methods over existing, more readily available and cheaper strain measurement techniques very carefully. Their unique selling point over other methods is the ability to provide residual strain information as a function of depth non-destructively. As a result, it is useful to consider the depth limitations for these radiations from the viewpoint of undertaking strain measurement. Normally the desired strain measurement accuracy is specified or predetermined by the engineering problem of interest. The largest allowable sampling gauge volume is determined by the spatial resolution required and the question is then how long will it take to make a measurement at a given depth, and how deep can measurements be made before they become prohibitively time-consuming? Furthermore, it is advisable to ask whether the task is better suited to laboratory X-rays, neutrons or synchrotron X-rays. Of course, in practice the answers to these questions are partly sample dependent and partly instrument dependent.

Sample-dependent effects often play an important role in determining whether it is practical to make a strain measurement. These effects include grain size, prior plastic work and texture effects. Such factors have been studied in some detail previously and will be covered only briefly here. Grain size and the degree of crystal perfection determine the mosaic size (Hirsch, 1956[Hirsch, P. B. (1956). Met. Phys. 6, 236-339.]), which in turn influences the minimum practical sampling volume for which conventional powder diffraction methods of analysis can be applied. Powder methods need a large number of individual diffracting regions. Whether this condition is met is a complex function of grain size, plastic work, incident-beam divergence and the angular discrimination of the detector (Andrews & Johnson, 1959[Andrews, K. W. & Johnson, W. (1959). Brit. J. Appl. Phys. 10, 321-325.]). Furthermore, these factors also affect the onset of primary and secondary extinction, which can limit the relative strengths of the incident and diffracted beams (Sears, 1989[Sears, V. F. (1989). Neutron Optics. Oxford University Press.]). In this paper, it is assumed that the sample is sufficiently fine grained that powder methods can be applied. The crystallographic texture is a measure of the degree to which the grains in a polycrystalline sample are not randomly oriented. Since manufacturing processes generally involve plastic deformation, and this has the effect of modifying the crystallographic orientation of the grains, engineering components are often textured to some extent. This can have a very strong effect on the strength of a given hkl diffraction peak measured in a particular direction. As a result, the peak may, in certain circumstances, vary by two or three orders of magnitude according to the measurement direction. The study of crystallographic texture is a well developed field with many applications in materials science which have been dealt with recently in the textbook by Kocks et al. (1998[Kocks, U. F., Tomé, C. N. & Wenk, H. R. (1998). Texture and Anisotropy. Cambridge University Press.]). This aspect will not be covered in this paper and it will be assumed that the material is untextured. In cases where there is strong texture, it may not be feasible to measure some peaks at the feasible path lengths calculated in this paper, but on the other hand this means that other peaks will probably be more intense than these simple estimates would suggest. Finally, at certain special wavelengths, incident-beam penetration may vary sharply due to the influence of Bragg edges. These sharp increases in transmission occur at wavelengths which just exceed twice the lattice spacing of important reflections in the polycrystal. At this point, the Bragg condition reaches 180° and Bragg diffraction is no longer possible for that reflection, causing the transmissivity to increase sharply (Santisteban et al., 2001[Santisteban, J. R., Edwards, L., Steuwer, A. & Withers, P. J. (2001). J. Appl. Cryst. 34, 289-297.]).

The above material-dependent factors mean that it is not possible to predict the diffracted signal precisely in any given situation without a detailed knowledge of the material's microstructure. Only rarely is all this microstructural information available; nevertheless, a general means of checking out what is likely to be practicable prior to costly experimentation and characterization is required. In this paper, a simple rigorous approach for calculating the maximum feasible path length is presented based on instrumental characteristics and engineering requirements, such that the best approach can be identified. It utilizes the expression for the uncertainty in diffraction peak position given by Withers et al. (2001[Withers, P. J., Daymond, M. R. & Johnson, M. W. (2001). J. Appl. Cryst. 34, 737-743.]). It is a systematic extension of the empirical concept of the maximum economic thickness developed by Webster, Wang et al. (1996[Webster, P. J., Wang, X. D. & Mills, G. (1996). Strain Scanning using Neutrons and Synchrotron Radiation, ECRS 4, Cluny, edited by S. Denis, pp. 127-134. Soc. Fr. Met. Mater.]) specifically for neutron diffraction.

As discussed in part II (Withers, 2004[Withers, P. J. (2004). J. Appl. Cryst. 37, 607-612.]), the power of the method lies in providing a benchmark for estimating the feasibility of running specific experiments on a given instrument. It provides a way to make reasoned choices and to compare different measurement strategies. In addition, it provides a framework for optimizing and monitoring the performance of existing, or for designing new, strain measurement instruments.

2. Attenuation length as a guide to penetration depth

Of course, one of the most important parameters governing the maximum feasible depth at which measurements can be made is the extent of attenuation with path length. It is a simple matter to calculate the thickness of a given material that will attenuate 63% of the incident beam (see Table 1[link]), called here the attenuation length (lμ). It has long been realised that the large attenuation lengths of neutrons relative to laboratory X-rays gives them advantages for measurements deep within engineering materials. It is also evident from Table 1[link] that at energies in excess of 150 keV, the attenuation lengths of X-ray photons are comparable with neutrons for many materials.

Table 1
Approximate attenuation lengths (lμ) (mm) for investigation by laboratory X-rays, synchrotron X-rays [Elleaume (1999[Elleaume, P. (1999). Excel Macro for Computing X-ray Cross Sections, http://www.esrf.fr/machine/groups/insertion_devices/Codes/Xray/Xray.html .]) using an approach described by Chapman et al. (1988[Chapman, D., Gmur, N., Lazarz, N. & Tomlinson, W. (1988). Nucl. Instrum. Methods A, 266, 191-194.])] and neutrons (Hutchings & Windsor, 1987[Hutchings, M. T. & Windsor, C. G. (1987). Industrial Applications. Methods of Experimental Physics, Vol. 23C, edited by K. Sköld & D. L. Price, pp. 405-482. New York: Academic Press.])

      Approximate attenuation length (lμ) (mm)
  Energy (keV) Wavelength (Å) Al Ti Fe Ni Cu
Thermal neutrons 2.5 × 10−5 1.80 96 18 8 5 10
ID 15 (ESRF) 150 0.08 39 14 7 5 5
ID 31 (ESRF) 60 0.21 13 3 1.1 0.8 0.7
ID 11 (ESRF) 49 0.25 10 2 0.7 0.5 0.4
BM16 (ESRF) 38 0.32 6.5 1 0.4 0.2 0.2
16.3 SRS 31 0.40 3.3 0.5 0.16 0.11 0.10
Laboratory (Cu Kα) 8.05 1.54 0.076 0.011 0.004 0.023 0.021

While the attenuation length gives an indication of the maximum depth/thickness from which it is feasible to obtain information for a given material, this does not provide the complete picture. On the whole, fluxes at neutron sources are orders of magnitude lower than laboratory X-ray photon fluxes, while synchrotron X-ray sources are orders of magnitude more intense. As a result, it may be possible to accept lower transmission probabilities, because the detected signal may still be comparable. Furthermore, the diffracting angles, typical background levels, detector configurations, sampling gauge volumes and peak widths characteristic of X-ray, synchrotron X-ray and neutron diffraction are quite different, making side-by-side comparisons difficult. As a result, it is important to consider the diffracted signal necessary for achieving the desired strain measurement accuracy and the rate at which it is attained. That is the purpose of this paper. The consequences for real experiments and instruments are investigated in part II.

