J. Synchrotron Rad. (2008). 15, 8-11 [ doi:10.1107/S0909049507044858 ]
Abstract: A simple variant of a Si internally cooled inclined X-ray monochromator of reasonable size is proposed. It has two inclined surfaces oriented into a V shape. This design substantially decreases the surface deformations introduced by radiation heat, and the size of the crystal is still feasible for a 50 mm broad impinging bending magnet or wiggler beam. The possibility of sagittal focusing of the diffracted beam is also discussed.
Keywords: inclined monochromator; heat load; internal cooling.
Copyright © International Union of Crystallography