3. The necessary signal

3.1. Negligible background

Clearly, the number of diffracted particles that must be detected (the total acquired signal N) depends on the required measurement accuracy. Furthermore, a lower signal is required to determine the position of a narrow diffraction peak than a broad one. In order to establish the maximum useful depth of penetration, a quantitative relationship is required between the stipulated strain measurement uncertainty Err() and the necessary signal.

The elastic strain is determined from the shift in the position of the diffraction peak either in terms of the diffracting angle (in radians) (θθ0) or wavelength (λλ0) according to whether a monochromatic beam or a white beam is used, where θ0 and λ0 are the corresponding unstrained values:

[{\rm monochromatic} \,\, \, \varepsilon = - \cot \theta (\theta-\theta_{0}) \eqno (1a)]

and

[{\rm white\,\,beam} \,\, \, \varepsilon = (\lambda-\lambda_{0})/\lambda ,\eqno (1b)]

from which uncertainty in strain is given by

[{\rm Err}\,(\varepsilon)^{2} = {\rm Err}\,(d)^{2}/d^{2} + {\rm Err}\,(d_{0})^{2}/d_{0}^{2}, \eqno (2)]

where d and d0 are the strained and strain-free lattice parameters and Err(d) and Err(d0) their respective uncertainties. It then follows that if d and d0 are very similar, which they always are, and their uncertainties are also similar, then the strain uncertainty is proportional to the uncertainty in d. In practice, the shift in lattice spacing is determined by fitting the diffraction peak to a Gaussian profile or other suitable functional form. Using the approach outlined by de Sivia (1996[Sivia, D. S. de (1996). Data Analysis - A Bayesian Tutorial. Oxford University Press.]), it has been shown that when the background is negligible these wavelength or angular shifts can be determined to a precision of Err(λ) or Err(θ), respectively, where (Withers et al., 2001[Withers, P. J., Daymond, M. R. & Johnson, M. W. (2001). J. Appl. Cryst. 34, 737-743.])

[{\rm Err}\,(\lambda)^{2} = u^{2}_{\lambda} / N \quad {\rm or} \quad {\rm Err}\,(\theta)^{2} = u^{2}_{\theta} / N ,\eqno (3)]

where u is the standard deviation of the peak profile in λ or θ, and N is the integrated neutron or photon count. The validity of this relation has been shown for a range of experimental data by Webster & Kang (2002[Webster, P. J. & Kang, W. (2002). J. Neutron Res. 10, 93-110.]).

Combining equations (1)[link][link] and (3),[link] the uncertainty in is given by

[{\rm Err}\,(\varepsilon) = u_{\lambda} /[\lambda (N ^{1/2})] \quad {\rm or}\quad {\rm Err}\,(\varepsilon) = u_{\theta} \cot \theta / (N ^{1/2}).\eqno (4)]

We shall see that for the conditions characteristic of neutron diffraction, the number of counts required to achieve 1 × 10−4 strain uncertainty at 2θ = 90° on a peak of full width at half-maximum 0.7° (in 2θ), an integrated signal of around 650 counts is required (Table 4). For synchrotron X-ray sources, the very narrow peak widths [e.g. ∼0.01° (in 2θ) for ID31 at the ESRF] mean that in theory very few photons are required to achieve 10−4 strain accuracy (see Table 4). In practice, one would probably wish to have a somewhat larger number of integrated counts, but certainly at least an order of magnitude less than required to achieve the same accuracy relative to neutron sources.

From equation (4)[link], the time (t) required to achieve a given strain uncertainty is

[t \simeq u^{2}_{\lambda} / [\lambda^{2}\,{\rm Err}\,(\varepsilon)^{2}n] \quad {\rm or} \quad t \simeq \cot^{2} \theta u^{2}_{\theta} / [{\rm Err}\,(\varepsilon)^{2}n], \eqno (5)]

where n is the integrated diffraction peak count recorded per unit time (N = nt).

3.2. Significant background

Increasing levels of background increase the signal required to achieve a given accuracy (Withers et al., 2001[Withers, P. J., Daymond, M. R. & Johnson, M. W. (2001). J. Appl. Cryst. 34, 737-743.]). This becomes increasingly important when making measurements at depth because of the long data acquisition times associated with weak signals, which mean that background levels can be significant. It has been shown analytically for Gaussian peak shapes that, for the same integrated diffraction peak count, the strain uncertainty increases by a penalty factor [1 + 2(21/2)B/Hhkl], where Hhkl is the height of the hkl peak and B the background level (note that Hhkl = hhklt, B = bt) (Withers et al., 2001[Withers, P. J., Daymond, M. R. & Johnson, M. W. (2001). J. Appl. Cryst. 34, 737-743.]):

[\eqalign{{\rm Err}\,(\varepsilon) =\hskip.2em& u_{\lambda}/\lambda \{N/[1 + 2 (2^{1/2})b /h^{hkl}]\}^{1/2} \quad {\rm or}\cr {\rm Err}\,(\varepsilon) =\hskip.2em& \cot \theta\, u_{\theta}/ \{N/[1 + 2 (2^{1/2})b /h^{hkl}]\} .}\eqno (6)]

Alternatively, the time t to achieve the same accuracy increases by the penalty factor [1 + 2(21/2)b/hhkl]:

[\eqalign{t =\hskip.2em& u^{2}_{\lambda} [1 + 2 (2^{1/2}) b/h^{hkl}] / n \,{\rm Err}\,(\lambda)^{2} \quad {\rm or}\cr t =\hskip.2em& u^{2}_{\theta} [1 + 2 (2^{1/2}) b/h^{hkl}] / n \,{\rm Err}\,(\theta)^{2} ,}\eqno (7)]

or

[\eqalign{t \simeq\hskip.2em& [1 + 2 (2^{1/2}) b/h^{hkl}] u^{2}_{\lambda} / [ \lambda ^{2}\, {\rm Err}\,(\varepsilon) ^{2} n]\quad {\rm or} \cr t \simeq\hskip.2em& [1 + 2 (2^{1/2}) b/h^{hkl}] u^{2}_{\theta} \cot^{2} \theta / [{\rm Err}\,(\varepsilon)^{2} n] .}\eqno (8)]

A large body of experimental strain measurement data has been collected for a round-robin study as part of the VAMAS TWA20 neutron strain measurement standardization project. The experimentally determined uncertainties are plotted against the diffracted peak height to background ratio in Fig. 1[link]. From an experimental perspective, it is interesting to note that the facilities have recorded data with a very broad spectrum of diffracted signal to background ratios (from 0.3 to 85). Despite the scatter in the data, the data are broadly in agreement with equation (6)[link].

[Figure 1]
Figure 1
The measured uncertainties normalized against equation (4)[link] for round-robin measurements on the two phases of an Al2O3/SiC nanocomposite sample performed by many neutron facilities across the world as part of the VAMAS TWA20 standardization exercise (all data fitted by E. C. Oliver) (Webster, 2001[Webster, G. A. (2001). Editor. Polycrystalline Materials - Determinations of Residual Stresses by Neutron Diffraction. Technology Trends Assessment. Geneva: ISO.]). Squares represent neutron data; open circles represent synchrotron data. The solid curve represents the function [1 + 2(21/2)b/hhkl]1/2.

4. Instrument performance

In this section, a method for obtaining performance indicators for neutron strain and synchrotron strain measurement instruments is presented. Throughout this paper, this analysis is applied, for illustrative purposes, to a generic neutron instrument (based on the NRU L3 instrument at Chalk River) and a variety of synchrotron instruments having different energies (60, 50, 40 and 30 keV). Unless specifically mentioned, it is assumed that the sampling gauge for neutrons is 40 mm3 and that for the synchrotron instruments is 1 mm3. The characteristic Bragg angles (θ) and peak widths (standard deviation uθ) assumed in the calculations are summarized in Table 4.

4.1. The detected signal

While instruments are often compared in terms of their incident flux, it is not always the case that the highest flux instruments acquire data at the fastest rates. Often, medium-flux instruments can out-perform high-flux instruments through having a more efficient detector arrangement.

In estimating the maximum feasible depth from which strain information can be obtained, i.e. the maximum depth beyond which it is not really practical to measure, one needs to estimate the signal detected as a function of depth. If we consider, for simplicity, a small cuboidal gauge cross section,1 then the detected signal varies according to the incident flux, the detector arrangement, gauge volume, scattering material and the scattering geometry. Of the N0 particles passing through the incident slit and incident on the sample, the number of particles diffracted, Ndiffhkl, from the sampling gauge volume, Vv = AvLv, involving a total path length l through the material of interest, is given by

[N_{\rm diff}^{hkl} = N_{0}P_{\rm transmit}(l) \,P_{\rm diff}^{hkl} L_{\rm v} =\Phi t A_{\rm v} \exp(- \mu l) \, P_{\rm diff}^{hkl} L_{\rm v}, \eqno(9)]

where Av is the area of the incident slit, μ is the linear absorption (attenuation) coefficient, Φ is the incident flux per unit area, t is the data acquisition time (s) and N0 is the number of particles incident on the sample, given by ΦtAv. Ptransmit(l) is the fraction of the incident signal transmitted a distance l through the material and Pdiffhkldl is the probability that a neutron/photon is diffracted on travelling a distance dl so that the probability of a particle being diffracted within the sampling gauge length is approximately PdiffhklLv. In the above relation, it is assumed that the probability of scattering or absorption over the sampling length Lv is small1 (i.e. Lvlμ). Equation (9)[link] can be reformulated in terms of the scattering by each scattering site. The atomic σcoh and the macroscopic Σcoh cross sections (see Table 2[link]) for coherent scattering are related by the total number of atoms per unit volume, natomsNc:

[\Sigma_{\rm coh} = n_{\rm atoms} N_{\rm c} \sigma_{\rm coh} = ( 1 / v_{0}) \sigma _{\rm coh}, \eqno (10)]

where Nc is the number of unit cells/volume, natoms is the number of atoms per cell and v0 is the average volume per atom. As a result, the total probability of a particle being diffracted by the sample taking into account all hkl, Pdiff, is given by ΣcohLv. For a single hkl reflection (Bacon, 1975[Bacon, G. E. (1975). Neutron Diffraction. Oxford: Clarendon Press.]),

[P_{\rm diff}^{hkl} = {{\lambda ^{3}}\over{4n_{\rm atoms}^{2} v_{0}^{2}}} \, \, {{m^{hkl} F^{hkl\,2}}\over{\sin \theta _{\rm B}^{hkl}}} ,\eqno (11)]

where [\theta _{\rm B}^{hkl}] is the Bragg angle for the hkl reflection, Fhkl 2 the structure amplitude factor per unit cell, and mhkl the multiplicity. For a simple metal, Fhkl 2 can be replaced by Mhklb2, where b = (σcoh/4π)1/2 is the coherent scattering length for the element, or the effective scattering length representing the constituent elements in a disordered alloy, and Mhkl describes the way the wavelets from atoms within the unit cell superpose and is crystal structure dependent. In which case, Fhkl 2 = Mhklv0Σcoh/4π.

Table 2
Macroscopic cross sections (mm−1) for various engineering materials calculated on the basis of the same databases as Table 1[link]

Note that lμ = (Σcoh + Σincoh + Σabsorb)−1. Σphoto is the cross section for creating a photoelectron and is analogous to Σabsorb for neutrons.

  Energy (keV) Σ Al Ti Fe Ni Cu
Neutron 1 × 10−5 Σcoh 0.009 0.0077 0.0971 0.1218 0.0634
    Σincoh 0.00005 0.0151 0.0033 0.0475 0.0044
    Σabsorb 0.00139 0.0345 0.0217 0.0410 0.03202
ID15 150 Σcoh 0.00152 0.0053 0.0125 0.0167 0.01707
    Σincoh 0.03451 0.0536 0.0934 0.109 0.10493
    Σphoto 0.001 0.014 0.046 0.069 0.074
ID31 60 Σcoh 0.009 0.030 0.069 0.092 0.094
    Σincoh 0.040 0.061 0.107 0.123 0.118
    Σphoto 0.026 0.250 0.760 1.117 1.194
ID11 49 Σcoh 0.012 0.041 0.095 0.126 0.129
    Σincoh 0.041 0.061 0.106 0.123 0.117
    Σphoto 0.047 0.435 1.304 1.920 2.050
BM16 40 Σcoh 0.018 0.061 0.139 0.185 0.189
    Σincoh 0.041 0.061 0.105 0.122 0.116
    Σphoto 0.096 0.857 2.54 3.74 3.96
SRS 31 Σcoh 0.029 0.098 0.22 0.294 0.3
    Σincoh 0.040 0.059 0.10 0.12 0.11
    Σphoto 0.237 2.03 5.95 8.7 9.2
X-ray (Cu Kα) 8.05 Σcoh 0.20 0.52 1.19 1.61 1.66
    Σincoh 0.03 0.04 0.06 0.06 0.06
    Σphoto 12.9 89.9 235 41.33 44.85
               
Density (g cm−3) 2.698 4.508 7.873 8.907 8.933
Atomic weight 26.98 47.9 55.85 58.71 63.54

Of course, not all diffracted particles are collected by the detector since it will occupy only a fraction of the total solid angle Ω. If the detector having efficiency %detect is ldetect high and at a distance r from the sample, then the fraction of particles diffracted from Vv towards, and detected by, the detector if correctly positioned at [\theta _{\rm B}^{hkl}] is given by

[P_{\rm detect} (\theta)2\Delta \theta = \% _{\rm detect} {{l_{\rm detect} } \over {2\pi r\sin 2\theta _{\rm B}^{hkl} }}\delta (2\theta _{\rm B}^{hkl} - 2\theta) ,\eqno (12)]

where δ is the Kronecker delta. In reality the diffracted signal is not all diffracted at [\theta _{\rm B}^{hkl}]. Instead, the Kronecker delta function could more realistically be replaced by a Gaussian profile of standard deviation 2uθ in 2θ:

[\eqalignno{P_{\rm detect} (\theta)2\Delta \theta =\hskip.2em& \% _{\rm detect} {{l_{\rm detect} } \over {2\pi r\sin 2\theta _{\rm B}^{hkl} }}\cr &\!\times\left \{{{1 \over {2u_\theta (2\pi)^{1/2} }}\exp \left[ { - {{(\theta - \theta _{\rm B})^2 } \over {2u_\theta ^2 }}} \right]} \right\}2\Delta \theta .\cr && \hfill\llap{(13)}}]

This means that the number of diffracted particles detected by a single detector, or detector bin of a position-sensitive detector (PSD), of angular width 2Δθ is given by

[\eqalignno{N_{\rm detect}^{ hkl} (2 \Delta \theta) =\hskip.2em& \Phi t A_{\rm v} P_{\rm transmit} (l) P_{\rm diff}^{hkl} L_{\rm v} P_{\rm detect} (\theta)2\Delta \theta\cr = \hskip.2em&(\Phi t A_{\rm v} ) [\exp(-\mu l)] \left [{{\lambda^{3} m^{hkl} F^{\,hkl\,2}}\over{4 n_{\rm atoms}^{2} v_{0}^{2} \sin \theta_{\rm B}^{hkl}}}\right] L_{\rm v} \cr &\!\times\bigg \{\% _{\rm detect} {{l_{\rm detect}}\over{2\pi r \sin 2\theta _{\rm B}^{hkl}}} {{1}\over{2u_{\theta} (2\pi)^{1/2}}} \cr &\!\times\exp \left [ -{1\over 2} {{(\theta - \theta _{\rm B})^{2}}\over{u_{\theta}^{2}}} \right ] \bigg\} 2 \Delta \theta.&(14)}]

It is worthwhile to consider the use of a PSD and a serial detector separately for a moment. For a serial detector, the effective angular width of the detector (2Δθdetect) is not obvious to the user, who is more concerned with the angular measurement interval (2Δθbin) and the time to measure a point tpt. The total measurement time then becomes t = tptnpts, where npts is the number of points recorded. However, since a serial detector has a fractional angular coverage of only 2Δθdetect/2Δθscan of the total scan, its performance in collecting the whole peak can simply be corrected by multiplying the efficiency by this factor (for a PSD this factor equals 1):

[\eqalignno{P_{\rm detect} =\hskip.2em& \%_{\rm detect} {{l_{\rm detect}}\over{2\pi r \sin 2\theta^{hkl}_{\rm B}}} {{2\Delta \theta_{\rm detect}}\over{2\Delta \theta_{\rm scan}}} \cr=\hskip.2em& \%_{\rm detect}^{\,\prime} {{l_{\rm detect}}\over{2\pi r \sin 2\theta^{hkl}_{\rm B}}} {{1}\over{2\Delta\theta_{\rm scan}}} .&\hfill\llap{ (15)}}]

In a sense, the serial detector can be thought of as an inefficient PSD and we can incorporate 2Δθdetect into the efficiency term [\% _{\rm detect}^{\,\prime}]. For neutrons F hkl 2 is independent of θ, while for X-rays it varies considerably with angle. For simplicity, let us model the scattering factor for X-rays as

[F^{\,hkl} = F_{\rm o}^{\,hkl} \{6^{-\sin\theta/\lambda} + (A/200) \,[1-\exp(-\sin\theta/\lambda)]\} \eqno (16)]

(with λ in Å), A representing atomic number and Fo hkl the scattering for θ = 0. While less accurate than conventional polynomial forms (Krawitz, 2001[Krawitz, A. D. (2001). Introduction to Diffraction in Materials Science and Engineering. New York: John Wiley.]), the error is considerably less than 20% over the whole range of θ for Al and Cu, and most materials between, and avoids the need for look-up tables and element-specific coefficients.

4.2. The instrumental detected flux constant

In the expression for the detected count integrated over the diffracted peak, Ndetecthkl, ΦPdetect concerns the instrument and VvPtransmit(l)Pdiffhkl is sample dependent. Since

[\Phi P_{\rm detect} = \Phi \% _{\rm detect}^{\,\prime} {{l_{\rm detect}}\over{2\pi r\sin 2\theta_{\rm B}^{hkl}}} {{1}\over{2\Delta \theta _{\rm scan}}} ,\eqno (17)]

we can characterize the performance of the instrument by grouping together the fixed terms to define an instrumental flux constant Φinstr:

[\Phi_{\rm instr} = \Phi \% _{\rm detect}^{\,\prime} {{l_{\rm detect}}\over{2\pi r}} = \Phi \%_{\rm detect}{{l_{\rm detect}}\over{2\pi r}}2\Delta_{\rm detect} , \eqno (18a)]

of which Φ, ldetect, %detect and 2Δθdetect are difficult to quantify, but Φinstr can also be expressed using equation (14)[link] as

[\Phi_{\rm instr} = \left [{{N_{\rm detect}^{hkl}}\over{V_{\rm eff}t}} {{4n_{\rm atoms}^{2} v_{0}^{2} \sin \theta _{\rm B}^{hkl} \sin 2\theta _{\rm B}^{hkl}}\over{\lambda^{3} m^{hkl} F^{\,hkl \,2} \exp(-\mu l)}} \right] 2\Delta\theta_{\rm scan} ,\eqno (18b)]

all of which can be measured.

Here the sampling volume Vv has been replaced by the effective sampling volume Veff. This is because the scattering is not strictly proportional to Vv. Veff takes into account two factors: firstly the attenuation across the gauge volume, which means that the whole volume is not illuminated to the same extent, and secondly, the fact that the gauge may only be partially filled. Veff would equal Vv if the attenuation coefficient were negligible. One way of establishing the detected flux constant for a given instrument is simply to use published instrumental flux data along with measurements/estimates of detector efficiency and capture angles using equation (18[link]a). However, these parameters are often either unavailable or unreliable. A more pragmatic approach is to use a standard calibration sample to evaluate the term experimentally using equation (18b[link]). Another useful benchmark is nA1311(x), the number of particles that would be detected in normal reflecting geometry from 1 mm3 of Al (311) in 1 s at a depth x.

4.3. Determination of instrumental constants

Since the effect of background is important in determining instrument performance, rather than employ a small instrument calibration sample, it was decided to use surface scans in reflection geometry on standard test blocks to determine the representative detected flux constant for various instruments. One such surface scan is shown in Fig. 2[link]. A number of important well known characteristic features of the intensity curve should be noted. Firstly, the intensity rises initially as more and more of the gauge enters the surface (this is because Veff increases); then the recorded intensity decreases approximately exponentially due to increasing attenuation once the gauge is completely submerged. Less well catalogued in the literature is the variation of the background count which is composed of an approximately constant term and an exponentially decreasing term.

[Figure 2]
Figure 2
The characteristic integrated (311) peak count rate, ndetecthkl (solid circles), peak height, hhkl (crosses), and background rate, b (open circles), as a function of depth upon entering a 4.2 mm thick flat-walled can containing Ni powder measured in reflection at the NRU reactor, Chalk River, Canada, for a 20 × 0.5 × 0.5 mm sampling gauge (Browne, 2001[Browne, P. A. (2001). PhD thesis, University of Salford, UK.]). The sudden rise in diffracted intensity at zero depth and the decrease at 4.2 mm are due to the gauge becoming partially filled. The corresponding model curves are also shown.

Clearly there are a number of ways in which Φinstr can be extracted from curves such as that in Fig. 2[link]. In reality, the actual value of Φinstr recorded will also vary according to the diffraction peak, or peaks, collected, as well as sample texture, but if a single standard calibration sample is used across a number of instruments, a comparative measure can be obtained. Besides, given the logarithmic dependence of the maximum path length on the instrumental flux and the inexact nature of the calculation, guideline values are sufficient.

Approximate values of Φinstr have been inferred using surface entrance scans similar to that in Fig. 2[link] (mostly collected for the 311 reflection using a standard Al sample) at a wide range of neutron and synchrotron sources. At a given instrument, the integrated intensity of the diffraction peak has been determined using the accepted method (scanning detector with or without analyser, PSD, energy-dispersive detector, etc.). The values of Φinstr and nA1311 (0) are shown in Table 3[link] using the data in Table 2[link]; the value of Φinstr recorded using instrument 16.3 at the SRS is large compared with that from BM16 because no analyser crystal was used in the case of the former. In the calculations, the diffraction angles and peak widths summarized in Table 4 were used. Because deep measurements at neutron sources usually involve much larger sampling gauges than are typical either for near-surface neutron or synchrotron X-ray measurements, a neutron sampling gauge of 40 mm3 has been assumed throughout this paper. Note that the integrated diffraction peak count rates at neutron sources are very small [nAl311 (0) = 3 neutrons every second for a 40 mm3 sample] whereas for synchrotron instruments they can be 104 times greater for the same sampling volume.

Table 3
Preliminary values of the instrumental detected flux, Φinstr, and the background constants, β0, βsample, extracted from entrance curves in reflection geometry

The particle count rates b0, bsample, bincoh, nAl311 (0) and hAl311 correspond to the various background and signal components that would be recorded in reflection in 1 s for 1 mm3 (40 mm3 for neutrons) sampling volume of Al at the surface. The peak profile constants expressed in counts can be converted to counts/degree by dividing by the bin width (2Δθbin).

  2Δθbin (°) Φinstr (counts m−2 s−1) β0 (counts m−2 s−1) βsample (counts m−2 s−1) b0 (counts s−1) bsample(0) (counts s−1) bincoh(0) (counts s−1) nAl311(0) (counts s−1) hAl311(0) (counts s−1) h/b
Neutrons 0.084 2.0 × 106 1 × 105 5 × 106 1 × 10−2 5 × 10−2 6 × 10−5 3 0.4 6
ID31 0.0010 2 × 1010 3 × 108 3 × 1011 9 70 0 1000 130 2
ID11 0.0005 5 × 1010 1 × 108 1 × 1010 10 9 × 10−1 0 2300 60 4
BM16 0.0033 8 × 108 4 × 105 2 × 1010 2 × 10−1 5 0 50 10 2
16.3 0.0100 7 × 109 3 × 107 1 × 1011 20 90 0 1000 425 4

4.4. The characteristic background flux

Equation (6)[link] dictates that the background flux is also an important factor in determining the achievable strain measurement accuracy, especially when the signal is low. Some level of background is unavoidable in that it is related to the scattering capability of the sampling volume and it cannot be removed by analysers. Incoherent neutron scattering from the sampling gauge volume comes into this category. The incoherent neutron scattering from Vv is given by

[P_{\rm incoh} L_{\rm v} = \Sigma_{\rm incoh} L_{\rm v}, \eqno (19)]

which means that the number of incoherently scattered particles detected is

[ N_{0}P_{\rm transmit} L_{\rm v} P_{\rm incoh} P^{\,\prime}_{\rm detect},]

noting that because the incoherent scattering is isotropic, the probability of capture by the detector ([P^{\,\prime}_{\rm detect}]) is lower than for the diffracted radiation, giving a coherent peak to incoherent background ratio of

[\eqalignno{{h^{hkl}\over b_{\rm incoh}} =\hskip.2em& { \lambda ^{3} \over 2n_{\rm atoms}^{2} v_{0}^{2}} { m^{hkl} F^{\,hkl\,2} \over \sin \theta _{\rm B}^{hkl} \sin 2\theta_{\rm B}^{hkl} \Sigma _{\rm incoh}} { 1 \over 2u_{\theta} (2\pi)^{1/2} }\cr =\hskip.2em& {\Sigma _{\rm coh} \over \Sigma _{\rm incoh} } { \lambda ^{3} m^{hkl} M^{hkl} \over 8\pi n_{\rm atoms}^{2} v_{0} \sin \theta _{\rm B}^{hkl} \sin 2\theta _{\rm B}^{hkl}} { 1 \over 2u_{\theta} (2\pi)^{1/2}}.&{(20)}}]

In other words, the broader the peak, the poorer the inherent neutron signal to background ratio. It is clear from Fig. 2[link] that for Ni (indeed for most materials), the incoherent background contribution is small. This is because many other sources of background exist. Air scattering can be significant, especially for synchrotron X-rays, and can enter the detector through the diffracted beam slits unless an analyser crystal is used. At elevated temperatures, scattering into the background arising from the Debye–Waller factor can also be important. Other sources of background include multiple scattering. This can occur within the sampling gauge if Lv/lμ is large (which it rarely is), or within the component if the sample dimensions are large compared with lμ (which they often are). An analyser crystal is most effective at precluding the latter.

In this paper, the many sources of background signal which are important at neutron and synchrotron sources are categorized into aperture-independent, sample-dependent and instrument-dependent background components. The first type of background signal is recorded from sources other than the sample, for example from very penetrating particles that can penetrate the shielding around the detector (e.g. fast gammas, high-energy X-rays, etc.), or electronic noise recorded by a CCD or PSD detector. This is usually a small contribution and is neglected here. The remainder enters the detector through the sampling gauge volume-defining aperture and is dependent on background scattered from the sample and from the local environment. These types of background would tend to zero as the aperture is closed. In order to characterize all the various contributions fully, a comprehensive study would be required before embarking on an experiment. What is needed here is a pragmatic approach which includes the major sources of background in a simple way, enabling estimates of the background to be made without the need for time-consuming experimentation on a case-by-case basis.

Analysis of many strain measurement scans, such as Fig. 2[link], suggests that the background can be conveniently represented by a constant term which is recorded even when the sampling volume is outside the sample, B0 (= b0t), combined with a sample-dependent term. Here both are taken to depend on the detector aperture, in that it is assumed that when the detector slit is closed, no signal is recorded (i.e. in many cases aperture-independent scattering can be neglected). The instrument-dependent term, B0, should probably be proportional to the detector aperture area; however, to keep the number of parameters to a minimum, it is probably sufficient to make this term Vv2/3 dependent. As is evident from Fig. 2[link], a large component of the background signal varies according to the depth of the sampling volume within the material and decays approximately exponentially with penetration distance. While this includes the incoherent scattering component (Bincoh), it is much greater than would be expected from this source alone. The extra contribution, Bsample (= bsamplet), is probably the result of multiple scattering and has thus been taken here to be proportional to Vv and to (Σcoh + Σincoh)2/Σtot (Bacon, 1975[Bacon, G. E. (1975). Neutron Diffraction. Oxford: Clarendon Press.]). In this way, the overall background count recorded at a given path length for a given material can be well represented by

[\eqalignno{B =\hskip.2em& B_{\rm incoh} + B_{\rm sample} + B_{0}\cr =\hskip.2em& \bigg\{ {1\over 2} \Phi_{\rm instr} \Sigma_{\rm incoh} \exp(-\mu l) V_{\rm v} + \bigg[\beta _{\rm sample} \exp(-\mu l) V_{\rm v} \cr &\!\times { (\Sigma_{\rm coh} + \Sigma_{\rm incoh} )^{2} \over \Sigma_{\rm tot}} \bigg] + \beta_{0} V_{\rm v}^{2/3} \bigg\} t { 2\Delta \theta _{\rm bin} \over 2\Delta \theta_{\rm scan}}, & \hfill\llap{(21)}}]

where βsample and β0 are the background constants and 2Δθbin represents the bin width for both a PSD and a serial detector. Note that here the degradation in the maximum feasible penetration length arising from the incoherent neutron scattering contribution due to an increase in background is predicted to be weaker than the Σcoh/(Σcoh + Σincoh) benchmark figure proposed by Webster, Mills et al. (1996[Webster, P. J., Mills, G., Wang, X. D., Kang, W. P. & Holden, T. M. (1996). J. Neutron Res. 3, 223-240.]). Because the incoherent contribution can be predicted from the instrumental flux and the incoherent scattering cross section, only three parameters (β0, βsample and Φinstr) need be determined for any given instrument configuration from the calibration depth scan. Typical values of β0, βsample, B0, Bsample and Bincoh are summarized in Table 3[link].

5. Experimental validation

The preceding sections have defined simple expressions for the variation of diffracted signal and background level as a function of material, diffraction peak, instrumental arrangement and sampling gauge volume and depth, as well as a procedure for establishing the key instrumental constants. The validity of the approach has been tested against a very wide range of neutron and synchrotron instruments and many engineering materials. In Fig. 3[link], the performance of the algorithms for predicting the integrated count and background levels are shown for the ID31 beamline. The experimental data were collected over a number of visits and the same values of the three instrumental constants were used in all cases. Note the large synchrotron count rates per unit sampling volume compared with the corresponding neutron values in Fig. 2[link]. This means that it is feasible to go to larger multiples of the attenuation length (see part II; Withers, 2004[Withers, P. J. (2004). J. Appl. Cryst. 37, 607-612.]). The fits are sufficient for our purposes (within a factor of 10), given that the primary aim was not to model precisely the diffraction peak and background behaviour; that would require a great deal of information about the sample texture, grain size, etc., and the exact experimental configuration. Instead, the aim was to make acceptable `ball-park' predictions based on just three instrumental parameters and basic materials scattering data. In this way, the measurement capability for a given problem can be estimated a priori. Note that beyond a normalized distance of around l/lμ = 6, the integrated count over the whole peak becomes less than the background signal recorded at a single point; this would be an extreme limit on the maximum depth that can be probed.

[Figure 3]
Figure 3
The integrated intensity (solid symbols) and background (open symbols) count rates (normalized by Vv) and the associated predictions (continuous and dashed lines, respectively) for depth scans on (a) Al (311), (b) Ti (10[\bar{1}]1), (c) Fe (200) and (211), and (d) Ni (111) recorded in reflection geometry on ID31 at 60 keV (ESRF). The data represented by square symbols were collected during the same beam period and are modelled by the bolder lines. The other data were collected at various times during the previous 15 months. The Φinstr, β0 and βsample instrumental parameters were obtained from the Ni data. The path lengths have been normalized by the respective attenuation lengths.

6. The maximum feasible path length

Combining equations (6)[link] and (18)[link][link], the signal needed to achieve a given strain uncertainty Err() for a conventional θ–2θ scan is given by

[\eqalignno{N_{\varepsilon} = \hskip.2em&\Phi_{\rm instr} t_{\varepsilon} V_{\rm eff} \exp(-\mu l) {{\lambda ^{3}}\over {4n_{\rm atoms}^{2} v_{0}^{2}}} {{m^{hkl} F^{\,hkl\,2}}\over{\sin \theta _{\rm B}^{hkl} \sin 2\theta _{\rm B}^{hkl} 2\Delta \theta _{\rm scan}}}\cr =\hskip.2em& \left[1 + 2(2^{1/2}) {{b}\over{h^{hkl}}} \right] {{\cot ^{2} \theta _{\rm B}^{hkl} u_{\theta}^{2}}\over{{\rm Err}\,(\varepsilon)^{2}}} &\hfill\llap{(22)}}]

where b/hhkl is given by

[\eqalignno{{b\over h^{hkl}} \!=\! \hskip.2em&\left [{\textstyle{1\over 2}} \Phi _{\rm instr} \Sigma _{\rm incoh}\! +\! \beta _{0} V_{\rm v}^{-1/3}\exp(\mu l) \!+ \!\beta _{\rm sample} {(\Sigma _{\rm coh}\! +\! \Sigma _{\rm incoh} )^{2} \over \Sigma _{\rm tot} }\right]\cr & \times{{\sin \theta _{\rm B}^{hkl} \sin 2\theta _{\rm B}^{hkl} 16 n_{\rm atoms}^{2} \pi v_{0}}\over{\Phi _{\rm instr} \Sigma _{\rm coh} \lambda ^{3} m^{hkl} M^{hkl}}} 2u_{\theta} (2\pi)^{1/2}.&\hfill\llap{(23)}}]

Strictly there is no absolute value of the path length beyond which it is not possible to make a measurement to a given accuracy. However, the time required to obtain sufficient diffracted signal relative to the background level becomes uneconomically long as the path length becomes greater. As a result, it is useful to define a maximum feasible path length. At least two criteria can be sensibly used to arrive at such a figure; namely a maximum acceptable time for the taking of a measurement, and a minimum diffracted peak height to background ratio.

6.1. Maximum acceptable acquisition time

The results of the above discussion of strain measurement accuracy can be used to assess the time required, t, to achieve a specified uncertainty, Err() from equation (8)[link]. This gives the well known result that the acquisition time is inversely proportional to the square of the required uncertainty. In other words, it takes four times as long to achieve twice the strain accuracy. The equation also highlights the need to minimize the background, especially when the signal intensity is low, such as, for example, when making measurements deep within a component.

Inserting equations (22)[link] and (23)[link] into equation (8)[link], we see that the well known observation that the time should increase exponentially with increasing depth is only true for the case of zero background (see the straight line response in Fig. 4[link]). This situation is not at all appropriate for the conditions representative of deep measurements. It is also evident from the figure that were the background to remain constant at the level recorded near the surface, the time to measure to a given strain would be considerably longer than that predicted on the basis of no background. However, in reality, the decreasing level of background observed with increasing depth of the type shown in Fig. 2[link] means that the actual picture is not so bleak. At shallow depths, the background decreases approximately exponentially at broadly the same rate as the diffracted signal, so that the time to achieve a given strain lies parallel to the zero background limit curve. At larger depths, the constant component of the background level becomes important and the curve remains lower than, but parallel to, the constant background curve (see Fig. 4[link]). For the acquired data, the depths at which 10−4 strain uncertainty is achievable in 60 min for a 40 mm3 Ni gauge are 9.1, 7.0 (see Table 1 of part II) and 5.0 mm for no background, the expected background and that representative of the near-surface, respectively.

[Figure 4]
Figure 4
The time needed to attain 10−4 strain accuracy in reflection geometry based on representative neutron calibration data (acquired at Chalk River) for a 40 mm3 Ni gauge volume, with no background, or a background level equal to that recorded near the surface, compared with the decreasing background response predicted from the calibration data (Fig. 2[link]). The 60 min maximum feasible time threshold is marked by the dashed line.

In practice, measurement times as long as 24 h have been used for point measurement by neutron diffraction, but this is exceptional. Progressively shorter acquisition times are required for line scans (∼1 h), area scans (∼10 min) and volume scans (∼1 min) because of the increase in the number of measurements (typically ∼10, >100 and >500, respectively) required to complete them. In order to probe larger depths, large sampling volumes are more commonly used for neutron measurements and so, as mentioned earlier, in our calculations the sampling volume is taken to be 1 mm3 and 40 mm3 for synchrotron X-rays and neutrons, respectively. Furthermore, in practice `acceptable' measurement times are not normally as long for synchrotron strain measurements, which are usually of the order of minutes as against hours for neutrons. We will denote the maximum path length for which the required strain accuracy can be achieved in 1 h by lt. A willingness to accept an acquisition time of 24 h would naturally increase the feasible path length. However, because of the logarithmic scale, increasing the measurement time has only a limited effect on increasing the maximum path length.

The maximum feasible path length under the time-based criteria is obtained by solving a quadratic in exp(μl) with t = 3600 s, formed by inserting equation (23)[link] into equation (22)[link]. The maximum feasible penetration lengths calculated on the basis of the maximum acceptable time are listed in Table 4[link]. Except for the Al case, these lengths are in startling agreement with the empirical economic thicknesses (x) proposed by Webster (1996[Webster, P. J. (1996). Mater. Sci. Forum, 228, 191-200.]) for neutron strain measurements on the basis of practical experience (see Table 4[link]) expressed as

[b^2 \exp( - \mu x) [\sigma _{\rm coh} /(\sigma _{\rm coh} + \sigma _{\rm inccoh})] = 1 , \eqno (24)]

where b is the scattering length, and σcoh and σincoh are the coherent and incoherent cross sections, respectively. The final term in equation (24)[link] was developed largely to account for the practical difficulty associated with measuring at depth when examining Ti samples due to the increased level of background arising from the large incoherent scattering contribution. Without this factor, Webster's economic thickness for Ti would be ∼50 mm, which is in excess of that for Fe. The more rigorous incorporation of the effect of the increased incoherent background contribution adopted here automatically captures this effect.

Table 4
Estimated maximum penetration lengths, lt, for neutrons and synchrotron sources based on 10−4 strain measurement precision and 1 h measurement time, or the penetration length, lh/b, at which a signal to background ratio of 1 would be recorded, using the scattering angles and peak widths shown

Also shown for Al are the time, th/b, at which the lh/b criterion is met and the counts necessary to achieve 10−4 accuracy in the case of no background, Nb=0, and at the maximum path length using the 1 h time criterion, Nt. For the neutron data, a sampling gauge volume of 40 mm3 has been assumed, while a gauge of 1 mm3 has been used for the synchrotron measurements. Data recorded at the NRU LR3 instrument at Chalk River have been taken as representative of current neutron sources. The times and necessary signals correspond to those for the Al sample.

      Al (311) Ti (100) Fe (211) Ni (311) Cu (311)      
  θ (°) uθ (°) lt lh/b lt lh/b lt lh/b lt lh/b lt lh/b th/b(Al) (min) Nb=0 Nt
Webster     237   23   37   21   40        
Neutrons 45 0.147 160 185 22 24 40 43 20 21 36 39 90 650 2000
ID31 8 0.002 115 45 29 16 13 9 8 5 8 5 0.01 3 750
ID11 8 0.004 85 55 19 13 8 6 5 4 5 3 0.2 25 1600
BM16 10 0.004 50 45 9.4 9 4 4 2.5 2.4 2.4 2.3 15 15 90
16.3 SRS 10 0.005 30 20 4.5 3 1 1.5 1.2 0.9 1.1 0.8 0.3 22 900

It is instructive to examine the necessary signals, Nt, listed in Table 4[link]. For the generic neutron source, the necessary signal is less than four times that for the ideal case of no background at all (Nb=0). This is quite acceptable. For the synchrotron instruments, however, the necessary signals are very much larger than the ideal case. This is because the large incident flux allows one to measure at many times the attenuation length, where the peak to background ratio is poor but there is still sufficient diffracted intensity to measure strain. As a result, large time penalty factors [1 + 2(21/2)b/hhkl] mean that the data acquisition process is very inefficient and it is probably wise to use a signal to background ratio criteria leading to slightly shorter penetration depths, but much faster acquisition times.

6.2. Minimum acceptable signal to background ratio

As we have seen, at large depths the signal tends to become lost in the background. This makes it increasingly time consuming to achieve a specified strain accuracy with increasing depth. One option is to say that beyond a diffraction peak height to background ratio of 1, data acquisition becomes uneconomic, placing a limit of the maximum feasible path length, represented as lh/b. This corresponds to an increase in signal of four times that required were there no background at all in order to achieve the same level of uncertainty [equation (6)[link]]. This can be taken as the point at which the time penalty associated with a poorer signal to noise ratio at larger depths starts to become prohibitively steep (see Fig. 5 in part II). The maximum feasible path length under the signal-to-background ratio-based criterion (hhkl/b = 1) is obtained directly from equation (23)[link]:

[\eqalignno{l_{h/b} =\hskip.2em& l_\mu \ln \Bigg\{\Bigg[ \Phi_{\rm instr}{ \lambda^{3} \over 4v_{0}^{2} } { m^{hkl}F^{\,hkl\,2} \over \sin \theta_{\rm B}^{hkl} 2u_{\theta} (2\pi)^{1/2}} \cr &\! - \bigg\{ \beta _{\rm sample} \bigg[{(\Sigma_{\rm coh} + \Sigma_{\rm incoh} ) ^{2} \over \Sigma _{\rm tot} }\bigg] + {1\over 2} \Phi _{\rm instr} \Sigma _{\rm incoh} \bigg\}\Bigg]\cr &\!\times{1\over \beta_{0} V_{\rm v}^{-1/3} } \Bigg\} .& \hfill\llap{(25)}}]

Note that the maximum path length under this criterion does not depend on the specified level of uncertainty, which only determines the necessary acquisition time. The maximum feasible path length can also be considered graphically. In Fig. 5[link], the variation in background and peak heights per second are shown for Al for various X-ray energy instruments (Vv = 1 mm3) and a generic thermal neutron instrument (Vv = 40 mm3). The maximum feasible path lengths are shown by the intersection of the height and background curves; under the present conditions, the lengths are summarized in Table 4[link]. The rate at which the peak height develops is 10–1000 times faster for the X-ray instruments than expected for neutron diffraction near the surface for Al (which is a weak neutron scatterer). It is clear from the figure that in the X-ray cases the exponential fall off of the sample background components is considerable and allows measurements to be made to much greater depths than would be the case were the background to remain constant as a function of depth. The absence of an analyser crystal on the SRS (16.3) means that the signal intensity is very high given the relatively low incident flux available, but the background is also high. Despite the low signal to background ratio, the neutron measurements can be made to much larger absolute depths because of the low level of attenuation and the very low background counts. Of course, as Table 4[link] shows, the maximum path length is larger for radiations with the larger attenuation lengths. Fig. 5[link] provides insight into the limitations of current instruments. Under this criterion, the penetration distance is strongly determined by the background constant β0. By taking measures to reduce β0, the background and peak height curves remain parallel to greater depth and thus the cross-over of the curves is achieved at greater depths, giving better depth capability. Under this criterion, the performance of BM16 at the ESRF is particularly noteworthy, having an especially low β0 constant and thus being capable of measurements to depths in excess of 7 times the attenuation length, whereas all the other instruments achieve a value of approximately 2–6 times the penetration length.

[Figure 5]
Figure 5
The variation in peak height (h) (solid symbols, continuous lines) and background level (b) (open symbols, dashed lines) expected per second for a sampling gauge of 1 mm3 (40 mm3 for neutrons) of Al (311) for BM16 (circles) and 16.3 (triangles) synchrotron X-rays, and for neutrons (diamonds), as a function of normalized path length based on the angles and peak widths listed in Table 4[link] and the calibration data acquired for the neutron (Chalk River), 16.3 (SRS) BM16 and ID31 (ESRF) synchrotron instruments.

For neutron sources, this criterion would lead to larger depths than for the 1 h time-limited criterion, but, as Table 4[link] indicates, the times become prohibitively long and the time-based criterion is probably more appropriate. In contrast, for the synchrotron instruments, the peak to background criterion gives a shorter maximum penetration length. Given that 1 h would be considered an excessively long period to acquire data at a synchrotron source and the time to achieve 10−4 accuracy at the lh/b limit is much shorter, this is probably the most appropriate criterion for synchrotron measurements.

7. Conclusions

Of course, the performance capability of a strain measurement instrument is not static. Instrument scientists are always endeavouring to improve strain sensitivity and reduce acquisition times. As a result, the exact values presented here should not be taken too literally. They should be regarded as preliminary values which nevertheless provide useful benchmarks for deciding whether an experiment is feasible on any specific instrument. Furthermore, they provide a tool for comparing different measurement strategies and for designing new instruments. The implications are investigated in part II (Withers, 2004[Withers, P. J. (2004). J. Appl. Cryst. 37, 607-612.]). While this paper demonstrates the utility of calibration experiments to obtain instrument performance benchmarks, a more rigorous and systematic endeavour is required to measure and collate calibration data from instruments worldwide. To date, the background response as a function of depth has been characterized at a wide range of instruments and found in all cases examined to be exponentially dependent on the depth of the sampling gauge. This behaviour has not been explained at a fundamental level. This is necessary to modify more rigorously the predicted background level as a function of instrument configuration and component constitution and shape, and to point to ways of lowering background levels. This might require a detailed experimental and simulation study. It appears from this preliminary study that through surface depth scans in reflection, a quick and easy way of extracting the main calibration parameters (Φinstr, β0, βsample) is provided.

A comprehensive and reliable set of data measured on specific reference samples would be of great value. It would benchmark the current capability of different instruments, allowing a quantitative assessment of the improvements brought by hardware upgrades and modifications. It would enable those considering an experimental measurement scheme to evaluate the time required and the experimental feasibility. It could also be used as the basis for intelligent systems developed by facilities to enable users to plan experiments more accurately. This work has concentrated on single peak analysis. Further work is required to compare instruments that deliver complete diffraction spectra, such as time-of-flight neutron instruments and energy-dispersive X-ray instruments, with those that provide individual diffraction peaks one at a time.

This paper has demonstrated that it is not simply incident flux that determines the capability of an instrument. Often instruments based at weaker sources can shorten measurement times through intelligent use of detectors. Of equal importance when considering penetration capability is the background level. This aspect has been rather neglected at X-ray sources where the near-surface count rates are extremely high and the peak to background ratios large. As a result, there is much that remains to be done, analogous to that undertaken at neutron sources, to cut background levels and thus extend the distances into materials to which we can reasonably measure.

The concept of the feasible path length has been introduced within a rigorous framework. Because of the very high X-ray intensities available at third-generation sources compared with neutron sources, the maximum feasible path length in order to achieve 10−4 uncertainty is predicted to be around 6–8lμ for synchrotrons and around 2lμ for neutrons, in very good agreement with practical experience. For synchrotron instruments, it is suggested that a peak:background criterion may be more appropriate than a maximum acceptable time, because it conveniently marks the point of diminishing returns when low signal to background seriously affects the measurement time. Its value is independent of the desired measurement accuracy. For neutrons, the peak:background ratio criterion gives longer times and greater depths than the elapsed-time criterion because of the low background signals characteristic of most neutron instruments.

Footnotes

1Provided that the gauge dimension along the beam direction is significantly shorter than the attenuation length, the scattered signal is dependent only on the sampling volume Vv, independent of the gauge shape; otherwise the level of scattering is shape dependent. This can be important especially when considering X-rays, in which case Vv can be replaced by Veff to account for attenuation within the gauge. In the following discussion, it is the sampling volume rather than the shape which is important.

Acknowledgements

This paper could not have been written without the help of all those responsible for running neutron and synchrotron strain instruments worldwide. Special thanks are owed to Drs A. Fitch, J. Wright and M. R. Daymond for special help extracting raw data, D. Hughes, A. Evans, J. Fonseca and P. A. Browne for making their experimental data available, and C. Ohms for providing data from the VAMAS TWA20 project. Helpful discussions have been had with Professor P. J. Webster and Drs M. W. Johnson, M. R. Daymond and A. Steuwer. Funding from the European Community, EPSRC and a Royal Society Wolfson Merit Award is also gratefully acknowledged.

References

First citationAndrews, K. W. & Johnson, W. (1959). Brit. J. Appl. Phys. 10, 321–325.  CrossRef CAS Web of Science Google Scholar
First citationBacon, G. E. (1975). Neutron Diffraction. Oxford: Clarendon Press.  Google Scholar
First citationBrowne, P. A. (2001). PhD thesis, University of Salford, UK.  Google Scholar
First citationChapman, D., Gmur, N., Lazarz, N. & Tomlinson, W. (1988). Nucl. Instrum. Methods A, 266, 191–194.  CrossRef Web of Science Google Scholar
First citationElleaume, P. (1999). Excel Macro for Computing X-ray Cross Sections, http://www.esrf.fr/machine/groups/insertion_devices/Codes/Xray/Xray.htmlGoogle Scholar
First citationHirsch, P. B. (1956). Met. Phys. 6, 236–339.  CrossRef CAS Web of Science Google Scholar
First citationHutchings, M. T. & Windsor, C. G. (1987). Industrial Applications. Methods of Experimental Physics, Vol. 23C, edited by K. Sköld & D. L. Price, pp. 405–482. New York: Academic Press.  Google Scholar
First citationKocks, U. F., Tomé, C. N. & Wenk, H. R. (1998). Texture and Anisotropy. Cambridge University Press.  Google Scholar
First citationKrawitz, A. D. (2001). Introduction to Diffraction in Materials Science and Engineering. New York: John Wiley.  Google Scholar
First citationSantisteban, J. R., Edwards, L., Steuwer, A. & Withers, P. J. (2001). J. Appl. Cryst. 34, 289–297.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationSears, V. F. (1989). Neutron Optics. Oxford University Press.  Google Scholar
First citationSivia, D. S. de (1996). Data Analysis – A Bayesian Tutorial. Oxford University Press.  Google Scholar
First citationWebster, P. J. (1996). Mater. Sci. Forum, 228, 191–200.  CrossRef Google Scholar
First citationWebster, G. A. (2001). Editor. Polycrystalline Materials – Determinations of Residual Stresses by Neutron Diffraction. Technology Trends Assessment. Geneva: ISO.  Google Scholar
First citationWebster, P. J. & Kang, W. (2002). J. Neutron Res. 10, 93–110.  CrossRef CAS Google Scholar
First citationWebster, P. J., Mills, G., Wang, X. D., Kang, W. P. & Holden, T. M. (1996). J. Neutron Res. 3, 223–240.  CrossRef CAS Google Scholar
First citationWebster, P. J., Wang, X. D. & Mills, G. (1996). Strain Scanning using Neutrons and Synchrotron Radiation, ECRS 4, Cluny, edited by S. Denis, pp. 127–134. Soc. Fr. Met. Mater.  Google Scholar
First citationWithers, P. J. (2004). J. Appl. Cryst. 37, 607–612.  Web of Science CrossRef CAS IUCr Journals Google Scholar
First citationWithers, P. J., Daymond, M. R. & Johnson, M. W. (2001). J. Appl. Cryst. 34, 737–743.  Web of Science CrossRef CAS IUCr Journals Google Scholar

© International Union of Crystallography. Prior permission is not required to reproduce short quotations, tables and figures from this article, provided the original authors and source are cited. For more information, click here.

Journal logoJOURNAL OF
APPLIED
CRYSTALLOGRAPHY
ISSN: 1600-5767
Follow J. Appl. Cryst.
Sign up for e-alerts
Follow J. Appl. Cryst. on Twitter
Follow us on facebook
Sign up for RSS feeds